Product codes

FDA product code

KMJ: Lubricant, Patient (FDA product code)

Follow KMJ: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class I
Regulation
21 CFR 880.6375
Review panel
General Hospital
Premarket route
510(k) exempt
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Not eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
490 in the last 12 months
Median review time
122 daysthe latest 25 decisions
Recall records
170 in the last 12 months
Adverse event reports
25last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code KMJ per year, by decision date, 49 since 1977. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20170
20180
20190
20200
20210
20220
20230
20240
20250
2026 (so far)0

Latest decisions

NumberDevice and applicantDecision and review time
K142473 Lubricating JellySion Biotext Medical , Ltd. Apr 13, 2015222 days
K141028 COLONGLIDE LUBRICANTPediatric Pharmaceuticals, Inc. Jul 24, 201493 days
K131617 COLONGLIDE(R) LUBRICANTPediatric Pharmaceuticals, Inc. Oct 3, 2013122 days
K121390 LUBE JELLY SURINGENurse Assist, Inc. Aug 10, 201293 days
K113689 DUKAL LUBRICATING JELLYDukal Corporation May 4, 2012141 days
K112110 STERILE LUBRICATING JELLYJianerkang Medical Dressing Company Apr 10, 2012263 days
K103718 DYNACOR LUBRICATION GELMedline Industries, Inc. May 10, 2011140 days
K101522 ULTRA SEAL STERILE LUBRICATING JELLYUltra Seal Corporation Dec 23, 2010204 days
K092488 DYNAREX STERILE LUBRICATING JELLYDynarex Corporation Dec 18, 2009127 days
K083189 STERI-LUB LUBRICATION GELAvail Medical Products, Inc. Nov 21, 200823 days

Recalls

Recall records for product code KMJ per year, by the date the recall began. Each recalled product is one record; these 17 records belong to 10 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20170
20180
20190
20200
20210
20221
20230
20240
20250
2026 (so far)0

Latest recalls

Recall and classFirm and reasonBegan
Z-1635-2022Class II Mckesson Medical-Surgical Inc. Corporate OfficeFacility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lo… May 25, 2022
Z-0984-2017Class II MEDLINE INDProduct was not sterilized. Product was shipped to distribution centers instead of the sterilization facility due to an operations error by a 3rd party shippin… Dec 12, 2016
Z-2249-2016Class II Amerx Health Care Corp.Product did not meet specifications for Zinc Acetate (OOS). May 20, 2016
Z-2058-2012Class II Westone Laboratories, Inc.Westone is recalling all Oto Ease ear lubricant in all packaging configurations and sizes due to potential microbial contamination of the product. Jul 13, 2012
Z-2191-2011 Eli Lilly and CompanyEli Lilly and Company announced 3/17/2011 that patients should not use the alcohol prep pads made by the Triad Group that are contained in the black starter ki… Mar 17, 2011

Adverse event reports (MAUDE)

Medical device reports received per year that name product code KMJ. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
20171
20185
20192
20201
20210
20224
20234
20243
20251
2026 (so far)25

Last 12 months by event type

  • Deaths0
  • Injuries24
  • Malfunctions1
  • Other or not stated0

Most reported device problems, last 12 months

  1. Adverse Event Without Identified Device or Use Problem10
  2. Product Quality Problem10
  3. Insufficient Device Problem Information4
  4. Unsealed Device Packaging1

In plain English: what the newest reports describe Written by AI

  • Reports describe Burkholderia cepacia, a bacterium, infections after transrectal ultrasound with prostate biopsy, a procedure that uses an ultrasound probe inserted into the rectum. Two reports describe 10 cases at four urology clinics, mostly found through emergency departments, with 40% becoming septic (a dangerous blood infection) and 80% admitted to hospital.

    6 reports: MW6003372, MW6003373, MW6003377, 1417592-2026-10781, 1417592-2026-10797, 1417592-2026-10674

  • Genetic testing (whole genome sequencing) of bacteria from six of the cases found the samples to be highly related to one another.

    3 reports: MW6003372, MW6003373, MW6003377

  • The reports say all affected patients used the same lubricant jelly, from two lot numbers (41126030002 and 41126030004), sold as 4 oz tubes, bottles, and 0.17 oz foil packets. The affected healthcare systems have stopped using these lots and set them aside.

    3 reports: MW6003372, MW6003373, MW6003377

  • The reporting facilities state that no malfunction has been reported and no product contamination has been confirmed. They also say the information so far does not indicate death or permanent harm, and that follow-up for more details has not produced new information.

    3 reports: 1417592-2026-10781, 1417592-2026-10797, 1417592-2026-10674

  • One report describes a patient with fever, chills, weakness, abdominal and back pain, and sepsis, who was admitted to hospital. Burkholderia cepacia was found in the blood and urine, and the report does not say what treatment was given.

    Report: 1417592-2026-10781

Written by AI from the narratives of the 6 newest reports that have one, received Oct 2, 2026 to Oct 19, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
146in the FDA’s Global Unique Device Identification Database
Establishment listings
140registered establishments that list this product code

Product codes classified under the same regulation.

Watch KMJ in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get KMJ every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in KMJ. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.