FDA product code
KLI: Enzyme Immunoassay, Non-Radiolabeled, Total Thyroxine (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 862.1700
- Review panel
- Clinical Chemistry
- Premarket route
- 510(k) exempt
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 860 in the last 12 months
- Median review time
- 77 daysthe latest 25 decisions
- Recall records
- 121 in the last 12 months
- Adverse event reports
- 93last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code KLI per year, by decision date, 86 since 1977. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K160495 | VITROS® Automation SolutionsOrtho-Clinical Diagnostics, Inc. | Oct 18, 2016239 days |
| K131515 | ACE -GT REAGENT; ACE LIPASE RAGENT; ACE T4 REAGENTAlfa Wassermann | Aug 14, 201378 days |
| K103484 | GSP NEONATAL THYROXINE (T4)Wallac Oy, A Subsidiary of Perkinelmer, Inc. | Apr 22, 2011147 days |
| K082638 | ADVIA CENTAUR SYSTEM WITH STREAMLAB ANALYTICAL WORKCELLSiemens Healthcare Diagnostics, Inc. | Feb 26, 2009169 days |
| K030860 | ACCUBIND NEONATAL T4 MICROPLATE EIA, MODEL 2625-300Monobind | Aug 22, 2003157 days |
| K023369 | ACCESS TOTAL T4 ASSAYBeckman Coulter, Inc. | Dec 9, 200262 days |
| K001603 | REPROBEAD T4 ENZYME IMMUNOASSAY REAGENT KIT (CATALOG# 105)Repromedix Corp. | Oct 5, 2000134 days |
| K000600 | SYVA EMIT 2000 THYROXINE ASSAY, CALIBRATORSSyva Co., Dade Behring, Inc. | Mar 27, 200033 days |
| K991432 | BIOCHECK TOTAL T4 ENZYME IMMUNOASSAY TEST KIT, MODEL BC-1007Biocheck, Inc. | Jun 2, 199937 days |
| K982298 | OPTICOAT T4 EIA KITBiotecx Laboratories, Inc. | Feb 3, 1999217 days |
Recalls
Recall records for product code KLI per year, by the date the recall began. Each recalled product is one record; these 12 records belong to 12 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 1 |
| 2018 | 1 |
| 2019 | 0 |
| 2020 | 1 |
| 2021 | 2 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 1 |
| 2025 | 0 |
| 2026 (so far) | 1 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-1922-2026Class II | Beckman Coulter, Inc.Beckman Coulter has identified that Access Total T4 Calibrator lots (PN 33805) exhibit a negative bias on DxI 600/800 instruments, with slopes outside product… | Mar 16, 2026 |
| Z-1303-2024Class II | Beckman Coulter, Inc.Four lots of Access Total T4 reagent lots demonstrated imprecision outside of the 10% specification which may result in erroneous or delayed results. | Feb 8, 2024 |
| Z-0310-2022Class II | DRG International, Inc.Firm has identified two lots of product that do not meet product shelf life, resulting in failed Controls. The clinical impact may be a delay in diagnosis and… | Oct 18, 2021 |
| Z-0214-2022Class II | BioCheck, Inc.Immunoassay TEST Kit do not meet product shelf-life standards resulting in failed controls. | Oct 14, 2021 |
| Z-1834-2020Class III | Ortho Clinical DiagnosticsThe specified reagent packs exhibit an increase in the frequency of unsuccessful calibrations. | Mar 10, 2020 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code KLI. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 0 |
| 2018 | 3 |
| 2019 | 1 |
| 2020 | 0 |
| 2021 | 3 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 93 |
Last 12 months by event type
- Deaths0
- Injuries0
- Malfunctions93
- Other or not stated0
Most reported device problems, last 12 months
- High Test Results87
- Imprecise or Erratic Results6
In plain English: what the newest reports describe Written by AI
Patients' total thyroxine (TT4, a thyroid hormone) results were reported as too high, and in one case too low, without matching the patients' clinical picture. Repeat testing or testing on another manufacturer's platform gave normal results in these reports.
13 reports: 2122870-2026-00825, 2122870-2026-00843, 2122870-2026-00516, 2122870-2026-00517, 2122870-2026-00480, 2122870-2026-00479, 2122870-2026-00704, 2122870-2026-00515, 2122870-2026-00525, 2122870-2026-00810, 2122870-2026-00748, 2122870-2026-00491, 2122870-2026-00485
Seven reports say the system check, calibration and quality control records were not provided for review. None of the reports describes injury, death or a change in patient treatment.
7 reports: 2122870-2026-00857, 2122870-2026-00843, 2122870-2026-00825, 2122870-2026-00810, 2122870-2026-00704, 2122870-2026-00516, 2122870-2026-00485
Where the manufacturer investigated discordant TT4 results, it found a malfunction but could not determine the cause. One report said medication interference was discussed, and another suspected sample interference and requested interference testing.
4 reports: 2122870-2026-00515, 2122870-2026-00516, 2122870-2026-00704, 2122870-2026-00517
Two reports link high results to calibration done with calibrator lots (the reference material used to set the test) that had already expired. In one, the analyzer's results were higher than those from another analyzer in the same lab.
2 reports: 2122870-2026-00738, 2122870-2026-00529
One report describes erratic repeat results over about two weeks, with about 47 percent of repeats differing by more than 7 percent from the lab's allowed bias, and some samples shifting across reference range limits. The manufacturer sent a specialist to the site.
Report: 2122870-2026-00857
Written by AI from the narratives of the 16 newest reports that have one, received Oct 8, 2026 to Oct 29, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 44in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 67registered establishments that list this product code
Other codes under 21 CFR 862.1700
Product codes classified under the same regulation.
Watch KLI in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get KLI every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in KLI. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:KLIRetrieved Nov 28, 2026, 9:43 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:KLIRetrieved Nov 29, 2026, 7:52 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:KLI · count decision_dateRetrieved Nov 28, 2026, 9:43 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:KLIRetrieved Nov 29, 2026, 7:52 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:KLI · count event_date_initiatedRetrieved Nov 28, 2026, 9:43 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:KLI · count res_event_numberRetrieved Nov 28, 2026, 9:43 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-1922-2026" "Z-1303-2024" "Z-0310-2022" "Z-0214-2022" "Z-1834-2020")Retrieved Nov 29, 2026, 7:52 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:KLI AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 7:52 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:KLI AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 7:52 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:KLI AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 7:52 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:KLIRetrieved Nov 28, 2026, 9:43 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:KLIRetrieved Nov 28, 2026, 9:43 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:862.1700Retrieved Nov 28, 2026, 9:43 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil