FDA product code
KGE: Forceps, Biopsy, Electric (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 876.4300
- Review panel
- Gastroenterology and Urology
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 461 in the last 12 months
- Median review time
- 126 daysthe latest 25 decisions
- Recall records
- 00 in the last 12 months
- Adverse event reports
- 12last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code KGE per year, by decision date, 46 since 1989. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 2 |
| 2018 | 3 |
| 2019 | 1 |
| 2020 | 0 |
| 2021 | 2 |
| 2022 | 1 |
| 2023 | 1 |
| 2024 | 0 |
| 2025 | 2 |
| 2026 (so far) | 1 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K251807 | Single Use Electrosurgical Hemostatic Forceps FD-410LR, FD-411UR, FD-412LROlympus Medical Systems Corp. | Mar 6, 2026267 days |
| K250187 | Disposable Hot Biopsy Forceps (FD-210U, FD-230U)Olympus Medical Systems Corporation | Oct 7, 2025258 days |
| K242857 | ClearHemograsperFinemedix Co., Ltd. | May 21, 2025243 days |
| K231721 | Disposable Hot Biopsy ForcepsBeijing Zksk Technology Co., Ltd. | Dec 14, 2023184 days |
| K220053 | Diathermic Slitter (FlushKnife) DK2620JI and DK2623JIFujifilm Corporation | Feb 4, 202229 days |
| K210406 | Bipolar Coagulation ForecepsHangzhou AGS MedTech Co., Ltd. | Oct 5, 2021237 days |
| K202438 | Ensure Single-Use Coagulation ForcepsMicro-Tech (Nanjing) Co., Ltd. | Mar 15, 2021202 days |
| K183428 | Avulsion ForcepsSTERIS Corporation | Apr 16, 2019126 days |
| K180018 | Disposable Hot Biopsy ForcepsMicro-Tech (Nanjing) Co., Ltd. | Jul 5, 2018184 days |
| K180134 | Ensizor Monopolar Endoscopic ScissorsSlater Endoscopy, LLC | Jun 29, 2018163 days |
Recalls
Recall records for product code KGE per year, by the date the recall began. Each recalled product is one record.
No recalls on record for this product code.
Adverse event reports (MAUDE)
Medical device reports received per year that name product code KGE. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 26 |
| 2018 | 26 |
| 2019 | 15 |
| 2020 | 7 |
| 2021 | 24 |
| 2022 | 16 |
| 2023 | 8 |
| 2024 | 9 |
| 2025 | 10 |
| 2026 (so far) | 8 |
Last 12 months by event type
- Deaths0
- Injuries2
- Malfunctions10
- Other or not stated0
Most reported device problems, last 12 months
- Energy Output Problem3
- Unintended Movement3
- Adverse Event Without Identified Device or Use Problem2
- Failure to Conduct1
- Insufficient Device Problem Information1
- Material Fragmentation1
In plain English: what the newest reports describe Written by AI
In five reports, forceps either sent no electrical current or could not stop bleeding by cautery (burning to seal vessels), and staff finished with another device. One of these was a short circuit, an unintended electrical path.
5 reports: 9610877-2025-00217, 9610877-2025-00215, 9610877-2025-00212, 9610877-2025-00175, 9610877-2025-00176
In four reports, the forceps cup or jaws would not open or close, so the forceps could not be used on the bleeding. One of these reports describes three forceps from the same batch that all failed this way.
4 reports: 9610877-2025-00224, 9610773-2026-00647, 9610877-2025-00007, 9610877-2024-00169
During a long dissection, an artery stuck to the forceps jaws and was torn free, causing renewed bleeding; a second forceps also stuck after about two uses. No patient harm was reported.
2 reports: 9610877-2026-00055, 9610877-2026-00056
In one report, the cup at the tip of the forceps stuck out farther than usual from its sheath during a papillectomy, exposing the internal wire. The wire pulled back when the handle was worked, but the cup stuck out again, so the device was stopped for safety.
Report: 9610877-2025-00070
In one report, metal fragments were found in a single-use hemostatic forceps during an endoscopic submucosal dissection. No patient injury was reported.
Report: 9614641-2026-01019
Written by AI from the narratives of the 19 newest reports that have one, received Nov 12, 2024 to Oct 14, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 175in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 80registered establishments that list this product code
Other codes under 21 CFR 876.4300
Product codes classified under the same regulation.
Watch KGE in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get KGE every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in KGE. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:KGERetrieved Nov 28, 2026, 9:37 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:KGERetrieved Nov 29, 2026, 5:01 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:KGE · count decision_dateRetrieved Nov 28, 2026, 9:37 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:KGERetrieved Nov 29, 2026, 5:01 PM UTC.
- device/recall · search product_code:KGE · count event_date_initiatedRetrieved Nov 28, 2026, 9:37 PM UTC.
- device/recall · search product_code:KGE · count res_event_numberRetrieved Nov 28, 2026, 9:37 PM UTC.
- device/event · search device.device_report_product_code:KGE AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 5:01 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:KGE AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 5:01 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:KGE AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 5:01 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:KGERetrieved Nov 28, 2026, 9:37 PM UTC. FDA data updated Nov 2, 2026.