Product codes

FDA product code

JZG: System, Test, Beta-2-Microglobulin Immunological (FDA product code)

Follow JZG: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 866.5630
Review panel
Immunology
Premarket route
510(k) exempt
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Not eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
530 in the last 12 months
Median review time
180 daysthe latest 25 decisions
Recall records
60 in the last 12 months
Adverse event reports
2last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code JZG per year, by decision date, 53 since 1984. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20170
20180
20190
20200
20210
20220
20230
20240
20250
2026 (so far)0

Latest decisions

NumberDevice and applicantDecision and review time
K161297 Beta-2-MicroglobulinBeckman Coulter Ireland, Inc. Jun 7, 201629 days
K122066 HUMAN BETA-2 MICROGLOBULIN URINE KIT FOR USE ON THE SPAPLUSThe Binding Site Group , Ltd. Aug 30, 201245 days
K110874 ADVIA CHEMISTRY REAGENTS, AND ADVIA CHEMISTRY B2-MICROGLOBULIN CALIBRATORSiemens Healthcare Diagnostics, Inc. Jan 20, 2012297 days
K083289 HUMAN BETA-2 MICROGLOBULIN KITThe Binding Site, Ltd. May 6, 2009180 days
K083463 DIMENSION VISTA BETA 2 MICROGLOBULIN, DIMENSION VISTA PROTEIN 1 CALIBRATOR, DIMENSION VISTA PROTEIN 1 CONTROL MSiemens Healthcare Diagnostics Mar 13, 2009109 days
K072078 QUANTIA BETA-2 MICROGLOBULIN, MODEL: 302822307Biokit, S.A. Dec 19, 2007142 days
K063272 DIMENSION VISTA B2MIC FLEX REAGENT CARTRIDGE, PROTEIN 1 CALIBRATOR, CONTROL MEDIUM, CONTROL HIGHDade Behring, Inc. Jan 26, 200788 days
K050613 QUANTIA BETA-2 MICROGLOBULINBiokit, S.A. Aug 16, 2005159 days
K032692 DAKOCYTOMATION BETA-2-MICROGLOBULIN KIT, CODE NO. OA 368Dako A/S Jan 22, 2004142 days
K014270 MININEPH HUMAN B2-MICROGLOBULIN KITThe Binding Site, Ltd. Aug 15, 2002231 days

Recalls

Recall records for product code JZG per year, by the date the recall began. Each recalled product is one record; these 6 records belong to 6 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20170
20182
20190
20200
20211
20220
20230
20241
20250
2026 (so far)0

Latest recalls

Recall and classFirm and reasonBegan
Z-0599-2025Class II Siemens Healthcare Diagnostics, Inc.Potential for falsely depressed auto-diluted results for samples above the measuring interval of 18.00 mg/L, up to 30.00 mg/L, when using the Atellica CH 2-Mic… Oct 21, 2024
Z-0440-2022Class II Siemens Healthcare Diagnostics, Inc.2 Microglobulin (B2M) Reagent Lot 211446 Bias with Quality Control and Patient Sample Results. Positive bias on patient samples of up to +20.9% at a B2M concen… Dec 1, 2021
Z-1301-2018Class II The Binding Site Group, Ltd.The product is not meeting the antigen excess protection claim of up to 760mg/L, as stated in the Instructions for Use. Mar 9, 2018
Z-2098-2018Class II Tosoh Bioscience IncAsfotase Alfa (Strensiq) interferes with certain Tosoh assays which potentially causes falsely increased or falsely decreased immunoassay test results, dependi… Mar 5, 2018
Z-1030-2016Class II Siemens Healthcare Diagnostics, Inc.Siemens Healthcare Diagnostics has confirmed an increase in the rate of Abnormal Assay errors and calibration failures with the Dimension Vista B2MIC Flex reag… Nov 3, 2015

Adverse event reports (MAUDE)

Medical device reports received per year that name product code JZG. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
20170
20180
20191
20200
20210
20220
20230
20241
20250
2026 (so far)2

Last 12 months by event type

  • Deaths0
  • Injuries0
  • Malfunctions2
  • Other or not stated0

Most reported device problems, last 12 months

  1. Non Reproducible Results2
  2. Incorrect, Inadequate or Imprecise Result or Readings1

In plain English: what the newest reports describe Written by AI

  • A warning flag called E143 (an automatic alert from the analyzer) appeared on calibration samples, patient samples, or both, for a beta-2 microglobulin test cartridge. The reports say the flags did not change patient treatment and show no sign of harm to patients.

    7 reports: 9610806-2015-00055, 9610806-2015-00056, 9610806-2015-00063, 9610806-2015-00065, 9610806-2015-00066, 9610806-2015-00067, 9610806-2015-00068

  • Some results from the analyzer were flagged as unreliable or were questioned, and retesting the samples after dilution (adding liquid to weaken them) gave higher values. One doctor said an initial result did not fit the patient's condition. One report describes two reagent lots giving different values.

    3 reports: 1823260-2026-02309, 1823260-2026-04080, 1823260-2024-02986

  • One report describes a malfunction in which the analyzer gave results that were too low for one female patient. The report does not describe what happened to the patient.

    Report: 1823260-2019-90113

Written by AI from the narratives of the 11 newest reports that have one, received Nov 11, 2015 to Oct 2, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
52in the FDA’s Global Unique Device Identification Database
Establishment listings
49registered establishments that list this product code

Watch JZG in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get JZG every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in JZG. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.