FDA product code
JXM: Enzyme Immunoassay, Benzodiazepine (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 862.3170
- Review panel
- Clinical Toxicology
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 900 in the last 12 months
- Median review time
- 111 daysthe latest 25 decisions
- Recall records
- 810 in the last 12 months
- Adverse event reports
- 2last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code JXM per year, by decision date, 90 since 1979. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 2 |
| 2018 | 1 |
| 2019 | 1 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 1 |
| 2023 | 0 |
| 2024 | 1 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K243498 | Alinity c Benzodiazepines Reagent KitMicrogenics Corporation | Dec 11, 202429 days |
| K221765 | ONLINE DAT Benzodiazepines IIRoche Diagnostics | Dec 23, 2022189 days |
| K190968 | CEDIA Benzodiazepine AssayMicrogenics Corporation | Dec 9, 2019241 days |
| K173963 | DRI Benzodiazepine AssayMicrogenics Corporation | Feb 21, 201855 days |
| K163590 | Psychemedics Microplate EIA for Benzodiazepines in HairPsychemedics Corporation | Sep 15, 2017269 days |
| K163704 | Healgen Multi-Drug Urine Test Cup; Healgen Multi-Drug Urine Test Dip CardHealgen Scientific,, LLC | Aug 2, 2017216 days |
| K153597 | Healgen Multi-Drug Urine Test Cup, Healgen Multi-Drug Urine Test Dip CardHealgen Scientific,, LLC | May 23, 2016159 days |
| K151771 | Immunalysis Benzodiazepines Urine Enzyme Immunoassay, Immunalysis Multi-Drug CalibratorsImmunalysis Corporation | Jan 5, 2016189 days |
| K151642 | Heagen Multi-Drug Urine Test Cup, Healgen Multi-Drug Urine Test Dip CardHealgen Scientific,, LLC | Aug 13, 201556 days |
| K150356 | Healgen Multi-Drug Urine Test Cup, Healgen Multi-Drug Urine Test Dip CardHealgen Scientific,, LLC | May 13, 201590 days |
Recalls
Recall records for product code JXM per year, by the date the recall began. Each recalled product is one record; these 81 records belong to 3 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 2 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-2876-2020Class III | OraSure Technologies, Inc.Incorrect lot number for Standard Oral Fluid Negative Calibrator on specification sheet included in kit and/or additionally same error on unit box | Jul 13, 2020 |
| Z-2877-2020Class III | OraSure Technologies, Inc.Incorrect lot number for Standard Oral Fluid Negative Calibrator on specification sheet included in kit and/or additionally same error on unit box | Jul 13, 2020 |
| Z-0484-2016Class III | Ameditech IncAmeditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which spec… | Nov 16, 2015 |
| Z-0485-2016Class III | Ameditech IncAmeditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which spec… | Nov 16, 2015 |
| Z-0486-2016Class III | Ameditech IncAmeditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which spec… | Nov 16, 2015 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code JXM. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 4 |
| 2018 | 4 |
| 2019 | 3 |
| 2020 | 0 |
| 2021 | 1 |
| 2022 | 0 |
| 2023 | 5 |
| 2024 | 6 |
| 2025 | 6 |
| 2026 (so far) | 2 |
Last 12 months by event type
- Deaths0
- Injuries0
- Malfunctions2
- Other or not stated0
Most reported device problems, last 12 months
- False Negative Result2
In plain English: what the newest reports describe Written by AI
Several reports describe the benzodiazepine test giving a positive result for patient urine or blood samples that confirmatory testing by LC-MS or GC/MS found negative, meaning the lab test flagged benzodiazepines that the more specific methods did not find.
10 reports: 1823260-2025-02507, 1823260-2025-02131, 1823260-2025-01596, 1823260-2025-00241, 1823260-2024-03017, 1823260-2024-01302, 1823260-2024-01265, 1823260-2023-02317, 1823260-2023-00687, 1823260-2024-00027
Several reports describe the test giving a negative result for samples that other methods found positive for benzodiazepines, with reported values far below the cutoff, such as 143 and 129 ng/mL where positive was expected. The manufacturer's corrected reports reclassified some of these results, and in one case the customer reported no treatment changed based on the result.
6 reports: 1319808-2026-00033, 1319808-2024-00040, 1319808-2021-00023, 1319808-2019-00037, 1823260-2023-02183, 1823260-2023-00762
A few reports say the results disagreed between the test instruments, with the same sample giving different values on two analyzers. One report describes a sample reading 59.5 on one analyzer and then a different value on another. Reports do not state a cause.
2 reports: 1823260-2025-02131, 1823260-2024-00027
One report describes a discordant negative result compared with confirmatory testing, with the customer stating that no patient treatment was changed. The report says there was no allegation of patient harm.
Report: 1319808-2026-00033
One report describes a new reagent version giving positive results where the older reagent gave negative results on four samples, which were then confirmed positive by mass spectrometry. The reporter did not state a cause.
Report: 1823260-2023-02403
Written by AI from the narratives of the 18 newest reports that have one, received Dec 23, 2019 to Jun 18, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 3,690in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 242registered establishments that list this product code
Other codes under 21 CFR 862.3170
Product codes classified under the same regulation.
Watch JXM in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get JXM every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in JXM. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:JXMRetrieved Nov 29, 2026, 5:43 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:JXMRetrieved Nov 29, 2026, 5:43 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:JXM · count decision_dateRetrieved Nov 29, 2026, 5:43 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:JXMRetrieved Nov 29, 2026, 5:43 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:JXM · count event_date_initiatedRetrieved Nov 29, 2026, 5:43 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:JXM · count res_event_numberRetrieved Nov 29, 2026, 5:43 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-2876-2020" "Z-2877-2020" "Z-0484-2016" "Z-0485-2016" "Z-0486-2016")Retrieved Nov 29, 2026, 5:43 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:JXM AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 5:43 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:JXM AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 5:43 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:JXM AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 5:43 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:JXMRetrieved Nov 29, 2026, 5:43 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:JXMRetrieved Nov 29, 2026, 5:43 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:862.3170Retrieved Nov 29, 2026, 5:43 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil