FDA product code
JPD: Hemoglobin A2 Quantitation (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 864.7400
- Review panel
- Hematology
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 130 in the last 12 months
- Median review time
- 69 daysthe latest 13 decisions
- Recall records
- 50 in the last 12 months
- Adverse event reports
- 0last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code JPD per year, by decision date, 13 since 1980. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K020489 | G7 AUTOMATED HPLC ANALYZER, BETA-THALASSEMIA MODETosoh Medics, Inc. | May 14, 200290 days |
| K991127 | VARIANT II B-THALASSEMIABio-Rad | Jun 10, 199969 days |
| K924122 | VARIANT BETA-THALASSEMIA SHORT PROGRAMBio-Rad | Mar 11, 1993209 days |
| K924496 | GLYCOMAT GREEN HBA2 ESTIMATION REAGENT KITDrew Scientific , Ltd. | Nov 16, 199273 days |
| K923318 | COLUMNMATE BETA-THAL HBA2 CONTROL - NORMALHelena Laboratories | Aug 25, 199249 days |
| K923317 | COLUMNMATE BETA THAL HBA2 CONTROL-ABNORMALHelena Laboratories | Aug 25, 199249 days |
| K913384 | COLUMNMATE(TM) BETA-THAL CAT. NO. 4515Helena Laboratories | Dec 17, 1991141 days |
| K896168 | HEMOGLOBIN A2 MICRO COLUMN TESTBio-Rad | Jan 18, 199085 days |
| K875102 | DETER. OF HEMOGLOBIN A2 USING THE DIAMATBio-Rad | Feb 24, 198876 days |
| K823870 | BETA THALQUIK COLUMN #5341Helena Laboratories | Feb 1, 198340 days |
Recalls
Recall records for product code JPD per year, by the date the recall began. Each recalled product is one record; these 5 records belong to 4 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-1537-2014Class II | Bio-Rad Laboratories, Inc.There is an issue with the buffers which impacts the quantification of the Hemoglobin F peak. Calibration and Quality control are likely to fail if the affecte… | Apr 8, 2014 |
| Z-3030-2011 | Tosoh Bioscience IncThe value after reconstitution appeared to be lower than the assigned value on the label. The measured value for Lot BCAL05R was 5.3%, whereas the value on the… | May 25, 2011 |
| Z-1075-2013Class III | Tosoh Bioscience IncThe Beta-thalessemia Mode of the G7 Analyzer-Beta Thalassemia Mode is an automated High Performance Liquid Chromatography (HPLC) system that separates and repo… | Aug 18, 2010 |
| Z-1588-2008 | Bio-Rad Laboratories IncSome product ROM cards have an incorrect program loaded and are unable to be updated. | Mar 18, 2008 |
| Z-1589-2008 | Bio-Rad Laboratories IncSome product ROM cards have an incorrect program loaded and are unable to be updated. | Mar 18, 2008 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code JPD. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
No adverse event reports received in the last ten years.
Last 12 months by event type
No reports received in the last 12 months.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 21in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 9registered establishments that list this product code
Watch JPD in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get JPD every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in JPD. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:JPDRetrieved Nov 29, 2026, 5:15 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:JPDRetrieved Nov 29, 2026, 5:15 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:JPD · count decision_dateRetrieved Nov 29, 2026, 5:15 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:JPDRetrieved Nov 29, 2026, 5:15 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:JPD · count event_date_initiatedRetrieved Nov 29, 2026, 5:15 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:JPD · count res_event_numberRetrieved Nov 29, 2026, 5:15 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-1537-2014" "Z-3030-2011" "Z-1075-2013" "Z-1588-2008" "Z-1589-2008")Retrieved Nov 29, 2026, 5:15 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:JPD AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 5:15 PM UTC.
- device/event · search device.device_report_product_code:JPD AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 5:15 PM UTC.
- device/event · search device.device_report_product_code:JPD AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 5:15 PM UTC.
- device/udi · search product_codes.code:JPDRetrieved Nov 29, 2026, 5:15 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:JPDRetrieved Nov 29, 2026, 5:15 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:864.7400Retrieved Nov 29, 2026, 5:15 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.