Product codes

FDA product code

JOQ: Generator, Pulse, Pacemaker, External Programmable (For Electrophysiological Studies Only) (FDA product code)

Follow JOQ: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 870.1750
Review panel
Cardiovascular
Premarket route
510(k)
Implant
No
Life-sustaining or supporting
Yes
GMP exempt
No
Third-party review
Not eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
180 in the last 12 months
Median review time
107 daysthe latest 18 decisions
Recall records
30 in the last 12 months
Adverse event reports
3last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code JOQ per year, by decision date, 18 since 1984. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20170
20181
20190
20200
20210
20220
20230
20240
20250
2026 (so far)0

Latest decisions

NumberDevice and applicantDecision and review time
K173439 Bloom2 Programmable Electrophysiology StimulatorFischer Medical Mar 29, 2018146 days
K092913 EP-4TM CARDIAC STIMULATOR, MODELS: EP-04-02, EP-04-04Ep Medsystems Oct 16, 200924 days
K072200 STIMLABMicropace Pty, Ltd. Oct 26, 200779 days
K070743 Z6 STIMULATORSt Cardio Technologies, LLC Jul 2, 2007108 days
K041442 EP-4 CLINICAL STIMUTATOREp Medsystems Sep 14, 2004105 days
K040207 EP-4 CLINICAL STIMULATOREp Medsystems Feb 27, 200428 days
K011826 EPS320 CARDIAC STIMULATORMicropace Pty, Ltd. Jan 24, 2002227 days
K991293 EPSTIM ELECTROPHYSIOLOGY STIMULATOR CONTROL PROGRAM, BLOOM DTU-215 ELECTROPHYSIOLOGY PROGRAMMABLE STIMULATORFischer Medical Technologies, Inc. May 5, 2000386 days
K963579 PACE-1A CARDIAC STIMULATORRadionics, Inc. Nov 8, 199663 days
K935590 EP MEDICAL, INC. EP-3 CLINICAL STIMULATOREp Medical, Inc. May 5, 1994168 days

Recalls

Recall records for product code JOQ per year, by the date the recall began. Each recalled product is one record; these 3 records belong to 3 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20170
20180
20190
20200
20210
20220
20230
20240
20250
2026 (so far)0

Latest recalls

Recall and classFirm and reasonBegan
Z-1422-2009 St. Jude MedicalNon-compliance with IEC Standard 60601-1 Clause 57.6 relative to fusing of AC input line. The neutral line is not fused. The hot line in fused. Clause 57.6 cal… Mar 23, 2009
Recall EP MedsystemsUnder certain circumstances the device may exhibit the following behaviors: 1)-Initiation of a rapid stimulation train without command, when the device is in u… Apr 4, 2006
Recall Fischer Imaging CorporationPossibility that after a loss of power or brown-out, the device may transmit unintended stimulation to patients. Jan 13, 2005

Adverse event reports (MAUDE)

Medical device reports received per year that name product code JOQ. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
20174
20183
20194
20202
20212
20224
202310
20248
20255
2026 (so far)2

Last 12 months by event type

  • Deaths0
  • Injuries0
  • Malfunctions3
  • Other or not stated0

Most reported device problems, last 12 months

  1. Pacing Problem2
  2. Communication or Transmission Problem1

In plain English: what the newest reports describe Written by AI

  • In several reports, the stimulator (the box that sends electrical pacing signals to the heart) stopped communicating with the computer or touchscreen, so pacing could not be done and the procedure was cancelled or stopped. Troubleshooting such as rebooting, checking or swapping cables, and replacing parts did not always fix the problem.

    4 reports: 2184149-2026-00222, 2184149-2025-00213, 2184149-2026-00049, 2124215-2023-54638

  • In a few reports, the stimulator or its display shut down, lost power, or kept rebooting during a procedure. Replacing or servicing the unit was sometimes needed, and in some cases the procedure was cancelled or rescheduled.

    4 reports: 2184149-2025-00119, 2124215-2024-13469, 2124215-2023-72188, 2124215-2023-44377

  • In one report, the computer's stimulation port was damaged, which blocked the touchscreen from connecting to the device. The manufacturer's and the reporter's account of this is that the port had collapsed, and the reporter alleged the computer contributed to cancelling the procedure.

    Report: 2184149-2026-00049

  • In a report, the pacing function stopped working after pulse field ablation (a heart procedure that uses electrical pulses to treat irregular heartbeat), and no backup pacing was available. The reporter did not say whether the system would be returned for analysis.

    Report: 2124215-2024-82776

  • In one report, a cable stopped working, and the reporter was not sure whether it had shorted out or broken. The reporter noted the device was the mapping system used with the procedure.

    Report: MW5161022

Written by AI from the narratives of the 19 newest reports that have one, received Aug 22, 2023 to Aug 24, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
25in the FDA’s Global Unique Device Identification Database
Establishment listings
12registered establishments that list this product code

Watch JOQ in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get JOQ every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in JOQ. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.