Product codes

FDA product code

JDL: Prosthesis, Hip, Semi-Constrained (Metal Cemented Acetabular Component) (FDA product code)

Follow JDL: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class III
Regulation
21 CFR 888.3320
Review panel
Orthopedic
Premarket route
PMA
Implant
Yes
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Not eligible
Summary malfunction reporting
Eligible
PMA decisions
00 original PMAs, the rest supplements
Median review time
Not availablethe latest 0 decisions
Recall records
51 in the last 12 months
Adverse event reports
25last 12 months, 0 deaths

PMA decisions

Original PMAs and supplements for product code JDL per year, by decision date. Review time is the days from receipt to decision.

All of them in the FDA’s PMA database

No decisions on record for this product code.

Recalls

Recall records for product code JDL per year, by the date the recall began. Each recalled product is one record; these 5 records belong to 4 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20170
20180
20190
20200
20210
20220
20230
20240
20250
2026 (so far)1

Latest recalls

Recall and classFirm and reasonBegan
Z-3196-2026Class II Howmedica Osteonics Corp.Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential… Jul 30, 2026
Z-2212-2010 Wright Medical Technology IncThe inner and outer product packaging was missing the following information: shelf life, translations, manufacturing date, CE marking and also listed the incor… Jun 7, 2010
Z-1715-2009 Wright Medical Technology IncIncorrect packaging, swapped sizes May 14, 2009
Z-1716-2009 Wright Medical Technology IncIncorrect packaging, swapped sizes May 14, 2009
Recall Stryker Howmedica OsteonicsThis device is not approved for sale in the US. Jan 9, 2004

Adverse event reports (MAUDE)

Medical device reports received per year that name product code JDL. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
2017336
2018205
2019140
202094
202168
202249
202332
202427
202516
2026 (so far)22

Last 12 months by event type

  • Deaths0
  • Injuries25
  • Malfunctions0
  • Other or not stated0

Most reported device problems, last 12 months

  1. Insufficient Device Problem Information10
  2. Corroded7
  3. Break4
  4. Entrapment of Device3
  5. Leak/Splash3
  6. Material Fragmentation3

In plain English: what the newest reports describe Written by AI

  • Five reports describe revision surgery for infection, and one of these describes extractor tools used to remove a neck and stem.

    5 reports: 3010536692-2026-00406, 3010536692-2026-00404, 3010536692-2026-00398, 3010536692-2026-00317, 3010536692-2026-00449

  • In three reports, the joint where the metal neck meets the stem (a modular taper) corroded, releasing metal debris and metal ions into the body. Surgeons described metal staining, metallosis (metal debris in tissue), and tissue damage or a pseudotumor (a mass of inflamed tissue).

    3 reports: 3010536692-2026-00485, 3010536692-2026-00466, 3010536692-2026-00198

  • Three reports describe the femoral stem or modular neck breaking. One patient felt a sudden pop in the hip and fell, and another report describes a broken neck alongside metallosis.

    3 reports: 3010536692-2026-00477, 3010536692-2026-00073, 3010536692-2026-00321

  • Two matching reports from the same reporter describe cobalt metal leaking into the bloodstream, with fatigue, headaches and trouble walking. The reporter links these to the implant, but the narrative does not confirm the cause of the symptoms.

    2 reports: 3010536692-2026-00226, MW5188325

  • Two reports describe an adverse soft tissue reaction to particulate debris, which is tissue irritation caused by tiny wear particles. One of them also describes the hip dislocating.

    2 reports: 3010536692-2026-00264, 3010536692-2026-00282

Written by AI from the narratives of the 18 newest reports that have one, received Apr 10, 2026 to Oct 5, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
575in the FDA’s Global Unique Device Identification Database
Establishment listings
236registered establishments that list this product code

Product codes classified under the same regulation.

Watch JDL in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get JDL every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in JDL. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.