FDA product code
JCT: Prosthesis, Tracheal, Expandable (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 878.3720
- Review panel
- Anesthesiology
- Premarket route
- 510(k)
- Implant
- Yes
- Life-sustaining or supporting
- Yes
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 450 in the last 12 months
- Median review time
- 125 daysthe latest 25 decisions
- Recall records
- 610 in the last 12 months
510(k) and De Novo decisions
Decisions for product code JCT per year, by decision date, 45 since 1991. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 0 |
| 2018 | 1 |
| 2019 | 0 |
| 2020 | 1 |
| 2021 | 2 |
| 2022 | 1 |
| 2023 | 1 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K230269 | Ultraflex™ Tracheobronchial Stent SystemBoston Scientific Corporation | Jul 14, 2023164 days |
| K220424 | Through the Scope Tracheal Stent SystemMicro-Tech (Nanjing) Co., Ltd. | Jul 8, 2022144 days |
| K212403 | Tracheal Stent System (Y-Shaped)Micro-Tech (Nanjing) Co., Ltd. | Oct 30, 202189 days |
| K202204 | Tracheal Stent SystemMicro-Tech (Nanjing) Co., Ltd. | Sep 1, 2021392 days |
| K201342 | HANAROSTENT Trachea/Bronchium (CCC)M.I.Tech Co., Ltd. | Dec 31, 2020225 days |
| K181200 | AEROmini Tracheobronchial Stent SystemMerit Medical Systems, Inc. | Sep 21, 2018137 days |
| K152842 | WALLSTENT RP ENDOPROSTHESIS TRACHEOBRONCHIAL, WALLSTENT ENDOPROSTHESIS TRACHEOBRONCHIALBoston Scientific Corporation | May 19, 2016233 days |
| K140382 | AEROMINI TRACHEOBRONCHIAL STENT TECHNOLOGYMerit Medical Systems, Inc. | Nov 25, 2014284 days |
| K140472 | BONASTENT TRACHEAL / BRONCHIALEndochoice, Inc. | Oct 17, 2014234 days |
| K141584 | ULTRAFLEX TRACHEOBRONCHIAL COVERED STENT SYSTEMBoston Scientific Corporation | Oct 16, 2014125 days |
Recalls
Recall records for product code JCT per year, by the date the recall began. Each recalled product is one record; these 61 records belong to 9 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 1 |
| 2018 | 5 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 2 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-1076-2022Class I | Atrium Medical CorporationFirm has observed increased rate of customer complaints related to separation of the balloon or catheter hub from the delivery catheter during delivery system… | Mar 3, 2022 |
| Z-1077-2022Class I | Atrium Medical CorporationFirm has observed increased rate of customer complaints related to separation of the balloon or catheter hub from the delivery catheter during delivery system… | Mar 3, 2022 |
| Z-3035-2018Class II | SPS Sterilization, IncThe expiration date on the device labeling exceeds its actual validated shelf life because the sterilization process used does not have adequate support for th… | Mar 7, 2018 |
| Z-3036-2018Class II | SPS Sterilization, IncThe expiration date on the device labeling exceeds its actual validated shelf life because the sterilization process used does not have adequate support for th… | Mar 7, 2018 |
| Z-3037-2018Class II | SPS Sterilization, IncThe expiration date on the device labeling exceeds its actual validated shelf life because the sterilization process used does not have adequate support for th… | Mar 7, 2018 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code JCT. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 109 |
| 2018 | 122 |
| 2019 | 106 |
| 2020 | 120 |
| 2021 | 106 |
| 2022 | 61 |
| 2023 | 110 |
| 2024 | 76 |
| 2025 | 50 |
| 2026 (so far) | 64 |
Last 12 months by event type
This section didn’t load from openFDA. It’s retried within the hour.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 460in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 73registered establishments that list this product code
Other codes under 21 CFR 878.3720
Product codes classified under the same regulation.
Watch JCT in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get JCT every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in JCT. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:JCTRetrieved Nov 24, 2026, 12:42 PM UTC. FDA data updated Nov 9, 2026.
- device/510k · search product_code:JCTRetrieved Nov 24, 2026, 12:42 PM UTC. FDA data updated Nov 9, 2026.
- device/510k · search product_code:JCT · count decision_dateRetrieved Nov 24, 2026, 12:42 PM UTC. FDA data updated Nov 9, 2026.
- device/recall · search product_code:JCTRetrieved Nov 24, 2026, 12:42 PM UTC. FDA data updated Nov 19, 2026.
- device/recall · search product_code:JCT · count event_date_initiatedRetrieved Nov 24, 2026, 12:42 PM UTC. FDA data updated Nov 19, 2026.
- device/recall · search product_code:JCT · count res_event_numberRetrieved Nov 24, 2026, 12:42 PM UTC. FDA data updated Nov 19, 2026.
- device/enforcement · search recall_number:("Z-1076-2022" "Z-1077-2022" "Z-3035-2018" "Z-3036-2018" "Z-3037-2018")Retrieved Nov 24, 2026, 12:42 PM UTC. FDA data updated Nov 11, 2026.
- device/event · search device.device_report_product_code:JCT AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 6:07 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:JCTRetrieved Nov 24, 2026, 12:42 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:JCTRetrieved Nov 24, 2026, 12:42 PM UTC. FDA data updated Nov 9, 2026.
- device/classification · search regulation_number:878.3720Retrieved Nov 28, 2026, 11:57 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.