Product codes

FDA product code

JAI: Couch, Radiation Therapy, Powered (FDA product code)

Follow JAI: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 892.5770
Review panel
Radiology
Premarket route
510(k)
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
320 in the last 12 months
Median review time
59 daysthe latest 25 decisions
Recall records
180 in the last 12 months
Adverse event reports
1last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code JAI per year, by decision date, 32 since 1983. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20171
20180
20191
20200
20210
20220
20230
20241
20250
2026 (so far)0

Latest decisions

NumberDevice and applicantDecision and review time
K231612 Eve Patient Positioner SystemLeo Cancer Care May 7, 2024340 days
K182683 Z-Robot Patient Positioning SystemChinan Biomedical Technology, Inc. Jun 27, 2019274 days
K170502 Patient Positioning SystemForte Automation Systems, Inc. Apr 13, 201751 days
K160518 LEONI ORION SystemLeoni Cia Cable System Jul 14, 2016141 days
K132084 PROTURA COUCH SOFTWAREMed-Tec, Inc. D/B/A Civco Medical Solutions Sep 26, 201383 days
K122413 PATIENT POSITIONING SYSTEMForte Automation Systems, Inc. Sep 13, 201236 days
K122201 PROTURA COUCH SOFTWAREMed-Tec, Inc. D/B/A Civco Medical Solutions Aug 30, 201236 days
K120520 EXACTRAC ROBOTICS NOVALIS ROBOTICSBrainlab AG Jun 29, 2012129 days
K120916 QFIX RADIOTHERAPY CHAIRWfr/Aquaplast Corp. Jun 20, 201285 days
K072898 HEXAPOD EVO RT COUCH TOPMedical Intelligence Medizintechnik GmbH Oct 25, 200714 days

Recalls

Recall records for product code JAI per year, by the date the recall began. Each recalled product is one record; these 18 records belong to 17 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20172
20181
20190
20200
20210
20220
20232
20240
20250
2026 (so far)0

Latest recalls

Recall and classFirm and reasonBegan
Z-0398-2024Class II LEONI CIA CABLE SYSTEMSUsers of the patient positioning system ORION System can be faced with a sudden short-term acceleration or with an unintended short-distance movement when the… Oct 6, 2023
Z-0399-2024Class II LEONI CIA CABLE SYSTEMSUsers of the patient positioning system ORION System can be faced with a sudden short-term acceleration or with an unintended short-distance movement when the… Oct 6, 2023
Z-2535-2018Class II Elekta, Inc.Upon completion of a 6D workflow, the HexaPOD evo Module may be in a tilted state if it is moved to the position used for 3D (non-iGUIDE) treatments resulting… May 16, 2018
Z-0671-2018Class II Med Tec IncWhen an error message remains displayed and is not cleared in the Protura software with the Elekta pedestal coordinates, moving the pedestal could cause the Pr… Nov 8, 2017
Z-2537-2018Class II Forte Automation Systems IncCommunications between the Patient Positioning System and the accuracy filter can periodically fail with no clear indication to the operator. Sep 5, 2017

Adverse event reports (MAUDE)

Medical device reports received per year that name product code JAI. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
20173
20180
20191
20200
20210
20220
20230
20241
20250
2026 (so far)1

Last 12 months by event type

  • Deaths0
  • Injuries0
  • Malfunctions1
  • Other or not stated0

Most reported device problems, last 12 months

  1. Fracture1
  2. Mechanical Problem1

In plain English: what the newest reports describe Written by AI

  • Several reports describe the couch ending up off its intended position, often after calibration or servicing, by amounts from a few millimetres to several centimetres. In one case a height calibration error was not caught until patients had been treated, with the largest offset reported as 16 cm.

    8 reports: 9617016-2013-00019, 9617016-2013-00012, 9617016-2013-00015, 3002466018-2019-84764, 3003900998-2017-00001, 1932738-2016-00001, MW5060337, 9617016-2013-00017

  • Several reports describe the treatment couch moving when it should not, sometimes shifting the patient's position. In one case a computer waking from sleep triggered a 5 cm vertical shift, and in another the couch moved during treatment and had to be halted until the patient was realigned.

    6 reports: 1932738-2012-00001, 1932738-2017-00002, 2910081-2012-10524, 9617016-2013-00002, 9710332-2014-00002, 2950679-2009-00005

  • One report describes a patient who slid off the couch after a side section came unlatched while he turned to vomit. Staff tried to catch him but could not, and he had head pain and was sent for a head CT scan.

    Report: 18659007

  • One report describes a gearbox part failing, causing the couchtop to tilt forward to the floor while a patient lay on it, held in place by a restraint mask. The manufacturer said no injury occurred, and the root cause was still under investigation.

    Report: 3012834372-2026-00001

  • One report describes a faulty power supply in the couch that caused nearby components to smolder and emit heavy smoke. The manufacturer said no injury or mistreatment was reported.

    Report: 2910081-2009-00079

Written by AI from the narratives of the 20 newest reports that have one, received Jul 31, 2009 to Oct 20, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
352in the FDA’s Global Unique Device Identification Database
Establishment listings
24registered establishments that list this product code

Watch JAI in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get JAI every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in JAI. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.