Product codes

FDA product code

IYB: Tube Mount, X-Ray, Diagnostic (FDA product code)

Follow IYB: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class I
Regulation
21 CFR 892.1770
Review panel
Radiology
Premarket route
510(k) exempt
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Not eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
400 in the last 12 months
Median review time
45 daysthe latest 25 decisions
Recall records
130 in the last 12 months
Adverse event reports
2last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code IYB per year, by decision date, 40 since 1978. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20170
20180
20190
20200
20210
20220
20230
20240
20250
2026 (so far)0

Latest decisions

NumberDevice and applicantDecision and review time
K935123 SUPERSTAND 300Wuestec Medical, Inc. Mar 27, 1996881 days
K946261 MTS-007, STS-008, SSMTS-010, EURO-40, 72, SUPER EURO 4072, 5000Scott Imaging Intl., Inc. Feb 15, 199564 days
K941004 CTS SYSTEMAcoma Medical Imaging, Inc. Jul 29, 1994150 days
K941017 F/C/W TUBESTANDAcoma Medical Imaging, Inc. Jul 29, 1994148 days
K940442 MPR-814Xri, Inc. Jun 9, 1994128 days
K935004 MODEL ALPHA TS-2Dpa Consulting, Inc. Dec 13, 199355 days
K933253 SUPERSTAND 3000Wuestec Medical, Inc. Aug 16, 199345 days
K925716 BENNETT OVERHEAD TUBE CRANE MODEL NO. B-OTCBennett X-Ray Corp. Dec 17, 199235 days
K925170 SHIMADZU X-RAY TUBE STAND BH-100 SERIESShimadzu Medical Systems Dec 7, 199255 days
K920727 CEILING TUBE SUPPORT CH-30XShimadzu Medical Systems Apr 16, 199258 days

Recalls

Recall records for product code IYB per year, by the date the recall began. Each recalled product is one record; these 13 records belong to 12 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20170
20182
20190
20200
20210
20220
20231
20240
20251
2026 (so far)0

Latest recalls

Recall and classFirm and reasonBegan
Z-0499-2026Class II SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISIONDue to an issue (potential breakage) with the shaft for the mounting of the X-ray tube assembly that could potentially result assembly coming off the rotating… Oct 16, 2025
Z-2270-2023Class II SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISIONEight fixing bolts at the base of the X-ray ceiling tube column assembly may be loose or missing so service is required to install a cable, bracket, and label… Apr 10, 2023
Z-2837-2018Class II Siemens Medical Solutions USA, IncIn rare cases of insufficient maintenance or high clinical workload, the first rope of the ceiling stand, which is designed to take the load, could break witho… Jun 29, 2018
Z-2838-2018Class II Siemens Medical Solutions USA, IncIn rare cases of insufficient maintenance or high clinical workload, the first rope of the ceiling stand, which is designed to take the load, could break witho… Jun 29, 2018
Z-0468-2015Class II Summit Industries Inc.The welds on J700 tube stands may be insufficient resulting in tube stand breakage, damage to the X-Ray machine, and/or patient or healthcare provider injury. Oct 28, 2014

Adverse event reports (MAUDE)

Medical device reports received per year that name product code IYB. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
20170
20181
20190
20200
20210
20220
20230
20240
20250
2026 (so far)2

Last 12 months by event type

  • Deaths0
  • Injuries0
  • Malfunctions2
  • Other or not stated0

Most reported device problems, last 12 months

  1. Unintended Movement1

In plain English: what the newest reports describe Written by AI

  • Several reports describe mounted arms, monitors and supports that broke loose or drifted down and hit staff, a patient or a bystander. One report describes a worker with a concussion after a drifting arm struck their head.

    5 reports: 1836145-2009-00003, 9611343-2005-00009, 1518293-2004-00019, 668099, 1221300-2007-00007

  • Several reports describe a ceiling-mounted X-ray tube that kept moving down or dropped without the operator commanding it, and in some cases struck people. In one report the tube kept descending after the button was released and hit a patient's knee, and in another a tube crane fell to the ground with no operator command.

    4 reports: 3005920099-2006-00001, 23943089, 9617251-2018-00001, 1418964-2015-00002

  • Two reports describe a tube crane or tube stand moving down onto a patient during positioning. In one, the operator pressed two buttons at once and could not release them quickly enough, so the crane hit a patient's hip with some force; no injury was diagnosed.

    2 reports: 1000404222-2009-00002, 1000404222-2009-00001

  • In a report about a power fault, the tube stand fell onto a patient after the locks that hold it in place lost power. The reporter said the fault was traced to a failed power supply part inside the generator.

    Report: MW5032245

  • A report describes a tube stand that fell onto a patient after a counterweight cable came loose because an anchor pin in the column was worn. The report does not state whether the patient was hurt.

    Report: 1418964-2008-00001

Written by AI from the narratives of the 20 newest reports that have one, received Nov 3, 2004 to Feb 26, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
95in the FDA’s Global Unique Device Identification Database
Establishment listings
37registered establishments that list this product code

Watch IYB in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get IYB every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in IYB. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.