FDA product code
IQA: System, Environmental Control, Powered (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 890.3725
- Review panel
- Physical Medicine
- Premarket route
- 510(k) exempt
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 240 in the last 12 months
- Median review time
- 59 daysthe latest 24 decisions
- Recall records
- 172 in the last 12 months
- Adverse event reports
- 1last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code IQA per year, by decision date, 24 since 1977. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K961306 | DELTATALKER WITH INFRAREDPrentke Romich Co. | Jun 27, 199684 days |
| K953905 | IMPERIUMTeledyne Brown Engineering | Nov 21, 199595 days |
| K953937 | LIAISON COMPUTER ACCESS & ENVIRONMENTAL CONTROL SYSTEMApt Technology, Inc. | Oct 31, 199574 days |
| K943244 | DEUCEApt Technology, Inc. | Sep 11, 1995432 days |
| K944585 | SCANNING DIRECTORPrentke Romich Co. | Feb 21, 1995155 days |
| K944345 | BEDSIDE BUTLERTrac Medical, Inc. | Oct 17, 199441 days |
| K940223 | LEARN MODE WATERPROOF PENDANT PANIC BUTTONInteractive Technologies, Inc. | Apr 29, 1994101 days |
| K926029 | INFRA-LINKZygo Industries, Inc. | Mar 26, 1993115 days |
| K921528 | HILL-ROM PATIENT ENVIRONMENTAL CONTROL SYSTEMHill-Rom, Inc. | Apr 20, 199218 days |
| K911101 | TONGUETOUCH KEYPAD AND ZOFCOM CONTROL SYSTEMZofcom, Inc. | Apr 25, 199143 days |
Recalls
Recall records for product code IQA per year, by the date the recall began. Each recalled product is one record; these 17 records belong to 13 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 1 |
| 2019 | 1 |
| 2020 | 0 |
| 2021 | 1 |
| 2022 | 0 |
| 2023 | 4 |
| 2024 | 4 |
| 2025 | 2 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-1031-2026Class II | Securitas Healthcare LLCSecuritas Healthcare has become aware of a firmware issue in the Call Stations in the Affected Lots that may cause Call Stations in the Affected Lots to (1) no… | Dec 8, 2025 |
| Z-1032-2026Class II | Securitas Healthcare LLCSecuritas Healthcare has become aware of a firmware issue in the Call Stations in the Affected Lots that may cause Call Stations in the Affected Lots to (1) no… | Dec 8, 2025 |
| Z-0570-2025Class II | Securitas Healthcare LLCCertain software versions will not notify end users of calls initiated by call points with the Arial Emergency and Nurse Call System. | Nov 1, 2024 |
| Z-2417-2024Class II | Securitas Healthcare LLCThere is a risk that alarms will not be received by the Arial Wireless Emergency Call System and premature depletion of the battery. | Jun 20, 2024 |
| Z-2418-2024Class II | Securitas Healthcare LLCThis is a risk that alarms will not be received by the Arial Wireless Nurse Call System and premature depletion of the battery. | Jun 20, 2024 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code IQA. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 1 |
| 2018 | 1 |
| 2019 | 0 |
| 2020 | 1 |
| 2021 | 0 |
| 2022 | 2 |
| 2023 | 18 |
| 2024 | 2 |
| 2025 | 2 |
| 2026 (so far) | 1 |
Last 12 months by event type
- Deaths0
- Injuries0
- Malfunctions1
- Other or not stated0
Most reported device problems, last 12 months
- Application Program Version or Upgrade Problem1
In plain English: what the newest reports describe Written by AI
Nurse call patient stations had buttons, such as Code Blue and stat assist, that did not register or light up. The manufacturer's failure analysis of returned stations found corrosion or damage to the flat flexible cable that carries signals to the keypad, which can cause loss of button or light function.
8 reports: 3009164985-2023-00010, 3009164985-2023-00011, 3009164985-2023-00012, 3009164985-2023-00014, 3009164985-2023-00015, 3009164985-2023-00017, 3009164985-2023-00019, 3009164985-2023-00020
In two reports from one facility, a resident's pendant or call light did not reliably summon staff. In one case the pendant showed a press but reset within a minute, and no pages were received; in the other, the call light switched itself on and off. A resident was reported to have fallen in one case.
2 reports: 1929691-2024-00002, 1929691-2024-00003
In one report, devices in a hospital room went offline, and the system's alert for the offline devices reached a staff console but was not answered. A patient was found unresponsive in that room, and the report states the patient died after resuscitation attempts failed. A manufacturer engineer restored the devices.
Report: 3008952681-2025-00001
Stat assist buttons were reported as non-responsive across six beds on one site, and the reports say the same type of failure was seen in a separate room earlier. The site reported these were not linked to a patient event.
Report: 3009164985-2023-00018
One patient station repeatedly set off a stat assist alarm on its own, without anyone pressing it, causing alarms to sound throughout the facility until the device was addressed. The manufacturer's support team said the station was the cause.
Report: 3009164985-2023-00013
Written by AI from the narratives of the 14 newest reports that have one, received Sep 5, 2023 to Oct 30, 2025. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 2,998in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 18registered establishments that list this product code
Watch IQA in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get IQA every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in IQA. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:IQARetrieved Nov 29, 2026, 12:51 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:IQARetrieved Nov 29, 2026, 12:51 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:IQA · count decision_dateRetrieved Nov 29, 2026, 12:51 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:IQARetrieved Nov 29, 2026, 12:51 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:IQA · count event_date_initiatedRetrieved Nov 29, 2026, 12:51 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:IQA · count res_event_numberRetrieved Nov 29, 2026, 12:51 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-1031-2026" "Z-1032-2026" "Z-0570-2025" "Z-2417-2024" "Z-2418-2024")Retrieved Nov 29, 2026, 12:51 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:IQA AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 12:51 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:IQA AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 12:51 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:IQA AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 12:51 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:IQARetrieved Nov 29, 2026, 12:51 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:IQARetrieved Nov 29, 2026, 12:51 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:890.3725Retrieved Nov 29, 2026, 12:51 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.