FDA product code
ILY: Lamp, Infrared, Therapeutic Heating (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 890.5500
- Review panel
- Physical Medicine
- Premarket route
- 510(k) exempt
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 2291 in the last 12 months
- Median review time
- 147 daysthe latest 25 decisions
- Recall records
- 201 in the last 12 months
- Adverse event reports
- 27last 12 months, 1 death
510(k) and De Novo decisions
Decisions for product code ILY per year, by decision date, 229 since 1985. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 5 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 1 |
| 2021 | 0 |
| 2022 | 2 |
| 2023 | 1 |
| 2024 | 1 |
| 2025 | 1 |
| 2026 (so far) | 1 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K253965 | CureLight Medical Diode Laser Systems (CureLight F2-A15; CureLight F2-B15; CureLight F2-A30; CureLight F2-B30; CureLight F3-AB30; CureLight F3-AB60.)Wuhan PHOMED Technology Company , Ltd. | Mar 19, 202698 days |
| K242755 | Medical Diode Laser Systems (VELAS II-30B, VELAS II-30A, GBOX-15AB, GBOX-20B, CHEESE II-10B)Gigaalaser Company , Ltd. | May 19, 2025249 days |
| K231997 | SIRONA 100 Therapy Laser System (OEM's Model EVANLAS)Cure Point Therapeutics, Inc. | Oct 15, 2024468 days |
| K223642 | LED Lip Perfector, model: ZC-05Light Tree Ventures Europe B.V. | Mar 9, 202393 days |
| K221189 | OLIZ LTB-1000ALtbio Co., Ltd. | Aug 29, 2022126 days |
| K213524 | CytonPro-5000Cytonsys, Inc. | Aug 10, 2022280 days |
| K202028 | dermalux Flex MDAesthetic Technology, Ltd. | Dec 16, 2020147 days |
| K171087 | AXON IR Heat LampCrystal Medtech, LLC | Jul 7, 201786 days |
| K163079 | HealthLight Microcontroller Models HL-FB (Footbed), HL-HB (Headband) and HL-FM (Facemask) LED Light Therapy SystemHealthlight, LLC | Jun 13, 2017222 days |
| K170068 | Softcure 96Wellray Medical Laser, Inc. | May 5, 2017116 days |
Recalls
Recall records for product code ILY per year, by the date the recall began. Each recalled product is one record; these 20 records belong to 12 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 1 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 4 |
| 2023 | 1 |
| 2024 | 0 |
| 2025 | 5 |
| 2026 (so far) | 1 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-2162-2026Class II | Technological Medical Advancements LLCSoftware update is needed for laser systems, for topical infrared heating to provide temporary relief of minor muscle and joint pain, spasms, stiffness associa… | Jan 9, 2026 |
| Z-2416-2025Class II | THOR Photomedicine LtdThe potential for a failed welded nut to shear off. When this occurs, the gas strut on one side of the device becomes detached from the canopy, making the cano… | Jul 24, 2025 |
| Z-2417-2025Class II | THOR Photomedicine LtdThe potential for a failed welded nut to shear off. When this occurs, the gas strut on one side of the device becomes detached from the canopy, making the cano… | Jul 24, 2025 |
| Z-2418-2025Class II | THOR Photomedicine LtdThe potential for a failed welded nut to shear off. When this occurs, the gas strut on one side of the device becomes detached from the canopy, making the cano… | Jul 24, 2025 |
| Z-2419-2025Class II | THOR Photomedicine LtdThe potential for a failed welded nut to shear off. When this occurs, the gas strut on one side of the device becomes detached from the canopy, making the cano… | Jul 24, 2025 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code ILY. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 12 |
| 2018 | 1 |
| 2019 | 7 |
| 2020 | 5 |
| 2021 | 5 |
| 2022 | 2 |
| 2023 | 15 |
| 2024 | 19 |
| 2025 | 17 |
| 2026 (so far) | 21 |
Last 12 months by event type
- Deaths1
- Injuries20
- Malfunctions5
- Other or not stated1
Most reported device problems, last 12 months
- Temperature Problem11
- Device Markings/Labelling Problem2
- Overheating of Device2
- Adverse Event Without Identified Device or Use Problem1
- Excessive Heating1
- Fluid/Blood Leak1
In plain English: what the newest reports describe Written by AI
Several reports describe the laser or treatment handpiece overheating or burning people. Staff reported blisters on their hands and fingers from the handpiece becoming very hot, and in some cases the handpiece reportedly heated without warning. Some of these reports do not give details on when injuries happened or whether first aid was given.
3 reports: 9616086-2026-00006, 9616086-2026-00007, 9616086-2026-00003
Reports describe burn injuries to patients after laser treatment, including a third-degree burn in one report and burn-like lesions on a patient's back in another. One report says the burn appeared hours after treatment ended. The details are brief and do not explain the cause.
3 reports: 9616086-2026-00016, 9616086-2026-00010, 9616086-2026-00004
Several reports describe burns or blistering from the device getting too hot. One patient reported burns on the temple after the mask-style wearable heated up during standard use, and a clinic diagnosed a burn. Another report describes second-degree burns on a patient's lower abdomen during treatment over a tattoo, which the clinic attributed to the patient moving.
2 reports: MW6003289, 2954354-2026-00007
One user reported a painful, itchy, bright red rash on the cheeks and eyebrows about four hours after using a light therapy device on two occasions. The rash needed prescription antihistamines and steroids and took about two weeks to clear.
2 reports: MW5186922, 3003899624-2026-00001
A user reported facial and hand volume loss, sunken temples and cheeks, and numbness above one eyebrow after many sessions of a near-infrared device. The user believed the device had shrunk fat under the skin, though no skin burning was seen.
2 reports: MW5187254, MW5181728
Written by AI from the narratives of the 20 newest reports that have one, received Jan 5, 2026 to Oct 1, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 1,548in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 628registered establishments that list this product code
Other codes under 21 CFR 890.5500
Product codes classified under the same regulation.
Watch ILY in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get ILY every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in ILY. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:ILYRetrieved Nov 28, 2026, 8:52 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:ILYRetrieved Nov 29, 2026, 4:16 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:ILY · count decision_dateRetrieved Nov 28, 2026, 8:52 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:ILYRetrieved Nov 29, 2026, 4:16 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:ILY · count event_date_initiatedRetrieved Nov 28, 2026, 8:52 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:ILY · count res_event_numberRetrieved Nov 28, 2026, 8:52 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-2162-2026" "Z-2416-2025" "Z-2417-2025" "Z-2418-2025" "Z-2419-2025")Retrieved Nov 29, 2026, 4:16 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:ILY AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 4:16 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:ILY AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 4:16 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:ILY AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 4:16 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:ILYRetrieved Nov 28, 2026, 8:52 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:ILYRetrieved Nov 28, 2026, 8:52 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:890.5500Retrieved Nov 28, 2026, 8:52 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil