FDA product code
HQO: Unit, Cautery, Thermal, Ac-Powered (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 886.4115
- Review panel
- Ophthalmic
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 120 in the last 12 months
- Median review time
- 80 daysthe latest 12 decisions
- Recall records
- 00 in the last 12 months
- Adverse event reports
- 119last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code HQO per year, by decision date, 12 since 1981. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K070871 | STARION UNIVERSAL POWER SUPPLYStarion Instruments | Apr 13, 200715 days |
| K043155 | STARION UNIVERSAL POWER SUPPLY (UPS)Starion Instruments | Dec 9, 200424 days |
| K002341 | THERMAL CAUTERY UNIT, MODEL 150Geiger Medical Technologies, Inc. | Oct 23, 200083 days |
| K000893 | AC POWERED THERMAL CAUTERY UNITStarion Instruments | Jun 14, 200086 days |
| K971538 | MENTOR GEMINI HEMOSTATIC ERASERMentor Ophthalmics, Inc. | Jul 14, 199777 days |
| K962135 | D.O.R.C. MICRODIATHERMY SYSTEMDutch Ophthalmic USA, Inc. | Aug 29, 199687 days |
| K943031 | BIPOLAR ELECTROSURGICAL PENCILBoston Surgical Products, Inc. | Sep 23, 199488 days |
| K900913 | ADLER DISPOSABLE BIPOLAR CORDAdler Instrument Co. | Apr 4, 199036 days |
| K874367 | VARIOUS CARDIAC CRYOPROBES HAVING DIA. & COS. DIFFSpembly Medical , Ltd. | Jan 4, 198873 days |
| K852075 | SITE-MALIS BIPOLAR COAGULATOR MODULEValley Forge Scientific Corp. | Aug 20, 198599 days |
Recalls
Recall records for product code HQO per year, by the date the recall began. Each recalled product is one record.
No recalls on record for this product code.
Adverse event reports (MAUDE)
Medical device reports received per year that name product code HQO. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 0 |
| 2018 | 17 |
| 2019 | 73 |
| 2020 | 44 |
| 2021 | 56 |
| 2022 | 53 |
| 2023 | 48 |
| 2024 | 66 |
| 2025 | 102 |
| 2026 (so far) | 107 |
Last 12 months by event type
- Deaths0
- Injuries1
- Malfunctions117
- Other or not stated1
Most reported device problems, last 12 months
- Electrical /Electronic Property Problem54
- Intermittent Loss of Power28
- Failure to Deliver Energy17
- Intermittent Continuity9
- Insufficient Device Problem Information8
- Crack4
In plain English: what the newest reports describe Written by AI
In multiple reports the power unit that supplies energy stopped delivering power even though its lights came on. Swapping cables did not fix it in these reports, but swapping the power unit did, and the procedures were finished with another unit.
7 reports: 2242352-2026-0000839, 2242352-2026-0000822, 2242352-2026-0000814, 2242352-2026-0000795, 2242352-2026-0000872, 2242352-2026-0000764, 2242352-2026-0000679
In two reports the device gave intermittent power, making a noise and then losing its cutting or burning ability, which the surgeons described as overheating. Swapping the cords appeared to fix the problem, and the same device completed all three procedures.
2 reports: 2242352-2026-0000682, 2242352-2026-0000683
During monopolar cautery used alongside the harvesting tool, two reports describe interference that caused loss of cautery function and difficulty cutting. The hospital's check found a power-unit-related issue, and one case was delayed by about 30 minutes.
2 reports: 2242352-2026-0000937, 2242352-2026-0000879
In two reports the device did not shut off after its 15-second activation cycle and could not be restarted by releasing and reengaging the trigger.
2 reports: 2242352-2026-0000839, 2242352-2026-0000879
Two reports describe cracks in the device's case, and one of them says the cable port was worn and no longer held the adapter tightly. In one report the device also stopped working intermittently, and the cracks were visible in photos the hospital provided.
2 reports: 2242352-2026-0000894, 2242352-2026-0000785
Written by AI from the narratives of the 17 newest reports that have one, received Aug 12, 2026 to Oct 30, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 34in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 2registered establishments that list this product code
Other codes under 21 CFR 886.4115
Product codes classified under the same regulation.
Watch HQO in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get HQO every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in HQO. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:HQORetrieved Nov 29, 2026, 6:47 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:HQORetrieved Nov 29, 2026, 6:47 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:HQO · count decision_dateRetrieved Nov 29, 2026, 6:47 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:HQORetrieved Nov 29, 2026, 6:47 PM UTC.
- device/recall · search product_code:HQO · count event_date_initiatedRetrieved Nov 29, 2026, 6:47 PM UTC.
- device/recall · search product_code:HQO · count res_event_numberRetrieved Nov 29, 2026, 6:47 PM UTC.
- device/event · search device.device_report_product_code:HQO AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 6:47 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:HQO AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 6:47 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:HQO AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 6:47 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:HQORetrieved Nov 29, 2026, 6:47 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:HQORetrieved Nov 29, 2026, 6:47 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:886.4115Retrieved Nov 29, 2026, 6:47 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil