Product codes

FDA product code

HPP: Headlamp, Operating, Battery-Operated (FDA product code)

Follow HPP: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class I
Regulation
21 CFR 886.4335
Review panel
Ophthalmic
Premarket route
510(k) exempt
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Not eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
00 in the last 12 months
Median review time
Not availablethe latest 0 decisions
Recall records
10 in the last 12 months
Adverse event reports
2last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code HPP per year, by decision date. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

No decisions on record for this product code.

Recalls

Recall records for product code HPP per year, by the date the recall began. Each recalled product is one record; these 1 records belong to 1 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20170
20180
20190
20200
20210
20220
20230
20240
20250
2026 (so far)0

Latest recalls

Recall and classFirm and reasonBegan
Z-1976-2014Class II Integra LifeSciences Corp.LED battery chargers may prematurely fail and will not charge the LED battery as intended Jun 11, 2014

Adverse event reports (MAUDE)

Medical device reports received per year that name product code HPP. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
20170
20180
20191
20202
20212
20221
20232
20241
20252
2026 (so far)1

Last 12 months by event type

  • Deaths0
  • Injuries1
  • Malfunctions1
  • Other or not stated0

Most reported device problems, last 12 months

  1. Overheating of Device1
  2. Smoking1

In plain English: what the newest reports describe Written by AI

  • In four reports, skin was burned during surgery with a headlamp, causing redness and blisters. Testing of some lamps found no fault, but one hospital measured significant heating.

    4 reports: MW5176076, MW5144689, 11116235, 3013682780-2023-00001

  • In three reports, a surgical headlamp was pointed at a surgeon's gown or scrubs, and the fabric began to smoke or steam. In one report the gown had no burn hole.

    3 reports: MW5190651, MW5180139, 12664630

  • In two reports, headlamp batteries overheated. One says two batteries swelled and cracked their housing during use. The other says a battery smoked while charging inside its carrying case, which the manufacturer says is not designed for charging.

    2 reports: 3006981798-2022-00033, 0002127132-2024-00001

  • In two reports, the headlamp cord was damaged, either where it joins the lamp head and headstrap or where a bezel cable tore. A piece fell into the surgical area, but in one report it landed on a sheet rather than the patient.

    2 reports: 3006981798-2020-00007, 3006981798-2020-00002

  • In one report, a headlamp was plugged into its charger, the charger caught fire, and the user threw it away. The headlamp stopped charging afterward.

    Report: 3006981798-2019-00006

Written by AI from the narratives of the 12 newest reports that have one, received May 24, 2019 to Jul 13, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
370in the FDA’s Global Unique Device Identification Database
Establishment listings
76registered establishments that list this product code

Product codes classified under the same regulation.

Watch HPP in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get HPP every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in HPP. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.