Product codes

FDA product code

HJO: Biomicroscope, Slit-Lamp, Ac-Powered (FDA product code)

Follow HJO: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 886.1850
Review panel
Ophthalmic
Premarket route
510(k)
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
1092 in the last 12 months
Median review time
194 daysthe latest 25 decisions
Recall records
80 in the last 12 months
Adverse event reports
14last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code HJO per year, by decision date, 109 since 1980. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20179
20182
20190
20201
20210
20221
20230
20241
20250
2026 (so far)2

Latest decisions

NumberDevice and applicantDecision and review time
K262637 SLIT LAMP SL-D4 (Type: LED)Topcon Corporation Sep 18, 202651 days
K253834 VISUREF 600Carl Zeiss Vision GmbH Mar 18, 2026107 days
K231760 HP-OCT (HP-OCT)Cylite Pty. , Ltd. Mar 8, 2024266 days
K222372 Kowa SL-19Kowa Company , Ltd. Nov 21, 2022108 days
K193188 MiiS Horus Eye Anterior CameraMedimaging Integrated Solution, Inc (Miis) Jan 14, 202057 days
K182306 Sunkingdom Slit LampChongqing Sunkingdom Medical Instrument Co., Ltd. Nov 22, 201890 days
K173771 IOLMaster 700Carl Zeiss Meditec, AG Aug 24, 2018256 days
K171877 SLM-1ER, SLM-2ER, SLM-3ERChongqing Kanghua Ruiming S&T Co., Ltd. Dec 21, 2017181 days
K170470 Digital Eye Anterior CameraMedimaging Integrated Solution, Inc (Miis) Oct 6, 2017232 days
K162778 Portable Slit Lamp MicroscopeSuzhou Kangjie Medical, Inc. Jul 21, 2017294 days

Recalls

Recall records for product code HJO per year, by the date the recall began. Each recalled product is one record; these 8 records belong to 6 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20170
20181
20190
20201
20211
20220
20230
20240
20250
2026 (so far)0

Latest recalls

Recall and classFirm and reasonBegan
Z-1560-2021Class II Haag-Streit USA IncExamination data and patient name may be mixed up when printing or generating a pdf with the interface at Pacific Coast and Laser Institute (PCLI). Apr 15, 2021
Z-1133-2021Class II Carl Zeiss Meditec AGWhen using software 1.90.2.09 or 1.90.8.06 and using modality worklist functionality for patient data transfer, the selection of the patient in the patient lis… Dec 7, 2020
Z-0108-2019Class II Carl Zeiss Meditec AGCalibration setting was incorrectly entered during manufacturing causing the optical coherence tomography (OCT) to obtain the incorrect two-dimensional images… Jun 25, 2018
Z-2587-2016Class II Nidek IncNidek Inc. received information from our manufacturer NIDEK CO. Japan that the shown axial lengths were much shorter at AL-Scan measurement. Jul 1, 2016
Z-0357-2016Class II Carl Zeiss Meditec AGIOL Master software versions 7.5 and 7.7 calculation printouts and exported reports can contain the wrong IOL power data. Oct 27, 2015

Adverse event reports (MAUDE)

Medical device reports received per year that name product code HJO. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
201715
201811
201913
202019
202114
20226
202318
202418
202520
2026 (so far)6

Last 12 months by event type

  • Deaths0
  • Injuries10
  • Malfunctions4
  • Other or not stated0

Most reported device problems, last 12 months

  1. Incorrect, Inadequate or Imprecise Result or Readings9
  2. Misassembly During Maintenance/Repair9
  3. Adverse Event Without Identified Device or Use Problem2
  4. Insufficient Device Problem Information1
  5. Output Problem1

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
413in the FDA’s Global Unique Device Identification Database
Establishment listings
121registered establishments that list this product code

Product codes classified under the same regulation.

Watch HJO in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get HJO every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in HJO. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.