FDA product code
HIF: Insufflator, Laparoscopic (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 884.1730
- Review panel
- Obstetrics and Gynecology
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Ineligible
- 510(k) and De Novo decisions
- 1793 in the last 12 months
- Median review time
- 134 daysthe latest 25 decisions
- Recall records
- 2915 in the last 12 months
- Adverse event reports
- 1,669last 12 months, 2 deaths
510(k) and De Novo decisions
Decisions for product code HIF per year, by decision date, 179 since 1978. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 5 |
| 2018 | 3 |
| 2019 | 2 |
| 2020 | 4 |
| 2021 | 1 |
| 2022 | 1 |
| 2023 | 7 |
| 2024 | 4 |
| 2025 | 3 |
| 2026 (so far) | 3 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K260209 | ENDOFLATOR+; Smart Smoke EvacuationKarl Storz SE & CO. KG | Aug 19, 2026208 days |
| K252584 | Insufflator (wAlcor C50A, wAlcor C50B, wAlcor C50C, wAlcor C35A, wAlcor C35B, wAlcor C35C)Weyo Surgical Technology , Ltd. | Jun 25, 2026314 days |
| K252318 | Insufflator (OPTO-IFL1000)Guangdong Optomedic Technologies, Inc. | Apr 16, 2026265 days |
| K250388 | ENDOFLATOR +Karl Storz SE & CO. KG | Sep 22, 2025223 days |
| K242772 | AirSurgN Insufflator (10030/AirSurgN)SmartsurgN, Inc. | Apr 11, 2025210 days |
| K243527 | High Flow Insufflation Unit (UHI-4)Olympus Medical Systems Corporation | Jan 13, 202560 days |
| K242536 | TauTona Pneumoperitoneum Assist Device (TPAD)Tautona Group | Oct 15, 202450 days |
| K232902 | EVA15 insufflatorPalliare , Ltd. | May 2, 2024227 days |
| K232464 | VereSee Optical Veres Needle SystemFreyja Healthcare, LLC | May 2, 2024261 days |
| K233020 | TauTona Pneumoperitoneum Assist Device (TPAD)Tautona Group Research & Development, Co., LLC | Jan 12, 2024112 days |
Recalls
Recall records for product code HIF per year, by the date the recall began. Each recalled product is one record; these 29 records belong to 15 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 2 |
| 2019 | 3 |
| 2020 | 1 |
| 2021 | 1 |
| 2022 | 0 |
| 2023 | 2 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 15 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-2776-2026Class II | Olympus Corporation of the AmericasPotential for unexpected instances where the front panel display turns off unexpectedly, and air delivery (insufflation) ceases. | Jun 16, 2026 |
| Z-2464-2026Class II | ConMed CorporationPotential for overpressure alerts. | May 14, 2026 |
| Z-2465-2026Class II | ConMed CorporationPotential for overpressure alerts. | May 14, 2026 |
| Z-2466-2026Class II | ConMed CorporationPotential for overpressure alerts. | May 14, 2026 |
| Z-2467-2026Class II | ConMed CorporationPotential for overpressure alerts. | May 14, 2026 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code HIF. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 68 |
| 2018 | 97 |
| 2019 | 128 |
| 2020 | 360 |
| 2021 | 654 |
| 2022 | 667 |
| 2023 | 890 |
| 2024 | 1,483 |
| 2025 | 1,519 |
| 2026 (so far) | 1,550 |
Last 12 months by event type
- Deaths2
- Injuries62
- Malfunctions1,603
- Other or not stated2
Most reported device problems, last 12 months
- Pressure Problem287
- Gas/Air Leak189
- Unexpected Shutdown100
- No Display/Image90
- Circuit Failure89
- Display or Visual Feedback Problem79
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 365in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 326registered establishments that list this product code
Other codes under 21 CFR 884.1730
Product codes classified under the same regulation.
Watch HIF in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get HIF every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in HIF. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:HIFRetrieved Nov 28, 2026, 10:21 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:HIFRetrieved Nov 29, 2026, 4:15 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:HIF · count decision_dateRetrieved Nov 28, 2026, 10:21 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:HIFRetrieved Nov 29, 2026, 4:15 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:HIF · count event_date_initiatedRetrieved Nov 28, 2026, 10:21 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:HIF · count res_event_numberRetrieved Nov 28, 2026, 10:21 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-2776-2026" "Z-2464-2026" "Z-2465-2026" "Z-2466-2026" "Z-2467-2026")Retrieved Nov 29, 2026, 4:15 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:HIF AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 4:15 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:HIF AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 4:15 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:HIF AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 4:15 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:HIFRetrieved Nov 28, 2026, 10:21 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:HIFRetrieved Nov 28, 2026, 10:21 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:884.1730Retrieved Nov 28, 2026, 10:21 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil