FDA product code
HGP: Electrode, Circular (Spiral), Scalp And Applicator (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 884.2675
- Review panel
- Obstetrics and Gynecology
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Eligible
- Summary malfunction reporting
- Ineligible
- 510(k) and De Novo decisions
- 170 in the last 12 months
- Median review time
- 91 daysthe latest 17 decisions
- Recall records
- 30 in the last 12 months
- Adverse event reports
- 2last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code HGP per year, by decision date, 17 since 1977. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 1 |
| 2026 (so far) | 0 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K243291 | Goldtrace Fetal Spiral Electrode (FSE) (CNS000004)Neoventa Medical AB | Jan 17, 202591 days |
| K030691 | FETAL SPIRAL ELECTRODE, MODEL FSE07000Clinical Innovations, Inc. | Sep 17, 2003196 days |
| K003458 | SOFT BEAT FETAL HEART RATE ELECTRODE MODEL SBT-7000Clinical Innovations, Inc. | May 3, 2001177 days |
| K943732 | LIFE TRACE FETAL SPIRAL ELECTRODE WITH MATERNAL REFERENCEGraphic Controls Corp. | Oct 20, 199479 days |
| K910815 | CR SPIRAL ELECTRODENeward Enterprises, Inc. | Dec 10, 1991287 days |
| K911657 | FETAL SCALP SPIRAL ELECTRODEAdvanced Medical Systems, Inc. | Jul 11, 199190 days |
| K905830 | COMBO FETAL SCALP ELECTRODE/LEGPLATEUtah Medical Products, Inc. | Mar 29, 199188 days |
| K904745 | MEDI-TRACE(R) FETAL MONITORING SPIRAL ELECTRODEGraphic Controls Corp. | Jan 29, 1991103 days |
| K905130 | FETAL MONITORING SPIRAL ELECTRODEGraphic Controls Corp. | Jan 29, 199175 days |
| K901152 | REDYPLATE DISPOSABLE LEG PLATEMedical Accessories, Inc. | Apr 4, 199022 days |
Recalls
Recall records for product code HGP per year, by the date the recall began. Each recalled product is one record; these 3 records belong to 3 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 1 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 1 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-0908-2023Class II | Philips North America LlcFetal Spiral Electrode (FSE) spiral tip of the FSE breaking off during use and requiring surgical intervention to remove the broken tip from the neonate patien… | Nov 21, 2022 |
| Z-1250-2019Class II | Philips North America, LLCDuring use of the Philips FSE, it is possible for the metal electrode tip to break off and remain in the scalp of the newborn following labor and delivery. | Jan 25, 2019 |
| Z-0330-2013Class II | Vital Signs Devices, a GE Healthcare CompanyThirty fetal scalp electrodes were found to have an incomplete package seal and lacked the printed lot number stamped on during sealing. | Oct 16, 2012 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code HGP. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 17 |
| 2018 | 44 |
| 2019 | 26 |
| 2020 | 25 |
| 2021 | 33 |
| 2022 | 27 |
| 2023 | 8 |
| 2024 | 6 |
| 2025 | 0 |
| 2026 (so far) | 2 |
Last 12 months by event type
- Deaths0
- Injuries1
- Malfunctions1
- Other or not stated0
Most reported device problems, last 12 months
- Mechanical Problem1
- Use of Device Problem1
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 32in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 19registered establishments that list this product code
Watch HGP in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get HGP every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in HGP. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:HGPRetrieved Nov 29, 2026, 5:03 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:HGPRetrieved Nov 29, 2026, 5:03 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:HGP · count decision_dateRetrieved Nov 29, 2026, 5:03 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:HGPRetrieved Nov 29, 2026, 5:03 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:HGP · count event_date_initiatedRetrieved Nov 29, 2026, 5:03 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:HGP · count res_event_numberRetrieved Nov 29, 2026, 5:03 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-0908-2023" "Z-1250-2019" "Z-0330-2013")Retrieved Nov 29, 2026, 5:03 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:HGP AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 5:03 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:HGP AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 5:03 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:HGP AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 5:03 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:HGPRetrieved Nov 29, 2026, 5:03 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:HGPRetrieved Nov 29, 2026, 5:03 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:884.2675Retrieved Nov 29, 2026, 5:03 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.