FDA product code
HFN: Transducer, Pressure, Intrauterine (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 884.2700
- Review panel
- Obstetrics and Gynecology
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Eligible
- Summary malfunction reporting
- Ineligible
- 510(k) and De Novo decisions
- 90 in the last 12 months
- Median review time
- 89 daysthe latest 9 decisions
- Recall records
- 10 in the last 12 months
- Adverse event reports
- 6last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code HFN per year, by decision date, 9 since 1989. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 1 |
| 2026 (so far) | 0 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K250897 | Koala Intrauterine Pressure Catheter (IPC-5000E)Clinical Innovations, LLC | May 22, 202558 days |
| K974389 | KOALA INTRAUTERINE PRESSURE CATHETER, KOALA EXTERNAL INTRAUTERINE PRESSURE CATHETERClinical Innovations, Inc. | Feb 19, 199890 days |
| K973746 | PRE-FILLED SYRINGE FOR KOALA IUPClinical Innovations, Inc. | Nov 24, 199754 days |
| K961424 | INTRAN 500Utah Medical Products, Inc. | Jul 10, 199689 days |
| K954955 | STORC INTRAUTERINE PRESSURE CATHETERClinical Innovations, Inc. | Apr 17, 1996173 days |
| K955443 | INTRAN PLUSUtah Medical Products, Inc. | Feb 20, 199684 days |
| K920375 | DISPOSABLE INTRAUTERINE PRESSURE TRANSDUCERTelos Medical Corp. | Oct 27, 1992272 days |
| K882224 | ULTRANS DISPOSABLE INTRAUTERINETelos Medical Corp. | Oct 12, 1989504 days |
| K893503 | SOFSENS(TM) MODEL LMP-100 INTRAUTERINE PRES. TRANSLaurel Medical Corp. | Jun 12, 198938 days |
Recalls
Recall records for product code HFN per year, by the date the recall began. Each recalled product is one record; these 1 records belong to 1 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-2200-2008 | Utah Medical Products, IncPossibility of compromised package sterility prior to the expiration date. | Jun 16, 2008 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code HFN. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 2 |
| 2021 | 3 |
| 2022 | 0 |
| 2023 | 1 |
| 2024 | 4 |
| 2025 | 0 |
| 2026 (so far) | 6 |
Last 12 months by event type
- Deaths0
- Injuries1
- Malfunctions5
- Other or not stated0
Most reported device problems, last 12 months
- Patient Device Interaction Problem1
In plain English: what the newest reports describe Written by AI
Several reports say the device did not give correct readings of contractions or uterine pressure. Staff changed cables, and in some cases replacing the catheter itself fixed the problem.
5 reports: MW5189988, MW5189594, MW5115662, 9995513, 2944237
Several reports say the catheter was hard to push into the uterus, and one says it felt stiffer than usual. In one case, the physician noted it was almost stuck in the introducer and needed more manipulation, and the patient later had a cesarean section.
4 reports: 24469546, 24469500, 24267935, MW5189594
Several reports describe pressure injuries or bruising on babies' faces or heads linked to the catheter's position. One facility said scraping injuries fell after staff were retrained on insertion, but it still saw pressure injuries from the catheter's position.
4 reports: 1722684-2026-00017, 1722684-2021-00004, 1722684-2021-00001, 1722684-2020-00001
In three reports, the baby's heart rate dropped after the catheter was placed, and an emergency cesarean section followed each time. One report describes bright red blood during insertion, and another describes a possible cut in a placental blood vessel. The babies died in two of these reports.
3 reports: MW5027361, 2224163, 2944237
Two reports describe the same event: a uterine rupture was found about 90 minutes after the catheter was used to add fluid into the uterus, a procedure called amnioinfusion. The patient needed an urgent cesarean section.
2 reports: 1722684-2024-00008, MW5152782
Written by AI from the narratives of the 19 newest reports that have one, received Aug 18, 2011 to Aug 20, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 26in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 15registered establishments that list this product code
Other codes under 21 CFR 884.2700
Product codes classified under the same regulation.
Watch HFN in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get HFN every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in HFN. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:HFNRetrieved Nov 29, 2026, 7:02 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:HFNRetrieved Nov 29, 2026, 7:02 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:HFN · count decision_dateRetrieved Nov 29, 2026, 7:02 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:HFNRetrieved Nov 29, 2026, 7:02 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:HFN · count event_date_initiatedRetrieved Nov 29, 2026, 7:02 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:HFN · count res_event_numberRetrieved Nov 29, 2026, 7:02 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-2200-2008")Retrieved Nov 29, 2026, 7:02 PM UTC.
- device/event · search device.device_report_product_code:HFN AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 7:02 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:HFN AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 7:02 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:HFN AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 7:02 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:HFNRetrieved Nov 29, 2026, 7:02 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:HFNRetrieved Nov 29, 2026, 7:02 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:884.2700Retrieved Nov 29, 2026, 5:01 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil