FDA product code
HEX: Colposcope (And Colpomicroscope) (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 884.1630
- Review panel
- Obstetrics and Gynecology
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 560 in the last 12 months
- Median review time
- 145 daysthe latest 25 decisions
- Recall records
- 00 in the last 12 months
- Adverse event reports
- 1last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code HEX per year, by decision date, 56 since 1976. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 1 |
| 2018 | 2 |
| 2019 | 1 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 1 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K233043 | Smart Scope® CXPeriwinkle Technologies Pvt., Ltd. | Jun 14, 2024263 days |
| K190187 | Gynescope SystemIllumigyn , Ltd. | May 8, 201996 days |
| K182764 | Biop Digital ColposcopeBiop Medical , Ltd. | Nov 27, 201860 days |
| K181034 | Pocket Colposcope SystemHadleigh Health Technologies, LLC | Sep 21, 2018156 days |
| K170446 | LT-300 SD Digital Video ColposcopeLutech Industries, Inc. | May 21, 201796 days |
| K161871 | EVA (Enhanced Visual Assessment) SystemMobileodt , Ltd. | Dec 19, 2016165 days |
| K160341 | Colposcope SystemXuzhou Kernel Medical Equipment Co., Ltd. | Sep 14, 2016219 days |
| K160380 | LT-300 HDLutech Industries, Inc. | May 13, 201693 days |
| K151878 | Video ColposcopeEdan Instruments, Inc. | Dec 1, 2015145 days |
| K143119 | LT-300 Video ColposcopeLutech Industries, Inc. | Dec 18, 201449 days |
Recalls
Recall records for product code HEX per year, by the date the recall began. Each recalled product is one record.
No recalls on record for this product code.
Adverse event reports (MAUDE)
Medical device reports received per year that name product code HEX. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 2 |
| 2020 | 1 |
| 2021 | 1 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 1 |
| 2025 | 1 |
| 2026 (so far) | 1 |
Last 12 months by event type
- Deaths0
- Injuries0
- Malfunctions1
- Other or not stated0
Most reported device problems, last 12 months
- Break1
- Device Dislodged or Dislocated1
- Unstable1
In plain English: what the newest reports describe Written by AI
Parts have come loose from the arm that supports the device (the balance arm), and in some cases parts broke or fell. One report describes a falling part hurting a nurse's knee and smashing the lens, and the procedure was briefly stopped.
3 reports: 3002808148-2024-00153, 8010047-2021-08344, 8010047-2019-01092
Pieces of a uterine manipulator, a device placed in the uterus during hysterectomy, were left in the vagina after surgery. Patients showed vaginal discharge or a foreign object, and the pieces were removed in clinic, under sedation, or under spinal anesthesia.
3 reports: 1216677-2011-00010, 1214248, MW5037350
One report describes a red burn on a user's neck where a necklace she wore was touching the device during colposcopy, which used a separate electrosurgical unit. A later safety test of the light source and camera found no electrical leakage.
Report: 8010047-2019-04243
One report describes an empty examination bed hitting the device while being lifted, breaking its adjustment handle and causing an attached monitor to fall to the floor. The monitor had outside damage but still displayed properly, and no patient injury was reported.
Report: 8010047-2020-07703
Written by AI from the narratives of the 12 newest reports that have one, received Apr 3, 2002 to Nov 4, 2025. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 158in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 23registered establishments that list this product code
Other codes under 21 CFR 884.1630
Product codes classified under the same regulation.
Watch HEX in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get HEX every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in HEX. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:HEXRetrieved Nov 29, 2026, 5:05 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:HEXRetrieved Nov 29, 2026, 5:05 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:HEX · count decision_dateRetrieved Nov 29, 2026, 5:05 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:HEXRetrieved Nov 29, 2026, 5:05 PM UTC.
- device/recall · search product_code:HEX · count event_date_initiatedRetrieved Nov 29, 2026, 5:05 PM UTC.
- device/recall · search product_code:HEX · count res_event_numberRetrieved Nov 29, 2026, 5:05 PM UTC.
- device/event · search device.device_report_product_code:HEX AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 5:05 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:HEX AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 5:05 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:HEX AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 5:05 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:HEXRetrieved Nov 29, 2026, 5:05 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:HEXRetrieved Nov 29, 2026, 5:05 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:884.1630Retrieved Nov 29, 2026, 5:05 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.