FDA product code
HBE: Drills, Burrs, Trephines & Accessories (Simple, Powered) (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 882.4310
- Review panel
- Neurology
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 651 in the last 12 months
- Median review time
- 127 daysthe latest 25 decisions
- Recall records
- 25819 in the last 12 months
- Adverse event reports
- 7,820last 12 months, 1 death
510(k) and De Novo decisions
Decisions for product code HBE per year, by decision date, 65 since 1987. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 2 |
| 2018 | 2 |
| 2019 | 5 |
| 2020 | 0 |
| 2021 | 6 |
| 2022 | 2 |
| 2023 | 2 |
| 2024 | 0 |
| 2025 | 6 |
| 2026 (so far) | 1 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K253627 | Surgify Halo (54.085.SHD.U1, 54.140.SHD. U1, 54.070.NVG.U1, 54.125.NVG.U1, 54.000.SEE.U1, 40.070.NVG.U1, 40.125.NVG.U1, 40.000.SEE.U1, 30.070.NVG.U2, 30.125.NVG.U2, 30.000.SEE.U2)Surgify Medical OY | Mar 26, 2026127 days |
| K250815 | Hubly Drill (H100)Hubly, Inc. | Nov 6, 2025234 days |
| K252060 | GREENPhasor Health, LLC | Oct 31, 2025122 days |
| K252662 | UniBurNakanishi, Inc. | Sep 19, 202528 days |
| K251433 | Surgify Halo (54.085.SHD.H1, 54.140.SHD.H1, 54.070.NVG.H1, 54.125.NVG.H1, 54.000.SEE.H1, 40.070.NVG.H1, 40.125.NVG.H1, 40.000.SEE.H1)Surgify Medical OY | Jun 4, 202527 days |
| K250380 | Surgify Halo (54.085.SHD.H1, 54.140.SHD.H1, 54.070.NVG.H1, 54.125.NVG.H1, 54.000.SEE.H1)Surgify Medical OY | Mar 13, 202530 days |
| K250374 | iBur 3.0mm Diamond Match Head, Distal Bend (8431-107-030D); iBur 3.0mm Precision Round, Distal Bend (8431-009-030); iBur 4.0mm Precision Round, Distal Bend (8431-009-040); iBur 2.0mm Diamond Round, Distal Bend (8431-012-020D); iBur 3.0mm Diamond Round, Distal Bend (8431-012-030D); iBur 4.0mm Diamond Round, Distal Bend (8431-012-040D); iBur 3.0mm Coarse Diamond Round, Distal Bend (8431-013-030DC); iBur 4.0mm Coarse Diamond Round, Distal Bend (8431-013-040DC); iBur 5.0mm Coarse Diamond RStryker Instruments | Mar 11, 202529 days |
| K232684 | Surgify HaloSurgify Medical OY | Nov 29, 202389 days |
| K230619 | Hubly Electric Drill (H100)Hubly, Inc. | May 4, 202359 days |
| K223199 | 2.1 x 255mm Drill, 22mm StopBiomet Microfixation | Nov 15, 202233 days |
Recalls
Recall records for product code HBE per year, by the date the recall began. Each recalled product is one record; these 258 records belong to 34 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 3 |
| 2018 | 1 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 1 |
| 2022 | 1 |
| 2023 | 3 |
| 2024 | 11 |
| 2025 | 2 |
| 2026 (so far) | 19 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-2555-2026Class II | SURGIFY MEDICAL OYPotential for burr breakage, occurring during bi-portal endoscopic spinal surgery (BESS). | Apr 30, 2026 |
| Z-2556-2026Class II | SURGIFY MEDICAL OYPotential for burr breakage, occurring during bi-portal endoscopic spinal surgery (BESS). | Apr 30, 2026 |
| Z-2557-2026Class II | SURGIFY MEDICAL OYPotential for burr breakage, occurring during bi-portal endoscopic spinal surgery (BESS). | Apr 30, 2026 |
| Z-2558-2026Class II | SURGIFY MEDICAL OYPotential for burr breakage, occurring during bi-portal endoscopic spinal surgery (BESS). | Apr 30, 2026 |
| Z-2559-2026Class II | SURGIFY MEDICAL OYPotential for burr breakage, occurring during bi-portal endoscopic spinal surgery (BESS). | Apr 30, 2026 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code HBE. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 848 |
| 2018 | 787 |
| 2019 | 836 |
| 2020 | 1,025 |
| 2021 | 1,047 |
| 2022 | 900 |
| 2023 | 948 |
| 2024 | 1,153 |
| 2025 | 1,943 |
| 2026 (so far) | 7,744 |
Last 12 months by event type
- Deaths1
- Injuries46
- Malfunctions7,768
- Other or not stated5
Most reported device problems, last 12 months
- No Apparent Adverse Event2,179
- Detachment of Device or Device Component1,316
- Naturally Worn1,153
- Overheating of Device998
- Difficult to Insert810
- Loss of Power344
In plain English: what the newest reports describe Written by AI
Bearings inside the attachment (the rotating parts that let it spin smoothly) were found detached or broken when the manufacturer evaluated returned units. This appears in many of the reports and is the most common finding described.
10 reports: 1625507-2026-08979, 1625507-2026-08981, 1625507-2026-08982, 1625507-2026-08983, 1625507-2026-08985, 1625507-2026-08987, 1625507-2026-08989, 1625507-2026-08994, 1625507-2026-08995, 1625507-2026-09001
Reports describe the attachment's rotating part not locking, not turning freely, running rough, or malfunctioning in rotation, and the device was reported to not work or to fail before the procedure.
5 reports: 1625507-2026-08978, 1625507-2026-08994, 1625507-2026-08995, 1625507-2026-09000, 1625507-2026-08985
Reports also mention an air leak near the foot switch connection, a variable collar that does not adjust freely with corrosion, and faded laser markings. Each is described in one report only, and no patient impact was reported.
3 reports: 1625507-2026-08992, 1625507-2026-08993, 1625507-2026-08983
The attachment was reported to overheat during use, and in one report the foot pedal was said to lose contact. The reports do not state a cause, and one report links the overheating to a misaligned bearing.
2 reports: 1625507-2026-08981, 1625507-2026-08988
A report describes the tip bearing breaking. Another describes worn bearings present before the attachment was used. Both reports state no patient impact.
2 reports: 1625507-2026-08980, 1625507-2026-08990
Written by AI from the narratives of the 20 newest reports that have one, received Oct 30, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 6,781in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 383registered establishments that list this product code
Other codes under 21 CFR 882.4310
Product codes classified under the same regulation.
Watch HBE in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get HBE every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in HBE. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:HBERetrieved Nov 28, 2026, 7:10 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:HBERetrieved Nov 29, 2026, 4:16 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:HBE · count decision_dateRetrieved Nov 28, 2026, 7:10 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:HBERetrieved Nov 29, 2026, 4:16 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:HBE · count event_date_initiatedRetrieved Nov 28, 2026, 7:10 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:HBE · count res_event_numberRetrieved Nov 28, 2026, 7:10 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-2555-2026" "Z-2556-2026" "Z-2557-2026" "Z-2558-2026" "Z-2559-2026")Retrieved Nov 29, 2026, 4:16 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:HBE AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 4:16 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:HBE AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 4:16 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:HBE AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 4:16 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:HBERetrieved Nov 28, 2026, 7:10 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:HBERetrieved Nov 28, 2026, 7:10 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:882.4310Retrieved Nov 28, 2026, 7:10 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil