Product codes

FDA product code

GKN: Timer, Clot, Automated (FDA product code)

Follow GKN: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 864.5400
Review panel
Hematology
Premarket route
510(k)
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
30 in the last 12 months
Median review time
232 daysthe latest 3 decisions
Recall records
00 in the last 12 months
Adverse event reports
47last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code GKN per year, by decision date, 3 since 1992. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20170
20180
20190
20200
20210
20220
20230
20240
20250
2026 (so far)0

Latest decisions

NumberDevice and applicantDecision and review time
K953870 ACTTRAC ELECTRONIC CONTROL DEVICE FOR THE ACT & ACT II COAGULATION INSTRUMENTSMedtronic Vascular Oct 31, 199575 days
K940426 MEDTRONIC HEMOTEC ACT II AUTOMATED COAGULATION TIMERMedtronic Hemotec, Inc. Sep 20, 1994232 days
K913320 MIL CM-II COAGULATION MACHINEMedical Innovations Corp. Jun 16, 1992327 days

Recalls

Recall records for product code GKN per year, by the date the recall began. Each recalled product is one record.

All of them in the FDA’s recall database

No recalls on record for this product code.

Adverse event reports (MAUDE)

Medical device reports received per year that name product code GKN. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
20172
20180
20191
20201
202112
202254
2023120
202454
202551
2026 (so far)44

Last 12 months by event type

  • Deaths0
  • Injuries0
  • Malfunctions47
  • Other or not stated0

Most reported device problems, last 12 months

  1. Electrical /Electronic Property Problem47
  2. Device Alarm System1
  3. Temperature Problem1

In plain English: what the newest reports describe Written by AI

  • Several reports describe the device giving inconsistent or inaccurate test readings, including readings that seemed wrong for the patient's clinical picture. In one case, a second device gave very different results on the same patient's sample. The manufacturer did not state a patient effect in these reports.

    9 reports: 2184009-2026-01400, 2184009-2026-01279, 2184009-2026-01322, 2184009-2026-01384, 2184009-2026-00611, 2184009-2026-00551, 2184009-2026-00502, 2184009-2026-01399, 2184009-2026-00607

  • Several reports describe the device failing its liquid quality control check, which is a test run with a known sample to confirm accuracy. In some cases the check had to be repeated several times before it passed, and in one case the electronic check passed while the liquid check failed.

    8 reports: 2184009-2026-00614, 2184009-2026-00532, 2184009-2026-00475, 2184009-2026-00767, 2184009-2026-00769, 2184009-2026-00770, 2184009-2026-00578, 2184009-2026-00595

  • One report describes the device giving correct readings but being reported as not functioning, and another describes it failing temperature checks below its set point. The manufacturer did not state a patient effect in these reports.

    2 reports: 2184009-2026-00768, 2184009-2026-00769

  • One report describes a channel, which is one of the device's separate testing paths, freezing a few seconds after start-up. The same report says the device had given inconsistent values on both channels for about a month, despite repeated internal cleaning.

    Report: 2184009-2026-01400

  • A report describes intermittent channel failures along with a loud fan noise from the device. The manufacturer did not state a patient effect.

    Report: 2184009-2026-00532

Written by AI from the narratives of the 20 newest reports that have one, received Apr 17, 2026 to Sep 23, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
11in the FDA’s Global Unique Device Identification Database
Establishment listings
5registered establishments that list this product code

Product codes classified under the same regulation.

Watch GKN in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get GKN every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in GKN. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.