FDA product code
GKN: Timer, Clot, Automated (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 864.5400
- Review panel
- Hematology
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 30 in the last 12 months
- Median review time
- 232 daysthe latest 3 decisions
- Recall records
- 00 in the last 12 months
- Adverse event reports
- 47last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code GKN per year, by decision date, 3 since 1992. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K953870 | ACTTRAC ELECTRONIC CONTROL DEVICE FOR THE ACT & ACT II COAGULATION INSTRUMENTSMedtronic Vascular | Oct 31, 199575 days |
| K940426 | MEDTRONIC HEMOTEC ACT II AUTOMATED COAGULATION TIMERMedtronic Hemotec, Inc. | Sep 20, 1994232 days |
| K913320 | MIL CM-II COAGULATION MACHINEMedical Innovations Corp. | Jun 16, 1992327 days |
Recalls
Recall records for product code GKN per year, by the date the recall began. Each recalled product is one record.
No recalls on record for this product code.
Adverse event reports (MAUDE)
Medical device reports received per year that name product code GKN. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 2 |
| 2018 | 0 |
| 2019 | 1 |
| 2020 | 1 |
| 2021 | 12 |
| 2022 | 54 |
| 2023 | 120 |
| 2024 | 54 |
| 2025 | 51 |
| 2026 (so far) | 44 |
Last 12 months by event type
- Deaths0
- Injuries0
- Malfunctions47
- Other or not stated0
Most reported device problems, last 12 months
- Electrical /Electronic Property Problem47
- Device Alarm System1
- Temperature Problem1
In plain English: what the newest reports describe Written by AI
Several reports describe the device giving inconsistent or inaccurate test readings, including readings that seemed wrong for the patient's clinical picture. In one case, a second device gave very different results on the same patient's sample. The manufacturer did not state a patient effect in these reports.
9 reports: 2184009-2026-01400, 2184009-2026-01279, 2184009-2026-01322, 2184009-2026-01384, 2184009-2026-00611, 2184009-2026-00551, 2184009-2026-00502, 2184009-2026-01399, 2184009-2026-00607
Several reports describe the device failing its liquid quality control check, which is a test run with a known sample to confirm accuracy. In some cases the check had to be repeated several times before it passed, and in one case the electronic check passed while the liquid check failed.
8 reports: 2184009-2026-00614, 2184009-2026-00532, 2184009-2026-00475, 2184009-2026-00767, 2184009-2026-00769, 2184009-2026-00770, 2184009-2026-00578, 2184009-2026-00595
One report describes the device giving correct readings but being reported as not functioning, and another describes it failing temperature checks below its set point. The manufacturer did not state a patient effect in these reports.
2 reports: 2184009-2026-00768, 2184009-2026-00769
One report describes a channel, which is one of the device's separate testing paths, freezing a few seconds after start-up. The same report says the device had given inconsistent values on both channels for about a month, despite repeated internal cleaning.
Report: 2184009-2026-01400
A report describes intermittent channel failures along with a loud fan noise from the device. The manufacturer did not state a patient effect.
Report: 2184009-2026-00532
Written by AI from the narratives of the 20 newest reports that have one, received Apr 17, 2026 to Sep 23, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 11in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 5registered establishments that list this product code
Other codes under 21 CFR 864.5400
Product codes classified under the same regulation.
Watch GKN in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get GKN every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in GKN. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:GKNRetrieved Nov 29, 2026, 7:29 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:GKNRetrieved Nov 29, 2026, 7:29 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:GKN · count decision_dateRetrieved Nov 29, 2026, 7:29 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:GKNRetrieved Nov 29, 2026, 7:29 PM UTC.
- device/recall · search product_code:GKN · count event_date_initiatedRetrieved Nov 29, 2026, 7:29 PM UTC.
- device/recall · search product_code:GKN · count res_event_numberRetrieved Nov 29, 2026, 7:29 PM UTC.
- device/event · search device.device_report_product_code:GKN AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 7:29 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:GKN AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 7:29 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:GKN AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 7:29 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:GKNRetrieved Nov 29, 2026, 7:29 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:GKNRetrieved Nov 29, 2026, 7:29 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:864.5400Retrieved Nov 29, 2026, 4:52 AM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.