Product codes

FDA product code

FWP: Prosthesis, Chin, Internal (FDA product code)

Follow FWP: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 878.3550
Review panel
General and Plastic Surgery
Premarket route
510(k)
Implant
Yes
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Not eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
661 in the last 12 months
Median review time
89 daysthe latest 25 decisions
Recall records
50 in the last 12 months
Adverse event reports
23last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code FWP per year, by decision date, 66 since 1976. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20170
20180
20192
20201
20210
20220
20231
20240
20250
2026 (so far)1

Latest decisions

NumberDevice and applicantDecision and review time
K254233 MEDPOR Customized Implant KitStryker Leibinger GmbH & Co KG Mar 12, 202673 days
K232593 FITme Customized Silicone ImplantKeosan Trading Co. Nov 22, 202389 days
K200610 Customized Contour ImplantImplantech Associates, Inc. Dec 23, 2020289 days
K191130 Customized Contour ImplantImplantech Associates, Inc. Aug 23, 2019116 days
K191916 Stryker CMF MEDPOR Priority Customized Implant KitStryker Leibinger GmbH & Co KG Aug 15, 201929 days
K153508 Stryker CMF MEDPOR Customized ImplantStryker Feb 5, 201660 days
K152463 SU-POR Patient-Specific Cranial Implant, SU-POR Patient-Specific Facial ImplantPoriferous, LLC Jan 20, 2016145 days
K143173 Stryker CMF MEDPOR Customized ImplantStryker Mar 23, 2015139 days
K141027 3D EPTFE; ANATOMICAL CHIN, CURVILINEAR CHIN, EXTENDED ANATOMICAL CHIN IMPLANT, SUBMALAR II IMPLANT, DORSAL COLUMELLA IMPImplantech Associates, Inc. Jul 10, 201479 days
K121315 STRYKER CMF MEDPOR CUSTOMIZED IMPLANTHowmedica Osteonics Corp. Nov 1, 2012183 days

Recalls

Recall records for product code FWP per year, by the date the recall began. Each recalled product is one record; these 5 records belong to 3 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20170
20180
20190
20200
20210
20220
20230
20240
20251
2026 (so far)0

Latest recalls

Recall and classFirm and reasonBegan
Z-0946-2026Class II DSAART LLCManufactured products deviated from established specifications and were distributed without obtaining the required FDA premarket approvals or clearances. Sep 26, 2025
Z-1090-2015Class II Stryker Craniomaxillofacial DivisionPotential for implant damage, implant cracking off/breaking intra-operatively. Dec 17, 2014
Z-1091-2015Class II Stryker Craniomaxillofacial DivisionPotential for implant deformation, implant mechanical stability too low, loss of barrier adhesion intra-operatively, implant damage, implant cracking off/break… Dec 17, 2014
Z-1092-2015Class II Stryker Craniomaxillofacial DivisionPotential for implant deformation, implant mechanical stability too low, loss of barrier adhesion intra-operatively, implant damage, implant cracking off/break… Dec 17, 2014
Z-2211-2013Class II Implantech Associates IncImplantech initiated a voluntary recall of certain lots of Terino Square Chin-Style II-Medicum (Lot # 848488 and 850156), after determining that the devices fr… Aug 30, 2013

Adverse event reports (MAUDE)

Medical device reports received per year that name product code FWP. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
201713
20187
201915
20206
202113
202212
20236
202410
202518
2026 (so far)21

Last 12 months by event type

  • Deaths0
  • Injuries12
  • Malfunctions11
  • Other or not stated0

Most reported device problems, last 12 months

  1. Device Appears to Trigger Rejection9
  2. Inadequacy of Device Shape and/or Size7
  3. Device Difficult to Setup or Prepare4
  4. Adverse Event Without Identified Device or Use Problem3
  5. Biocompatibility1
  6. Break1

In plain English: what the newest reports describe Written by AI

  • Multiple reports describe poor fit. Implants did not match the patient's bone, left gaps, or pressed toward the brain, so surgeons had to trim them heavily or use a different implant. In one report, an implant split while the doctor tried to make it lie flat.

    5 reports: 0008010177-2026-00114, 0008010177-2026-00073, 0008010177-2026-00025, 0008010177-2026-00026, 2028924-2026-00003

  • Some patients developed an orbital cutaneous fistula, an abnormal opening between the eye socket and the skin, 5 to 9 years after implantation. Revision surgery was needed. The reports come from a published medical journal article.

    2 reports: 0008010177-2026-00174, 0008010177-2026-00172

  • Some patients developed chronic granulomatous inflammation, a long-lasting inflammation made of clusters of immune cells, around the implant, and revision surgery was needed. The reports come from a published medical journal article.

    2 reports: 0008010177-2026-00171, 0008010177-2026-00173

  • Infection after implantation was reported in two reports. One patient needed revision surgery, and in the other, fluid leaked from the eye and part of the implant had to be removed during the investigation.

    2 reports: 0008010177-2026-00186, 0008010177-2026-00036

  • A report says the needles used to fix muscles could not pass through the narrow openings and suture channels of the implant. The surgeon had to apply significant force, the needle deviated sideways, and the surgeon was injured. No surgical delay or medical intervention was reported at intake.

    Report: 0008010177-2026-00046

Written by AI from the narratives of the 14 newest reports that have one, received Mar 3, 2026 to Sep 29, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
4,497in the FDA’s Global Unique Device Identification Database
Establishment listings
77registered establishments that list this product code

Product codes classified under the same regulation.

Watch FWP in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get FWP every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in FWP. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.