FDA product code
FRZ: Device, Patient Transfer, Powered (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 880.6775
- Review panel
- General Hospital
- Premarket route
- 510(k) exempt
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 50 in the last 12 months
- Median review time
- 125 daysthe latest 5 decisions
- Recall records
- 20 in the last 12 months
- Adverse event reports
- 2last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code FRZ per year, by decision date, 5 since 1984. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K121929 | ZEPHYR 'X-SERIES PATIENT TRANSFER SLEDDiacor, Inc. | Jan 17, 2013199 days |
| K910830 | INVALIFTAmerican Invalift Corp. | Jun 9, 1992468 days |
| K841728 | LCI MEDICAL INC AIR/PALLehigh Consolidated Industries | Jun 5, 198441 days |
| K834585 | CELLTRAK-BR&D Systems, Inc. | May 2, 1984125 days |
| K840796 | TENDER LIFTING CARELancer Assoc. | Apr 20, 198457 days |
Recalls
Recall records for product code FRZ per year, by the date the recall began. Each recalled product is one record; these 2 records belong to 2 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 1 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-2394-2023Class II | Stryker Sustainability Solutions39-inche-wide inflatable air transfer mattress packages contain 34-Inch-wide mattresses, so there may be a decreased surface area available to distribute the p… | Jun 21, 2023 |
| Recall | D.T. Davis Enterprises Ltd.Instructions for use are not adhering to the product | Sep 13, 2005 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code FRZ. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 3 |
| 2018 | 2 |
| 2019 | 2 |
| 2020 | 0 |
| 2021 | 3 |
| 2022 | 0 |
| 2023 | 4 |
| 2024 | 2 |
| 2025 | 3 |
| 2026 (so far) | 0 |
Last 12 months by event type
- Deaths0
- Injuries2
- Malfunctions0
- Other or not stated0
Most reported device problems, last 12 months
- Adverse Event Without Identified Device or Use Problem1
- Device-Device Incompatibility1
In plain English: what the newest reports describe Written by AI
Several reports describe patients sliding or falling off mats. Staff linked this to slippery mat surfaces, and to mats that deflated unevenly, so the weight shifted. In one report a patient had mild abrasions to the head and shoulder after sliding off an operating table.
4 reports: 12287282, 8615460, 9134533, 7921762
Three reports describe hover mats found stained when they were unwrapped or set up for surgery. Two of these say the mat or its packaging came from a reprocessed (previously used and cleaned) item. In one case the stained mat was removed from the operating room.
3 reports: 21046808, 20320597, 12703541
Two reports describe burns on the outer thigh of patients who had their legs in metal stirrups during an MRI scan. The reports say the thigh rested against the stainless steel part of a stirrup. The patients had spinal epidurals, so they may not have felt the heat. The manufacturer's stirrup part was named in both reports, and one report says the burn was not recognized until a second patient had a similar wound.
2 reports: MW5180551, MW5180428
One report describes a battery overheating and giving off fumes, which caused a burning smell in the building. The reporter's facility found several battery and charging problems on the same device model and advised staff to keep the unit plugged in and check its indicator light.
Report: MW5146491
One report describes a lift sling that broke while a patient was being lifted off a bed. The patient dropped only about 3 inches before reaching the bed. The sling was used with a weight scale.
Report: MW5146341
Written by AI from the narratives of the 20 newest reports that have one, received Feb 3, 2016 to Dec 15, 2025. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 335in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 65registered establishments that list this product code
Watch FRZ in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get FRZ every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in FRZ. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:FRZRetrieved Nov 29, 2026, 7:52 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:FRZRetrieved Nov 29, 2026, 7:52 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:FRZ · count decision_dateRetrieved Nov 29, 2026, 7:52 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:FRZRetrieved Nov 29, 2026, 7:52 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:FRZ · count event_date_initiatedRetrieved Nov 29, 2026, 7:52 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:FRZ · count res_event_numberRetrieved Nov 29, 2026, 7:52 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-2394-2023")Retrieved Nov 29, 2026, 7:52 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:FRZ AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 7:52 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:FRZ AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 7:52 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:FRZ AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 7:52 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:FRZRetrieved Nov 29, 2026, 7:52 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:FRZRetrieved Nov 29, 2026, 7:52 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:880.6775Retrieved Nov 29, 2026, 7:52 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil