FDA product code
FRI: Scale, Stand-On, Patient (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class I
- Regulation
- 21 CFR 880.2700
- Review panel
- General Hospital
- Premarket route
- 510(k) exempt
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- Yes
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 30 in the last 12 months
- Median review time
- 30 daysthe latest 3 decisions
- Recall records
- 20 in the last 12 months
- Adverse event reports
- 13last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code FRI per year, by decision date, 3 since 1977. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest decisions
Recalls
Recall records for product code FRI per year, by the date the recall began. Each recalled product is one record; these 2 records belong to 2 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-2217-2017Class II | Cardinal Scale Mfg CoOne complaint of a seat on a scale failing unexpectedly which can cause an injury. | Dec 23, 2015 |
| Z-0294-2012 | Pelstar, LlcThere is a potential for an electrical short when installing or replacing the batteries in the remote display head of the scales. This short may cause the batt… | Oct 7, 2011 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code FRI. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 0 |
| 2018 | 1 |
| 2019 | 6 |
| 2020 | 3 |
| 2021 | 5 |
| 2022 | 3 |
| 2023 | 1 |
| 2024 | 4 |
| 2025 | 2 |
| 2026 (so far) | 12 |
Last 12 months by event type
- Deaths0
- Injuries0
- Malfunctions13
- Other or not stated0
Most reported device problems, last 12 months
- Insufficient Device Problem Information3
- Material Split, Cut or Torn3
- Calibration Problem2
- Failure to Power Up2
- Battery Problem1
- Crack1
In plain English: what the newest reports describe Written by AI
Several reports describe power cords or power supplies with exposed or damaged wiring. One report describes a power supply whose end was melted with copper wires showing, and a facility says its power adapters fail often, with cords that thin, wear through and kink easily. The manufacturer reportedly told the facility the damage was due to user damage, not design.
6 reports: 1316463-2025-00093, 1316463-2025-00087, 1316463-2024-00159, 1316463-2024-00082, 3011196194-2022-00021, 13757188
Several reports describe batteries leaking, swelling, peeling or corroding inside the scale's battery compartment. In one, the swollen batteries stopped the scale working even after new batteries were fitted. In another, a user found leaking batteries and an error message.
4 reports: 3011196194-2026-00040, 3011196194-2022-00021, 3011196194-2021-00008, 3011196194-2021-00006
Some reports describe the scale failing to work at all. One says it would not turn on, and another says stepping on it produced nothing. A third says the scale was no longer calibrating and read up to 20 pounds off.
3 reports: 1417592-2026-47300, 1417592-2026-26054, 1417592-2026-57651
A report describes a patient stepping on an emergency department triage scale and its weighing surface shattering or exploding. Another report describes a scale's glass surface found cracked after a loud noise. A third reports a scale that was physically damaged, with no injury stated.
3 reports: 19163192, 3011196194-2026-00047, 1316463-2026-00311
A report describes a user falling while stepping off the scale and fracturing a hip. The device was not returned to the manufacturer for investigation.
Report: 3004594267-2021-00001
Written by AI from the narratives of the 20 newest reports that have one, received Mar 29, 2021 to Oct 5, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 665in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 97registered establishments that list this product code
Watch FRI in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get FRI every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in FRI. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:FRIRetrieved Nov 29, 2026, 9:47 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:FRIRetrieved Nov 29, 2026, 9:47 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:FRI · count decision_dateRetrieved Nov 29, 2026, 9:47 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:FRIRetrieved Nov 29, 2026, 9:47 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:FRI · count event_date_initiatedRetrieved Nov 29, 2026, 9:47 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:FRI · count res_event_numberRetrieved Nov 29, 2026, 9:47 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-2217-2017" "Z-0294-2012")Retrieved Nov 29, 2026, 9:47 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:FRI AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 9:47 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:FRI AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 9:47 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:FRI AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 9:47 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:FRIRetrieved Nov 29, 2026, 9:47 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:FRIRetrieved Nov 29, 2026, 9:47 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:880.2700Retrieved Nov 29, 2026, 9:47 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.