Product codes

FDA product code

FFX: System, Gastrointestinal Motility (Electrical) (FDA product code)

Follow FFX: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 876.1725
Review panel
Gastroenterology and Urology
Premarket route
510(k)
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
702 in the last 12 months
Median review time
120 daysthe latest 25 decisions
Recall records
72 in the last 12 months

510(k) and De Novo decisions

Decisions for product code FFX per year, by decision date, 70 since 1976. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20172
20181
20194
20201
20210
20220
20233
20241
20251
2026 (so far)1

Latest decisions

NumberDevice and applicantDecision and review time
K262621 Rapid Barostat Bag (RBB) Pump and Catheter SystemMui Scientific, Inc. Sep 30, 202664 days
K252605 mcompass Anorectal Balloon Expulsion Catheter (RMD-003-001)Medspira, LLC Dec 5, 2025109 days
K240007 Solar Compact (G4-1)Laborie Medical Technologies, Corp. May 6, 2024125 days
K231861 EndoflipTM 300 SystemCovidien, LLC Jul 21, 202328 days
K223705 EndoflipTM 300Covidien, LLC Apr 14, 2023123 days
K222000 Transit-PelletsMedifactia AB Jan 18, 2023195 days
K201106 IntraMarX 3D Radiopaque MarkerAnx Robotica Corp May 21, 202027 days
K191087 IntraMarX Radiopaque MarkersAnx Robotica Corp Dec 18, 2019238 days
K190208 Diversatek Healthcare High Resolution Impedance Manometry (HRiM) ProbeDiversatek Healthcare Oct 3, 2019241 days
K181760 Transit-PelletsMedifactia AB Aug 8, 2019401 days

Recalls

Recall records for product code FFX per year, by the date the recall began. Each recalled product is one record; these 7 records belong to 7 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20170
20180
20190
20200
20210
20220
20231
20240
20251
2026 (so far)2

Latest recalls

Recall and classFirm and reasonBegan
Z-2669-2026Class II Laborie Medical Technologies CorpUnder extended clinical use conditions (e.g., long duration studies), moisture generated within the pneumatic circuit may accumulate inside the system's intern… May 21, 2026
Z-2950-2026Class II Covidien, LLCThe system's 24-hour usage limit results in catheters incorrectly flagged as expired, making them unavailable for clinical procedures. May 13, 2026
Z-2324-2025Class III Diversatek HealthcareAnorectal manometry procedure kit outer kit label is mislabeled with an incorrect expiration date; the manufacturing date was used in place of the expiration d… Jun 4, 2025
Z-2521-2023Class III Diversatek HealthcareGI Software labelled as revision 3.3.0 is incorrectly labeled and is in fact revision 3.2.0. Jul 25, 2023
Z-0135-2012 Sierra Scientific Instruments IncThe recall was initiated after Given Imaging received a test report during a Canadian licensing process where the AR ManoShield exhibited pinholes and did not… Oct 12, 2011

Adverse event reports (MAUDE)

Medical device reports received per year that name product code FFX. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
201710
201817
201927
202020
202126
202231
202351
202459
202532
2026 (so far)99

Last 12 months by event type

This section didn’t load from openFDA. It’s retried within the hour.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
1,348in the FDA’s Global Unique Device Identification Database
Establishment listings
65registered establishments that list this product code

Watch FFX in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get FFX every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in FFX. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.