Product codes

FDA product code

EWL: Prosthesis, Laryngeal (Taub) (FDA product code)

Follow EWL: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 874.3730
Review panel
Ear, Nose and Throat
Premarket route
510(k)
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Not eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
400 in the last 12 months
Median review time
122 daysthe latest 25 decisions
Recall records
42 in the last 12 months
Adverse event reports
23last 12 months, 1 death

510(k) and De Novo decisions

Decisions for product code EWL per year, by decision date, 40 since 1980. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20170
20180
20190
20200
20210
20220
20230
20241
20250
2026 (so far)0

Latest decisions

NumberDevice and applicantDecision and review time
K240763 FirstFit Surgical KitFreudenberg Medical, LLC Dec 18, 2024273 days
K141380 PROVOX FREEHANDS FLEXIVOICEAtos Medical AB Sep 26, 2014122 days
K131947 PROVOX VEGA PUNCTURE SETAtos Medical AB Oct 18, 2013113 days
K092593 PROVOX ACTIVALVE (SIZE 4.5MM, 6MM, 8MM, 10MM, 12.5MM)Atos Medical AB Oct 14, 2010416 days
K093258 BLOM-SINGER INDWELLING TEP OCCLUDERHelix Medical, LLC Mar 3, 2010135 days
K090455 PROVOX VEGA, MODEL 8130-8135Atos Medical AB Jun 5, 2009102 days
K082206 BLOM-SINGER DUAL VALVE INDWELLING VOICE PROSTHESIS, MODELS DV-1600 SERIES, DV-2000 SERIESHelix Medical, LLC Sep 4, 200830 days
K043138 PROVOX NID VOICE REHABILITATION SYSTEM, MODEL REF 7101 TO 7134Atos Medical AB Nov 22, 200410 days
K022125 PROVOX FREEHANDS HME, MODEL 7710Atos Medical AB Jul 30, 200229 days
K991587 BLOM-SINGER INDWELLING 2000 VOICE PROSTHESISHelix Medical, Inc. Jun 25, 199949 days

Recalls

Recall records for product code EWL per year, by the date the recall began. Each recalled product is one record; these 4 records belong to 3 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20170
20180
20190
20200
20210
20220
20230
20240
20250
2026 (so far)2

Latest recalls

Recall and classFirm and reasonBegan
Z-2909-2026Class II Freudenberg Medical, LLCIncorrect labeling- The labeled box for the Large Esophageal Flange contains a Large Esophageal and Tracheal Flange, and the labeled box for the Large Esophage… Jun 17, 2026
Z-2910-2026Class II Freudenberg Medical, LLCIncorrect labeling- The labeled box for the Large Esophageal Flange contains a Large Esophageal and Tracheal Flange, and the labeled box for the Large Esophage… Jun 17, 2026
Z-0128-2014Class III Helix Medical LLCInHealth Technologies (IHT) is recalling the Blom-Singer Adjustable Tracheostoma Valve II (ATSV II) because the body component of the device is missing from th… Oct 1, 2013
Z-2336-2012Class III Helix Medical LLCHelix Medical LLC is recalling the BLOM-SINGER indwelling TEP Occluder due to an invalid expiration date. Mar 8, 2012

Adverse event reports (MAUDE)

Medical device reports received per year that name product code EWL. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
201715
20180
20197
20202
20217
20225
20234
20248
202512
2026 (so far)23

Last 12 months by event type

  • Deaths1
  • Injuries20
  • Malfunctions2
  • Other or not stated0

Most reported device problems, last 12 months

  1. Biofilm coating in Device6
  2. Fluid/Blood Leak4
  3. Device Fell2
  4. Fracture2
  5. Installation-Related Problem2
  6. Material Separation2

In plain English: what the newest reports describe Written by AI

  • Fluid leaked around or through the device into the airway, sometimes causing pneumonia and hospital stays. In one report, leakage began right after placement and the device was later presumed swallowed; the patient died, but the manufacturer found no link between the death and the device.

    4 reports: 3007501574-2026-00005, 3002806269-2026-00024, 2025182-2026-00002, 3002806269-2026-00017

  • Devices moved out of place: a device came loose and traveled into the airway and lungs, requiring surgery to remove it. In another case a device stuck partly in the throat wall and had to be retrieved in an operation.

    3 reports: 3002806269-2026-00029, 3007501574-2026-00010, 3007501574-2026-00003

  • A part of the valve broke off or was missing, letting food, fluid and mucus enter the airway. One patient was hospitalized with shortness of breath and extra mucus, and the missing piece, a small plastic flap, was later found deep in the windpipe.

    3 reports: 3002806269-2026-00027, 3007501574-2026-00008, 2025182-2026-00001

  • During removal, the flange at the outer end of the valve tore away, leaving the rest of the valve in the puncture site. The patient swallowed the valve and was then checked by scope and chest X-ray to make sure it had not entered the airway.

    2 reports: 3002806269-2026-00026, 3007501574-2026-00007

  • The throat narrowed, blocking food from passing and leaving one patient unable to speak. Balloon dilation was stopped partway because of the device, and steroid injections gave partial relief, allowing small amounts of liquid, though the patient lost weight.

    2 reports: 3002806269-2026-00021, 3007501574-2026-00002

Written by AI from the narratives of the 16 newest reports that have one, received Feb 18, 2026 to Oct 12, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
348in the FDA’s Global Unique Device Identification Database
Establishment listings
23registered establishments that list this product code

Product codes classified under the same regulation.

Watch EWL in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get EWL every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in EWL. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.