Product codes

FDA product code

ETB: Prosthesis, Partial Ossicular Replacement (FDA product code)

Follow ETB: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 874.3450
Review panel
Ear, Nose and Throat
Premarket route
510(k)
Implant
Yes
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Not eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
742 in the last 12 months
Median review time
28 daysthe latest 25 decisions
Recall records
31 in the last 12 months
Adverse event reports
1last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code ETB per year, by decision date, 74 since 1976. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20170
20180
20190
20200
20210
20220
20230
20240
20252
2026 (so far)2

Latest decisions

NumberDevice and applicantDecision and review time
K260720 mWING Stapes Prosthesis (0.4 x 3.50 mm) (58717); mWING Stapes Prosthesis (0.4 x 3.75 mm) (58719); mWING Stapes Prosthesis (0.4 x 4.00 mm) (58721); mWING Stapes Prosthesis (0.4 x 4.25 mm) (58723); mWING Stapes Prosthesis (0.4 x 4.50 mm) (58725); mWING Stapes Prosthesis (0.4 x 4.75 mm) (58727); mWING Stapes Prosthesis (0.4 x 5.00 mm) (58729); mWING Stapes Prosthesis (0.4 x 5.25 mm) (58731); mWING Stapes Prosthesis (0.4 x 5.50 mm) (58733); mWING Stapes Prosthesis (0.5 x 3.50 mm) (58735)MED-EL Elektromedizinische Geräte GmbH Apr 3, 202629 days
K260718 mGRIP Partial Prosthesis (0.75 mm) (58669); mGRIP Partial Prosthesis (1.00 mm) (58671); mGRIP Partial Prosthesis (1.25 mm) (58673); mGRIP Partial Prosthesis (1.50 mm) (58675); mGRIP Partial Prosthesis (1.75 mm) (58677); mGRIP Partial Prosthesis (2.00 mm) (58679); mGRIP Partial Prosthesis (2.25 mm) (58681); mGRIP Partial Prosthesis (2.50 mm) (58683); mGRIP Partial Prosthesis (3.00 mm) (58685); mGRIP Partial Prosthesis (3.50 mm) (58687); mDISC Partial Prosthesis (58850)MED-EL Elektromedizinische Geräte GmbH Apr 3, 202629 days
K241142 mAXIS Stapes Prosthesis, mLOOP Stapes Prosthesis, mZAM Stapes Prosthesis and mFIX Stapes ProsthesisMED-EL Elektromedizinische Geräte GmbH Jan 17, 2025267 days
K241261 mCLIP Partial Prosthesis, mCLIP ARC Partial Prosthesis, mXACT Partial Prosthesis and mXACT PRO Partial Prosthesis KitMED-EL Elektromedizinische Geräte GmbH Jan 17, 2025266 days
K130512 NITIFLEX STAPES PROSTHESIS, DETROIT PISTON, SKARZYNSKI PISTON, ROBERSON STAPES PROSTHESISHeinz Kurz GmbH Medizintechnik Dec 2, 2013278 days
K112616 NITIBOND STAPES PROSTHESIS, NITIPLAST STAPES PROSTHESISHeinz Kurz GmbH Medizintechnik Feb 8, 2013519 days
K080070 K-HELIX PISTON AND PORP PARTIAL OSSICULAR REPLACEMENT PROSTHESESGrace Medical, Inc. Apr 17, 200898 days
K063374 NITINOL LOCKING INCUDO-STAPEDIAL JOINT, ANGULAR PISTON, PISTON, MALLEUS PISTON, PORP AND FOOTPLATE SHOEGrace Medical, Inc. Dec 6, 200628 days
K061853 GRACE MEDICAL ADJUSTABLE AND FIXED LENGTH PARTIAL & TOTAL OSSICULAR REPLACEMENT PROSTHESESGrace Medical, Inc. Aug 18, 200649 days
K060518 PARTIAL OSSICULAR REPLACEMENT PROSTHESESGrace Medical, Inc. Jun 28, 2006121 days

Recalls

Recall records for product code ETB per year, by the date the recall began. Each recalled product is one record; these 3 records belong to 2 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20170
20180
20190
20202
20210
20220
20230
20240
20250
2026 (so far)1

Latest recalls

Recall and classFirm and reasonBegan
Z-2308-2026Class II Grace Medical, Inc.Due to incorrect functional length on device labeling. Apr 3, 2026
Z-2813-2020Class II Olympus Corporation of the AmericasMislabeled outer carton: packaged with the outer carton label stating the contents were the 70143662 Smart Piston but the inside package contents was the 70142… Jul 14, 2020
Z-2814-2020Class II Olympus Corporation of the AmericasMislabeled outer carton: packaged with the outer carton label stating the contents were the 70143662 Smart Piston but the inside package contents was the 70142… Jul 14, 2020

Adverse event reports (MAUDE)

Medical device reports received per year that name product code ETB. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
20173
20180
20196
20204
20211
20226
20231
20243
20255
2026 (so far)1

Last 12 months by event type

  • Deaths0
  • Injuries0
  • Malfunctions1
  • Other or not stated0

Most reported device problems, last 12 months

  1. Break1

In plain English: what the newest reports describe Written by AI

  • In several reports, the implant was found loose or uncapped inside an opened sterile package, or out of its holder, before use. In most of these, staff opened a new implant instead.

    5 reports: 1057421-2025-00001, 21525709, 19350507, 1057421-2024-00001, MW5150793

  • Several reports describe the implant being bent, crimped (folded or pinched), or having a hook too closed, so it could not be placed properly. In one case a bent implant was used, and in another a bent shaft found at a later operation led to a second operation.

    4 reports: 1037007-2022-00020, 1037007-2022-00003, 19350507, 1000022662-2020-00001

  • Two reports describe the implant breaking or coming apart during placement. In one, the implant broke while being inserted. In the other, the nitinol (a shape-memory metal) part separated from its wire, and a new implant was opened.

    2 reports: 50237501, 1037007-2022-00012

  • Two reports describe the implant moving out of position after surgery. In one, a patient felt a pop with ear discharge about three months after surgery, and the implant was found in the ear canal, stuck on the eardrum. In the other, the implant fell out of place after a patient strained while shoveling snow, and it is unclear whether harm occurred.

    2 reports: 9710014-2025-00849, 1037007-2022-00008

  • Two reports say the patient's hearing did not improve after surgery. In both, the surgeon removed the implant and placed a new one, and the narratives do not state a specific cause.

    2 reports: 9710014-2025-00449, 1000022662-2020-00001

Written by AI from the narratives of the 18 newest reports that have one, received Jul 2, 2020 to Aug 13, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
1,103in the FDA’s Global Unique Device Identification Database
Establishment listings
101registered establishments that list this product code

Product codes classified under the same regulation.

Watch ETB in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get ETB every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in ETB. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.