Product codes

FDA product code

DWC: Controller, Temperature, Cardiopulmonary Bypass (FDA product code)

Follow DWC: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 870.4250
Review panel
Cardiovascular
Premarket route
510(k)
Implant
No
Life-sustaining or supporting
Yes
GMP exempt
No
Third-party review
Not eligible
Summary malfunction reporting
Ineligible
510(k) and De Novo decisions
300 in the last 12 months
Median review time
97 daysthe latest 25 decisions
Recall records
520 in the last 12 months
Adverse event reports
526last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code DWC per year, by decision date, 30 since 1981. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20170
20180
20190
20201
20210
20222
20230
20240
20250
2026 (so far)0

Latest decisions

NumberDevice and applicantDecision and review time
K220635 Heater-Cooler System 3TLivanova Deutschland, GmbH Nov 16, 2022257 days
K212657 Quantum Heater-CoolerSpectrum Medical , Ltd. May 6, 2022256 days
K191402 Heater-Cooler System 3TLivanova Deutschland, GmbH Feb 25, 2020277 days
K140012 SORIN FLEXTHERMSorin Group Deutschland GmbH Jan 31, 201429 days
K130300 HEATER-COOLER UNIT HCU 40Maquet Cardiopulmonary, AG Jul 1, 2013144 days
K122813 HEMOTHERM MODEL 400CE COOLER/HEATER SYSTEMCincinnati Sub-Zero Products, Inc. Dec 18, 201296 days
K102147 CARDIOQUIP MODULAR COOLER-HEATER MODEL MCH-1000Cardioquip Llp Nov 19, 2010112 days
K071521 TERUMO HX2 TEMPERATURE MANAGEMENT SYSTEMTerumo Cardiovascular Systems Corp. Nov 21, 2007170 days
K052601 STOCKERT HEATER-COOLER SYSTEM 3TSorin Group Deutschland GmbH Jun 6, 2006258 days
K031544 HEATER COOLER UNIT, MODEL HCU 30Jostra AB Aug 7, 200383 days

Recalls

Recall records for product code DWC per year, by the date the recall began. Each recalled product is one record; these 52 records belong to 29 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20178
20188
20190
20201
202110
20220
20235
20240
20250
2026 (so far)0

Latest recalls

Recall and classFirm and reasonBegan
Z-0565-2024Class II CardioQuip, LLCThere is the potential that cooler-heater devices may contain bacterial contamination which could result in patient infection. Oct 24, 2023
Z-0566-2024Class II CardioQuip, LLCThere is the potential that cooler-heater devices may contain bacterial contamination which could result in patient infection. Oct 24, 2023
Z-0438-2024Class II CardioQuip, LLCThere is a potential for the heating-chamber to malfunction resulting in melting/burning of the heater chamber components. Oct 19, 2023
Z-0439-2024Class II CardioQuip, LLCThere is a potential for the heating-chamber to malfunction resulting in melting/burning of the heater chamber components. Oct 19, 2023
Z-0073-2024Class II LivaNova Deutschland GmbHThe firm's labeling (Quick Reference Guide) contains incorrect information in the maintenance schedule for the Heater-Cooler device. Aug 3, 2023

Adverse event reports (MAUDE)

Medical device reports received per year that name product code DWC. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
2017716
2018564
2019697
2020548
2021622
2022605
2023928
2024555
2025680
2026 (so far)480

Last 12 months by event type

  • Deaths0
  • Injuries13
  • Malfunctions512
  • Other or not stated1

Most reported device problems, last 12 months

  1. Microbial Contamination of Device301
  2. Temperature Problem65
  3. Pumping Problem33
  4. Insufficient Cooling32
  5. Pumping Stopped27
  6. Electrical Shorting9

In plain English: what the newest reports describe Written by AI

  • Pump blockage error codes appeared during priming, the step where the circuit is filled with fluid before use. The displays showed codes meaning a pump was blocked or too slow. No patient injury was reported.

    4 reports: 9611109-2026-90524, 9611109-2026-90513, 9611109-2026-90503, 9611109-2026-90474

  • Several reports described cooling or condenser parts failing: a compressor found defective, a unit unable to provide proper cooling during a procedure, and a condenser that could not function properly. One report noted that water heating still worked during a priming event. No patient injury was reported.

    4 reports: 9611109-2026-90501, 9611109-2026-90502, 9611109-2026-90454, 9611109-2026-90450

  • Water contamination inside the device was reported in several reports. In a number of these, the customer or a technician suspected contaminated internal tubing, and lab tests of the water (HPC, a count of bacteria-type organisms) were outside the company's specification, indicating device contamination. No patient involvement was reported.

    3 reports: 3007899424-2026-00240, 3007899424-2026-00241, 3007899424-2026-00211

  • One report described a power cord plug that had melted. A separate report described a loaner unit showing a pump malfunction code during water replacement, and another described a patient-side pump that did not work properly during a service visit. No patient harm was reported in these.

    3 reports: 1516825-2026-00030, 9611109-2026-90534, 9611109-2026-90436

  • One report described a unit found contaminated with Nontuberculous Mycobacteria, a group of environmental bacteria that can cause lung and other infections. The report stated there was no patient injury.

    Report: 9611109-2026-90521

Written by AI from the narratives of the 15 newest reports that have one, received Sep 21, 2026 to Oct 29, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
87in the FDA’s Global Unique Device Identification Database
Establishment listings
13registered establishments that list this product code

Watch DWC in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get DWC every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in DWC. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.