Product codes

FDA product code

DTS: Sucker, Cardiotomy Return, Cardiopulmonary Bypass (FDA product code)

Follow DTS: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 870.4420
Review panel
Cardiovascular
Premarket route
510(k) exempt
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Not eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
90 in the last 12 months
Median review time
64 daysthe latest 9 decisions
Recall records
110 in the last 12 months
Adverse event reports
12last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code DTS per year, by decision date, 9 since 1979. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20170
20180
20190
20200
20210
20220
20230
20240
20250
2026 (so far)0

Latest decisions

NumberDevice and applicantDecision and review time
K053021 ESTECH CLEARVIEW MV ATRIAL DEPRESSOREstech, Inc. Jan 26, 200692 days
K982891 CALIFORNIA MEDICAL LABORATORIES INC. STANDARD AND SMALL RIGID TIP SUCTION WANDSCalifornia Medical Laboratories, Inc. Sep 30, 199844 days
K963756 IBC CARDIAC SUCTION WARDInternational Biophysics Corp. Nov 14, 1997423 days
K905008 FLEXIBLE INTRACARDIAC SUCTION DEVICEGish Biomedical, Inc. Jan 23, 199178 days
K900339 HAND-HELD INTRACARDIAC SUCKERS W/ERGONOMIC HANDLESShiley, Inc. Mar 28, 199064 days
K891906 ROBICSEK INTRACARDIAC SUCTION WANDGish Biomedical, Inc. Jun 13, 198978 days
K801641 OHIO DISPOSABLE CLEAR SUCTION INSTRUMENTOhio Medical Products Jul 28, 198011 days
K801400 DISPOSABLE SUCKER SYSTEMTexas Medical Products, Inc. Jun 30, 198017 days
K791846 DISPOSSABLE SUCTION TUBE, 6 FRENCHDlp, Inc. Oct 2, 197913 days

Recalls

Recall records for product code DTS per year, by the date the recall began. Each recalled product is one record; these 11 records belong to 7 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20170
20180
20190
20200
20210
20220
20230
20241
20250
2026 (so far)0

Latest recalls

Recall and classFirm and reasonBegan
Z-1804-2024Class II Medtronic Perfusion SystemsPotential for unsealed sterile packing. Mar 19, 2024
Z-2422-2015Class II International Biophysics Corp.It was discovered that the heat seal placed onto sterile pouches may not meet strength acceptance criteria. Aug 7, 2015
Z-0649-2013Class II California Medical Laboratories IncCalifornia Medical Laboratories, Inc. has identified an issue with the Rigid Tip Suction Wand. Due to insufficient adhesive application, the distal tip has the… Nov 2, 2012
Z-0650-2013Class II California Medical Laboratories IncCalifornia Medical Laboratories, Inc. has identified an issue with the Rigid Tip Suction Wand. Due to insufficient adhesive application, the distal tip has the… Nov 2, 2012
Z-0273-2012 International Biophysics CorpPotential for the sterility to be compromised. Oct 5, 2011

Adverse event reports (MAUDE)

Medical device reports received per year that name product code DTS. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
20172
20183
20193
20202
202111
202211
202312
202420
202517
2026 (so far)11

Last 12 months by event type

  • Deaths0
  • Injuries4
  • Malfunctions8
  • Other or not stated0

Most reported device problems, last 12 months

  1. Delivered as Unsterile Product3
  2. Material Separation3
  3. Obstruction of Flow3
  4. Detachment of Device or Device Component1
  5. Material Deformation1
  6. Mechanical Problem1

In plain English: what the newest reports describe Written by AI

  • Several reports describe the wire or wire mesh inside the cannula (the tube placed in the heart) deforming, loosening, or protruding, in one case causing trauma (injury) to heart tissue. The manufacturer's findings are not stated in these reports.

    3 reports: 3003955307-2026-00198, 3003955307-2026-00085, 2184009-2025-00784

  • Two reports describe a tip or end piece detaching inside the body. One report describes a spring uncoiling and a metal ball end falling off during surgery, which had to be located inside the patient's chest. Another describes a bullet-shaped tip of a weighted sump tube breaking off inside the chest; the patient says it remains there and that the manufacturer told them an MRI is not allowed because of it.

    3 reports: 3003955307-2026-00314, MW5174345, 2184009-2025-01591

  • Several reports describe suction devices (suckers) that were blocked or would not apply vacuum, either straight from the package or during use. In each case the device was replaced, and the cause was not clear in these reports.

    3 reports: 3003955307-2026-00020, 2184009-2025-00712, 2184009-2025-00364

  • Several reports describe packaging problems found before use, including damaged outer packaging, a broken sterile seal, and a hair seen inside the sterile packaging. In each case the device was replaced and no patient was involved.

    3 reports: 2184009-2026-00139, 2184009-2026-00007, 2184009-2025-00680

  • One report describes a cannula's introducer (the guide used to insert it) hitting the cannula tip, with the tip reported broken through. It is unknown whether the devices were replaced before the procedure.

    Report: 2184009-2025-00785

Written by AI from the narratives of the 14 newest reports that have one, received Feb 28, 2025 to Oct 28, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
94in the FDA’s Global Unique Device Identification Database
Establishment listings
26registered establishments that list this product code

Watch DTS in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get DTS every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in DTS. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.