FDA product code
DPT: Probe, Blood-Flow, Extravascular (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 870.2120
- Review panel
- Cardiovascular
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 480 in the last 12 months
- Median review time
- 115 daysthe latest 25 decisions
- Recall records
- 240 in the last 12 months
- Adverse event reports
- 62last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code DPT per year, by decision date, 48 since 1978. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 1 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 1 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K233076 | Laser Speckle Imaging System (RFLSI CZW)Rwd Life Science Co., Ltd. | May 28, 2024245 days |
| K163339 | SPECTRALMD DEEPVIEW WOUND IMAGING SYSTEM 2.0Spectralmd, Inc. | Aug 17, 2017262 days |
| K152749 | moorLDI2-IR Infrared laser Doppler imager, moorLDI2-HIR High Resolution Infrared laser Doppler imager, moorLDI2-VR Visible Red laser Doppler imagerMoor Instruments, Ltd. | May 12, 2016232 days |
| K142932 | Deltex Medical KDP72 Doppler ProbeDeltex Medical Limited | Jan 22, 2015105 days |
| K132730 | BIO-PROBE DISPOSABLE INSERT WITH; BALANCE BIOSURFACE, CARMEDA BIOACTIVE SURFACE, TRILLIUM BIOSURFACEMedtronic, Inc. | May 30, 2014269 days |
| K133903 | BIO-PROBE TRANSDUCER (ADULT)Medtronic, Inc. | Apr 10, 2014111 days |
| K132163 | MOORLDLS-BI LASER DOPPLER BURNS IMAGERMoor Instruments, Ltd. | Jan 29, 2014201 days |
| K124049 | DEEPVIEW DIGITAL VIDEO PHYSIOLOGICAL PORTABLE IMAGING SYSTEMSpectralmd, Inc. | Apr 18, 2013108 days |
| K123216 | AIMAGO EASYLDI MICROCIRCULATION CAMERA (STAND PLUS CAMERA ARM), AIMAGO EASY LDI STAND, INCLUDING 4-WHEELED STAND BASE ANAimago SA | Feb 7, 2013115 days |
| K122943 | MOORFLPI-2 FULL-FIELD LASER PERFUSION IMAGERMoor Instruments, Ltd. | Jan 3, 2013101 days |
Recalls
Recall records for product code DPT per year, by the date the recall began. Each recalled product is one record; these 24 records belong to 3 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 22 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-0184-2025Class II | Smiths Medical ASD, Inc.Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device… | Sep 24, 2024 |
| Z-0185-2025Class II | Smiths Medical ASD, Inc.Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device… | Sep 24, 2024 |
| Z-0186-2025Class II | Smiths Medical ASD, Inc.Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device… | Sep 24, 2024 |
| Z-0187-2025Class II | Smiths Medical ASD, Inc.Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device… | Sep 24, 2024 |
| Z-0188-2025Class II | Smiths Medical ASD, Inc.Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device… | Sep 24, 2024 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code DPT. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 0 |
| 2018 | 3 |
| 2019 | 1 |
| 2020 | 3 |
| 2021 | 13 |
| 2022 | 46 |
| 2023 | 21 |
| 2024 | 169 |
| 2025 | 53 |
| 2026 (so far) | 57 |
Last 12 months by event type
- Deaths0
- Injuries1
- Malfunctions61
- Other or not stated0
Most reported device problems, last 12 months
- Fluid/Blood Leak46
- Incorrect, Inadequate or Imprecise Result or Readings5
- Leak/Splash5
- Break2
- Crack2
- Gas/Air Leak2
In plain English: what the newest reports describe Written by AI
Fluid leaked from the pressure sensor or its connections, such as a T-valve joint, a saline line connection, or the arterial needle connection. In each of these reports the leaking part was replaced, and in one report the patient needed ventilator support for respiratory failure while monitoring continued.
5 reports: 9616567-2026-00060, 9616567-2026-00031, 9616567-2026-00065, 1526863-2026-00140, 1526863-2026-00094
Blood leaked at the connection, and the sensor showed no pressure waveform, so the arterial blood pressure was not displayed. In one report the blood pressure was unstable and the sensor was replaced immediately, and the manufacturer considered a sensor quality issue.
3 reports: 1526863-2026-00130, 1526863-2026-00103, 1526863-2026-00096
Two reports say the product leaked and could not be zeroed, which is the calibration step done at setup. In a third report, the manufacturer could not confirm the leak because the device was not returned for testing, so no cause was found.
3 reports: 9616567-2026-00004, 9616567-2026-00007, 9616567-2026-00006
Two reports describe an email stating that three units of a monitoring kit showed liquid leakage. The reports do not say what the manufacturer found or whether any patient was affected.
2 reports: 1526863-2026-00092, 1526863-2026-00091
One report says blood flowed back into the tubing after the transducer was connected to a radial artery catheter, which is a thin tube placed in the wrist artery. Blood return and leakage stopped after the transducer was replaced.
Report: 9616567-2026-00033
Written by AI from the narratives of the 14 newest reports that have one, received Apr 10, 2026 to Oct 21, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 209in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 34registered establishments that list this product code
Other codes under 21 CFR 870.2120
Product codes classified under the same regulation.
Watch DPT in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get DPT every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in DPT. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:DPTRetrieved Nov 29, 2026, 5:15 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:DPTRetrieved Nov 29, 2026, 5:15 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:DPT · count decision_dateRetrieved Nov 29, 2026, 5:15 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:DPTRetrieved Nov 29, 2026, 5:15 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:DPT · count event_date_initiatedRetrieved Nov 29, 2026, 5:15 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:DPT · count res_event_numberRetrieved Nov 29, 2026, 5:15 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-0184-2025" "Z-0185-2025" "Z-0186-2025" "Z-0187-2025" "Z-0188-2025")Retrieved Nov 29, 2026, 5:15 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:DPT AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 5:15 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:DPT AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 5:15 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:DPT AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 5:16 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:DPTRetrieved Nov 29, 2026, 5:15 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:DPTRetrieved Nov 29, 2026, 5:15 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:870.2120Retrieved Nov 29, 2026, 5:15 PM UTC. FDA data updated Nov 16, 2026.