FDA product code
DOF: Lead, Atomic Absorption (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 862.3550
- Review panel
- Clinical Toxicology
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Ineligible
- 510(k) and De Novo decisions
- 60 in the last 12 months
- Median review time
- 195 daysthe latest 6 decisions
- Recall records
- 214 in the last 12 months
510(k) and De Novo decisions
Decisions for product code DOF per year, by decision date, 6 since 1985. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K142705 | LeadCare Plus Blood Lead Testing SystemMagellan Diagnostics | Jul 7, 2015288 days |
| K123563 | LEADCARE (R) ULTRA (TM) BLOOD LEAD TESTING SYSTEMMagellan Diagnostics | Aug 20, 2013274 days |
| K052549 | LEADCARE II BLOOD LEAD TESTING SYSTEMEsa Biosciences, Inc. | Oct 6, 200520 days |
| K971640 | LEADCARE BLOOD LEAD TESTING SYSTEMEsa, Inc. | Sep 9, 1997130 days |
| K936060 | ATOMSPEC GF BLOOD LEAD WORKSTATIONThermo Jarrell Ash Corp.Div.Thermo Inst. | Sep 6, 1994260 days |
| K843839 | STAT/PB REAGENTTri-Tech, Inc. | Jan 3, 198594 days |
Recalls
Recall records for product code DOF per year, by the date the recall began. Each recalled product is one record; these 21 records belong to 13 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 2 |
| 2018 | 3 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 3 |
| 2022 | 1 |
| 2023 | 1 |
| 2024 | 1 |
| 2025 | 0 |
| 2026 (so far) | 4 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-0162-2027Class III | Magellan Diagnostics, Inc.In kit sub-lot 2530M-01 the date of expiration printed on the outer labeling may indicate the wrong expiration date of July 22, 2026 while the correct expirati… | Sep 4, 2026 |
| Z-1893-2026Class II | Magellan Diagnostics, Inc.Customers reported observing unexpectedly elevated results when using certain third-party micro-collection devices to collect capillary samples for testing wit… | Mar 13, 2026 |
| Z-1894-2026Class II | Magellan Diagnostics, Inc.Customers reported observing unexpectedly elevated results when using certain third-party micro-collection devices to collect capillary samples for testing wit… | Mar 13, 2026 |
| Z-1895-2026Class II | Magellan Diagnostics, Inc.Customers reported observing unexpectedly elevated results when using certain third-party micro-collection devices to collect capillary samples for testing wit… | Mar 13, 2026 |
| Z-0584-2026Class II | Magellan Diagnostics, Inc.Calibration Buttons being used in the kitting process were incorrectly labeled as Lot 2323M for lot 2333M. The use of mislabeled calibration Button with Lot #2… | Feb 23, 2024 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code DOF. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
This section didn’t load from openFDA. It’s retried within the hour.
Last 12 months by event type
This section didn’t load from openFDA. It’s retried within the hour.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 19in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 6registered establishments that list this product code
Other codes under 21 CFR 862.3550
Product codes classified under the same regulation.
Watch DOF in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get DOF every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in DOF. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:DOFRetrieved Nov 28, 2026, 8:39 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:DOFRetrieved Nov 28, 2026, 8:39 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:DOF · count decision_dateRetrieved Nov 28, 2026, 8:39 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:DOFRetrieved Nov 28, 2026, 8:39 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:DOF · count event_date_initiatedRetrieved Nov 28, 2026, 8:39 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:DOF · count res_event_numberRetrieved Nov 28, 2026, 8:39 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-0162-2027" "Z-1893-2026" "Z-1894-2026" "Z-1895-2026" "Z-0584-2026")Retrieved Nov 28, 2026, 8:39 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:DOFRetrieved Nov 28, 2026, 8:39 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:DOFRetrieved Nov 28, 2026, 8:39 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:862.3550Retrieved Nov 28, 2026, 6:36 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil