DDR: Myoglobin, Antigen, Antiserum, Control (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 866.5680
- Review panel
- Immunology
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 460 in the last 12 months
- Median review time
- 37 daysthe latest 25 decisions
- Recall records
- 120 in the last 12 months
- Adverse event reports
- 7last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code DDR per year, by decision date, 46 since 1977. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 1 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K231832 | Access MyoglobinBeckman Coulter, Inc. | Sep 18, 202388 days |
| K123785 | DIAZYME MYOGLOBIN; ASSAY, CALIBRATOR SET, CONTROL SETDiazyme Laboratories | Jun 7, 2013179 days |
| K122599 | ADVIA CHEMISTRY MYOGLOBIN REAGENT, ADVIA CHEMISTRY MYOGLBIN CALIBRATORSiemens Healthcare Diagnostics, Inc. | Oct 3, 201237 days |
| K083260 | ELECSYS MYOGLOBIN STAT, AND ELECSYS MYOGLOBINRoche Diagnostics | Oct 16, 2009345 days |
| K080481 | ACCESS IMMUNOASSAY SYSTEMS MYOGLOBIN AND ACCESS IMMUNOASSAY SYSTEMS MYOGLOBIN CALIBRATORSBeckman Coulter, Inc. | Feb 20, 2009364 days |
| K061683 | TINA-QUANT MYOGLOBIN GEN.2 TEST SYSTEMRoche Diagnostics Corp. | Jul 21, 200636 days |
| K042924 | ARCHITECT STAT MYO IMMUNOASSAYFisher Diagnostics | Jan 24, 200594 days |
| K042982 | QUANTEX MYOGLOBINInstrumentation Laboratory CO | Dec 22, 200454 days |
| K033300 | VITROS IMMUNODIAGNOSTIC PRODUCTS MYOGLOBIN REAGENT PACK; CALIBRATORS; RANGE VERIFIERSOrtho-Clinical Diagnostics, Inc. | Nov 5, 200322 days |
| K021229 | MYOGLOBIN AND MYOGLOBIN CALIBRATORS ON THE ACCESS IMMUNOASSAY SYSTEMS, MODEL 973243, 973244Beckman Coulter, Inc. | Jun 28, 200271 days |
Recalls
Recall records for product code DDR per year, by the date the recall began. Each recalled product is one record; these 12 records belong to 10 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 1 |
| 2019 | 1 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 1 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-1476-2024Class II | Abbott LaboratoriesThe reason for the recall is the failure of calibration and quality controls due to a manufacturing issue of microparticles not meeting labeled claim. This may… | Feb 21, 2024 |
| Z-2445-2019Class II | Roche Diagnostics Operations, Inc.The firm conducted internal investigations that confirmed customer complaints of quality control (QC) imprecision and calibration failures for the Tina-quant M… | Jul 2, 2019 |
| Z-2072-2018Class II | Tosoh Bioscience IncAsfotase Alfa (Strensiq) interferes with certain Tosoh assays which potentially causes falsely increased or falsely decreased immunoassay test results, dependi… | Mar 5, 2018 |
| Z-2899-2016Class II | Siemens Healthcare Diagnostics, IncFailure of analytical sensitivity with Calibrator U kit lots ending in 63 and 64 and a drift from the internal standardization | Aug 8, 2016 |
| Z-0612-2016Class II | Siemens Healthcare Diagnostics IncStratus CS Cartridge Barcode labels may not scan on the Stratus CS systems | Nov 11, 2015 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code DDR. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 5 |
| 2018 | 3 |
| 2019 | 6 |
| 2020 | 2 |
| 2021 | 1 |
| 2022 | 1 |
| 2023 | 7 |
| 2024 | 31 |
| 2025 | 5 |
| 2026 (so far) | 7 |
Last 12 months by event type
- Deaths0
- Injuries0
- Malfunctions7
- Other or not stated0
Most reported device problems, last 12 months
- Low Test Results5
- High Test Results2
In plain English: what the newest reports describe Written by AI
Several reports describe myoglobin results that were falsely low: a retest on the same sample gave much higher values, for example 12 ng/mL first versus 40 ng/mL on retest, and 70 ng/mL versus 336.6 ng/mL. The reports say no impact on patient management was reported.
3 reports: 3008344661-2026-00143, 1823260-2025-01475, 1823260-2025-00291
Several reports describe myoglobin results that were falsely high or inconsistent with the patient's condition, such as a value of 3486 ng/L for a related troponin test in one case. In one report a sample read far above other tubes drawn at the same time, and in another a repeat run on a different analyzer disagreed with the first.
3 reports: 1823260-2026-01862, 1823260-2025-03908, 1823260-2025-04731
Several reports describe quality control fluid results that read lower than expected on the instrument, such as 110.9 ng/mL against an expected value near 134.8 ng/mL in one case. The reports say the bias could lead to inappropriate physician action if undetected on patient samples, and no patient harm was alleged.
2 reports: 3007111389-2026-00045, 3007111389-2026-00033
One report describes a 56-year-old male whose initial myoglobin result was 208.3 ug/L against a reference range of 0-140.1 ug/L, while repeat results were 139.9 and 137.2 ug/L. The report says no impact to patient management was reported.
Report: 3008344661-2026-00098
One report describes a falsely decreased myoglobin result for a 59-year-old male, 131.7 ng/mL on a new reagent versus 163.8 ng/mL on the previous reagent, with a reference range below 140 ng/mL. The report says no impact to patient management was reported.
Report: 3008344661-2026-00089
Written by AI from the narratives of the 17 newest reports that have one, received May 17, 2024 to Sep 17, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 64in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 70registered establishments that list this product code
Other codes under 21 CFR 866.5680
Product codes classified under the same regulation.
Watch DDR in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get DDR every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in DDR. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:DDRRetrieved Nov 29, 2026, 5:38 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:DDRRetrieved Nov 29, 2026, 5:38 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:DDR · count decision_dateRetrieved Nov 29, 2026, 5:38 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:DDRRetrieved Nov 29, 2026, 5:38 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:DDR · count event_date_initiatedRetrieved Nov 29, 2026, 5:38 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:DDR · count res_event_numberRetrieved Nov 29, 2026, 5:38 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-1476-2024" "Z-2445-2019" "Z-2072-2018" "Z-2899-2016" "Z-0612-2016")Retrieved Nov 29, 2026, 5:38 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:DDR AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 5:38 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:DDR AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 5:38 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:DDR AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 5:38 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:DDRRetrieved Nov 29, 2026, 5:38 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:DDRRetrieved Nov 29, 2026, 5:38 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:866.5680Retrieved Nov 29, 2026, 5:38 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.