Product codes

FDA product code

CIT: Nadh Oxidation/Nad Reduction, Ast/Sgot (FDA product code)

Follow CIT: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 862.1100
Review panel
Clinical Chemistry
Premarket route
510(k) exempt
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Not eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
1390 in the last 12 months
Median review time
48 daysthe latest 25 decisions
Recall records
130 in the last 12 months
Adverse event reports
3last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code CIT per year, by decision date, 139 since 1976. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20170
20180
20190
20200
20210
20220
20230
20240
20250
2026 (so far)0

Latest decisions

NumberDevice and applicantDecision and review time
K093883 AST/GOT 4+1 SL, AST/GOT 4+1 SL, ELICAL 2, ELITROL I, ELITROL II, MODEL ASSL-0250, ASSL-0455, CALI-0580, CONT-0080, CONT-Seppim S.A.S. Sep 20, 2010276 days
K080485 ASPARTATE AMINO TRANSFERASE (AST/SGOT) TEST SYSTEM, ASAT (GOT) FSDiasys Diagnostics Systems GmbH Dec 4, 2008286 days
K071525 ACTIVATED ASPARTATE AMINOTRANSFERASEAbbott Laboratories Mar 19, 2008289 days
K071580 AST/GOT (IFCC), PYRIDOXAL PHOSPHATE, ECAL, NORTROL AND ABTROLThermofisher Scientific OY Oct 3, 2007117 days
K071211 COBAS C 111 ANALYZERRoche Diagnostics Jul 30, 200790 days
K060318 HEPATIC ENZYMES ON ABX PENTRA 400 CLINICAL CHEMISTRY ANALYZERHoriba Abx Aug 8, 2006181 days
K023551 WIENER LAB GOT (AST) UV AA LIQUIDAWiener Laboratories Saic Dec 9, 200248 days
K013698 AST (SGOT)Jas Diagnostics, Inc. Dec 21, 200144 days
K013145 WIENER LAB. GOT (AST) UV AA. WEINER LAB. GOT (AST) UV UNITEST, MODELS 10 * 20 ML CAT. NR. 1751302 AND 20 * 2 ML CAT NR.Wiener Laboratories Saic Nov 27, 200168 days
K010488 ATAC PAK AST REAGENT KITElan Holdings, Inc. Apr 9, 200148 days

Recalls

Recall records for product code CIT per year, by the date the recall began. Each recalled product is one record; these 13 records belong to 11 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20172
20180
20190
20200
20210
20220
20231
20240
20250
2026 (so far)0

Latest recalls

Recall and classFirm and reasonBegan
Z-0199-2024Class II Randox Laboratories Ltd.RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not meeting sample re-run conditions when samples are above or below… Sep 6, 2023
Z-1651-2018Class II Siemens Healthcare Diagnostics, Inc.Various assays used to measure different chemical levels in blood, urine, or cerebrospinal fluid may interfere with the drugs sulfasalazine and/or sulfapyridin… Jun 29, 2017
Z-1657-2018Class II Siemens Healthcare Diagnostics, Inc.Various assays used to measure different chemical levels in blood, urine, or cerebrospinal fluid may interfere with the drugs sulfasalazine and/or sulfapyridin… Jun 29, 2017
Z-0006-2015Class III ELITech Clinical Systems SASSome users of ENVOY 500 AST Reagent Kit, reference 55255, for Envoy 500 systems are observing that Quality Control (QC) fails before the product expiry is reac… Aug 11, 2014
Z-1530-2014Class II Alfa Wassermann, Inc.Extremely low absorbance readings. Mar 5, 2014

Adverse event reports (MAUDE)

Medical device reports received per year that name product code CIT. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
20172
20181
20192
20201
20215
20222
20232
20248
202510
2026 (so far)3

Last 12 months by event type

  • Deaths0
  • Injuries0
  • Malfunctions3
  • Other or not stated0

Most reported device problems, last 12 months

  1. High Test Results1
  2. Non Reproducible Results1
  3. Output Problem1

In plain English: what the newest reports describe Written by AI

  • Several reports describe AST (a liver enzyme blood test) results that were questioned because they seemed wrong for the patient. Repeat tests on the same sample or on other analyzers often gave very different values, and the manufacturer or lab sometimes judged the repeat values correct.

    10 reports: 1823260-2026-01267, 1823260-2025-04868, 1823260-2025-04768, 1823260-2024-03844, 1823260-2025-00968, 1823260-2024-00580, 1823260-2024-02544, 1823260-2025-01180, 1823260-2024-01311, 1823260-2025-00091

  • Several reports describe an AST result that was far higher than repeat or other results. Some repeat tests on the same sample gave no value at all, only a reaction-problem flag, and dilution testing produced very high numbers.

    4 reports: 1823260-2025-01180, 1823260-2025-00968, 1823260-2024-03219, 1823260-2024-01311

  • Some reports describe AST results carrying data flags or alarms from the analyzer, which are warnings that a result may not be reliable. In some cases the lab could not provide the underlying numbers.

    4 reports: 1823260-2024-02574, 1823260-2025-04173, 1823260-2025-01362, 1823260-2025-04768

  • One report says the manufacturer received similar concerns and reported that the AST test showed a positive bias, meaning higher results, with a new test formulation and with samples drawn in light green lithium heparin tubes. The report states no harm to the patient and was filed out of caution.

    Report: 22663494

  • One report describes AST and ALT results that were questioned after a patient had received red blood cells, which the customer suspected might cause interference. Diluted repeat results gave very different and sometimes negative or implausible values, so a new sample was collected.

    Report: 1823260-2025-04768

Written by AI from the narratives of the 20 newest reports that have one, received Feb 28, 2024 to May 6, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
140in the FDA’s Global Unique Device Identification Database
Establishment listings
94registered establishments that list this product code

Product codes classified under the same regulation.

Watch CIT in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get CIT every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in CIT. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.