FDA product code
CHP: Radioimmunoassay, Estradiol (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class I
- Regulation
- 21 CFR 862.1260
- Review panel
- Clinical Chemistry
- Premarket route
- 510(k) exempt
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 500 in the last 12 months
- Median review time
- 72 daysthe latest 25 decisions
- Recall records
- 310 in the last 12 months
- Adverse event reports
- 64last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code CHP per year, by decision date, 50 since 1979. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K041349 | LUMINESCENT IMMUNOASSAY KIT FOR THE DETECTION OF ESTRADIOL IN SALIVA AND SERUMIbl GmbH | Sep 24, 2004127 days |
| K973901 | 17 B-ESTRADIOL (ELISA)Kmi Diagnostics, Inc. | Nov 24, 199741 days |
| K973743 | ACCESS ESTRADIOL REAGENTS ON THE ACCESS IMMUNOASSAY ANALYZERBeckman Instruments, Inc. | Oct 31, 199730 days |
| K970386 | ACS ESTADIOL-6 IIChiron Diagnostics Corp. | Jun 4, 1997121 days |
| K970643 | VITROS IMMUNODIAGNOSTICS PRODUCTS ESTRADIOL REAGENT PACK(GEM. 1050)/ESTRADIOL CALIBRATORS (GEM.C050)Johnson & Johnson Clinical Diagnostics, Inc. | Mar 18, 199726 days |
| K970126 | ACCESS ESTRADIOL REAGENTS ON THE ACCESS IMMUNOASSAY ANALYZER (33550,33540,33545)Bio-Rad Laboratories, Inc. | Feb 25, 199742 days |
| K965109 | ELECSYS ESTRADIOL ASSAYBoehringer Mannheim Corp. | Feb 18, 199760 days |
| K955647 | VIDAS ESTRADIOL II (E2II) ASSAYBiomerieux Vitek, Inc. | Feb 6, 199656 days |
| K954697 | CIBA CORNING ACS ESTRADIOL-6 ASSAYCiba Corning Diagnostics Corp. | Dec 13, 199562 days |
| K953605 | ULTRA-SENSITIVE ESTRADIOLDiagnostic Systems Laboratories, Inc. | Oct 2, 199561 days |
Recalls
Recall records for product code CHP per year, by the date the recall began. Each recalled product is one record; these 31 records belong to 19 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 3 |
| 2018 | 4 |
| 2019 | 2 |
| 2020 | 2 |
| 2021 | 1 |
| 2022 | 7 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-0867-2023Class II | Biomerieux IncAs the temperature and time out of range was exceeded, product performances cannot be guaranteed | Nov 9, 2022 |
| Z-1020-2022Class II | Siemens Healthcare Diagnostics, Inc.Falsely Elevated Results Observed with Plasma Specimens, could lead a clinician to misinterpret patient results | Mar 3, 2022 |
| Z-1021-2022Class II | Siemens Healthcare Diagnostics, Inc.Falsely Elevated Results Observed with Plasma Specimens, could lead a clinician to misinterpret patient results | Mar 3, 2022 |
| Z-1022-2022Class II | Siemens Healthcare Diagnostics, Inc.Falsely Elevated Results Observed with Plasma Specimens, could lead a clinician to misinterpret patient results | Mar 3, 2022 |
| Z-1023-2022Class II | Siemens Healthcare Diagnostics, Inc.Falsely Elevated Results Observed with Plasma Specimens, could lead a clinician to misinterpret patient results | Mar 3, 2022 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code CHP. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 39 |
| 2018 | 38 |
| 2019 | 36 |
| 2020 | 135 |
| 2021 | 43 |
| 2022 | 61 |
| 2023 | 60 |
| 2024 | 65 |
| 2025 | 86 |
| 2026 (so far) | 56 |
Last 12 months by event type
- Deaths0
- Injuries2
- Malfunctions62
- Other or not stated0
Most reported device problems, last 12 months
- High Test Results41
- Non Reproducible Results31
- Low Test Results27
- Incorrect Measurement11
- Incorrect, Inadequate or Imprecise Result or Readings5
- Chemical Problem2
In plain English: what the newest reports describe Written by AI
Several reports describe estradiol results that were far higher or lower than repeat results on the same patient sample, and the customer repeated the tests because the first values did not fit the patient. In most of these cases the repeat value was judged correct, but the reports do not explain the cause.
8 reports: 1823260-2026-02734, 1823260-2026-02339, 1823260-2026-02044, 1823260-2026-02277, 1823260-2026-02043, 1823260-2026-02198, 1823260-2026-03501, 1823260-2026-03240
Several reports describe results that appeared to contradict other tests or the patient's condition, such as estradiol values in a patient with no ovaries. One report found that a sample sent to another laboratory with a different method gave a result that matched the patient's condition.
3 reports: 1823260-2026-02259, 1823260-2026-02786, 1823260-2026-03814
A report with reagent lots shows the same patient samples gave very different estradiol values depending on which reagent lot was used. The reporter could not say which lot produced which result, and it is unclear which results were believed correct.
2 reports: 1823260-2026-02198, 1823260-2026-04454
One report describes fertility patients whose results were reported without a flag and may have affected treatment. One report says two patients were hospitalized for ovarian hyperstimulation, a condition where the ovaries swell after fertility drugs, and the customer suspects an interfering substance in the blood, but the report does not confirm this link.
Report: 1823260-2026-02074
One report describes a very high estradiol result that was repeated as a much lower value on a second run. The reporter had concerns about how the sample was handled, and the manufacturer's investigation found frequent sample clot and short sample alarms, which it said pointed to poor handling before testing at the customer site.
Report: 1823260-2026-03734
Written by AI from the narratives of the 20 newest reports that have one, received May 13, 2026 to Oct 29, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 88in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 59registered establishments that list this product code
Other codes under 21 CFR 862.1260
Product codes classified under the same regulation.
Watch CHP in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get CHP every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in CHP. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:CHPRetrieved Nov 29, 2026, 5:17 AM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:CHPRetrieved Nov 29, 2026, 6:09 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:CHP · count decision_dateRetrieved Nov 29, 2026, 5:17 AM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:CHPRetrieved Nov 29, 2026, 6:09 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:CHP · count event_date_initiatedRetrieved Nov 29, 2026, 5:17 AM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:CHP · count res_event_numberRetrieved Nov 29, 2026, 5:17 AM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-0867-2023" "Z-1020-2022" "Z-1021-2022" "Z-1022-2022" "Z-1023-2022")Retrieved Nov 29, 2026, 6:09 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:CHP AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 6:09 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:CHP AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 6:09 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:CHP AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 5:17 AM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:CHPRetrieved Nov 29, 2026, 5:17 AM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:CHPRetrieved Nov 29, 2026, 5:17 AM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:862.1260Retrieved Nov 29, 2026, 5:48 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil