FDA product code
CGZ: Electrode, Ion-Specific, Chloride (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 862.1170
- Review panel
- Clinical Chemistry
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Eligible
- Summary malfunction reporting
- Ineligible
- 510(k) and De Novo decisions
- 380 in the last 12 months
- Median review time
- 61 daysthe latest 25 decisions
- Recall records
- 110 in the last 12 months
- Adverse event reports
- 2last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code CGZ per year, by decision date, 38 since 1979. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 1 |
| 2018 | 1 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K182072 | VITROS Chemistry Products Cl-Slides, VITROS Chemistry Products ECO2 Slides, VITROS Chemistry Products GLU SlidesOrtho-Clinical Diagnostics, Inc. | Oct 30, 201890 days |
| K170320 | Alinity c ICT Sample DiluentAbbott Laboratories | Oct 24, 2017265 days |
| K162020 | VITROS Chemistry Products Cl- SlidesOrtho Clinical Diagnostics | Dec 16, 2016148 days |
| K082099 | VITROS CHEMISTRY PRODUCTS CI-DT SLIDES; VITROS CHEMISTRY PRODUCTS DT CALIBRATOR KIT, MODELS CI-DT SLIDES:131 4905Ortho-Clinical Diagnostics, Inc. | Aug 21, 200827 days |
| K070889 | AUTODELFIA NEONATAL IRT L KIT, MODEL B022-112Wallac OY | Apr 14, 2008381 days |
| K070531 | ISE MODULE AND BAR CODE READER ADDITION FOR THE DATAPRO CLINICAL CHEMISTRY ANALYZERThermo Fisher Scientific | Sep 4, 2007193 days |
| K060108 | COBAS INTEGRA CHLORIDE ELCTRODE GEN.2Roche Diagnostics Corp. | Feb 16, 200634 days |
| K052027 | OPTI LION ELECTROLYTE ANALYZER, MODEL GD7200Osmetech, Inc. | Sep 29, 200564 days |
| K020148 | pHoenix ISE Reagents FOR Olympus AU Chemistry SystemsPhoenix Diagnostics, Inc. | Feb 4, 200219 days |
| K013451 | PHOENOIX ISE REAGENTS FOR ROCHE HITACHI, MODELS 700/900 SERIESPhoenix Diagnostics, Inc. | Dec 18, 200161 days |
Recalls
Recall records for product code CGZ per year, by the date the recall began. Each recalled product is one record; these 11 records belong to 10 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 3 |
| 2019 | 2 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-0267-2020Class II | Ortho-Clinical DiagnosticsUsers may experience intermittent slide dispense issues with some cartridges due to the anti-backup platen failing to advance inside the cartridge. | Sep 10, 2019 |
| Z-0268-2020Class II | Ortho-Clinical DiagnosticsUsers may experience intermittent slide dispense issues with some cartridges due to the anti-backup platen failing to advance inside the cartridge. | Sep 10, 2019 |
| Z-1296-2019Class II | Abbott Laboratories, IncThere is a potential to generate falsely elevated serum or plasma chloride results when using the product. | Aug 28, 2018 |
| Z-3043-2018Class III | Ortho-Clinical DiagnosticsThe slides do not meet the maximum allowable interference (MAI) claims for triglyceride interference when using serum samples. Triglyceride interference on the… | Jul 11, 2018 |
| Z-3184-2018Class III | Ortho-Clinical DiagnosticsThe slides failed to meet current claims for Limit of Blank (LoB), Limit of Detection (LoD) and Limit of Quantification (LoQ) for urine samples. The measuring… | Jul 11, 2018 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code CGZ. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 2 |
| 2018 | 19 |
| 2019 | 3 |
| 2020 | 3 |
| 2021 | 0 |
| 2022 | 5 |
| 2023 | 9 |
| 2024 | 6 |
| 2025 | 5 |
| 2026 (so far) | 2 |
Last 12 months by event type
- Deaths0
- Injuries0
- Malfunctions2
- Other or not stated0
Most reported device problems, last 12 months
- High Test Results1
- Low Test Results1
In plain English: what the newest reports describe Written by AI
Some chloride results were questioned because they were accompanied by a negative anion gap, a calculated value that can signal an electrolyte imbalance. In these reports the customer repeated the test, and the repeat chloride value was judged correct.
2 reports: 1823260-2025-04600, 1823260-2023-03486
In one report, a chloride result of 88 mmol/L was low and prompted a rerun of quality control, which was out of range until the customer troubleshot the analyzer. The repeat result of 101 mmol/L was judged correct.
Report: 1823260-2023-01858
In one report, the same sample gave a chloride result of 138 mmol/L initially and then 119, 126 and 239 mmol/L on three repeats. The customer did not know which result was correct, and the results were not reported outside the laboratory.
Report: 1823260-2023-02412
One report describes a positive bias in chloride results on an integrated sensor since 2021, but states there were no discordant patient results, no delays, and no known patient harm. It also says the manufacturer's data showed no effect on patient care.
Report: 2432235-2025-00034
Written by AI from the narratives of the 20 newest reports that have one, received Jun 5, 2023 to Jan 7, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 480in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 260registered establishments that list this product code
Other codes under 21 CFR 862.1170
Product codes classified under the same regulation.
Watch CGZ in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get CGZ every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in CGZ. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:CGZRetrieved Nov 28, 2026, 11:28 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:CGZRetrieved Nov 29, 2026, 7:43 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:CGZ · count decision_dateRetrieved Nov 28, 2026, 11:28 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:CGZRetrieved Nov 29, 2026, 7:43 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:CGZ · count event_date_initiatedRetrieved Nov 28, 2026, 11:28 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:CGZ · count res_event_numberRetrieved Nov 28, 2026, 11:28 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-0267-2020" "Z-0268-2020" "Z-1296-2019" "Z-3043-2018" "Z-3184-2018")Retrieved Nov 29, 2026, 7:43 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:CGZ AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 8:07 AM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:CGZ AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 8:07 AM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:CGZ AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 8:07 AM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:CGZRetrieved Nov 28, 2026, 11:28 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:CGZRetrieved Nov 28, 2026, 11:28 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:862.1170Retrieved Nov 28, 2026, 11:28 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil