CGZ: Electrode, Ion-Specific, Chloride (FDA product code)

Follow CGZ: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 862.1170
Review panel
Clinical Chemistry
Premarket route
510(k)
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Eligible
Summary malfunction reporting
Ineligible
510(k) and De Novo decisions
380 in the last 12 months
Median review time
61 daysthe latest 25 decisions
Recall records
110 in the last 12 months
Adverse event reports
2last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code CGZ per year, by decision date, 38 since 1979. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20171
20181
20190
20200
20210
20220
20230
20240
20250
2026 (so far)0

Latest decisions

NumberDevice and applicantDecision and review time
K182072 VITROS Chemistry Products Cl-Slides, VITROS Chemistry Products ECO2 Slides, VITROS Chemistry Products GLU SlidesOrtho-Clinical Diagnostics, Inc. Oct 30, 201890 days
K170320 Alinity c ICT Sample DiluentAbbott Laboratories Oct 24, 2017265 days
K162020 VITROS Chemistry Products Cl- SlidesOrtho Clinical Diagnostics Dec 16, 2016148 days
K082099 VITROS CHEMISTRY PRODUCTS CI-DT SLIDES; VITROS CHEMISTRY PRODUCTS DT CALIBRATOR KIT, MODELS CI-DT SLIDES:131 4905Ortho-Clinical Diagnostics, Inc. Aug 21, 200827 days
K070889 AUTODELFIA NEONATAL IRT L KIT, MODEL B022-112Wallac OY Apr 14, 2008381 days
K070531 ISE MODULE AND BAR CODE READER ADDITION FOR THE DATAPRO CLINICAL CHEMISTRY ANALYZERThermo Fisher Scientific Sep 4, 2007193 days
K060108 COBAS INTEGRA CHLORIDE ELCTRODE GEN.2Roche Diagnostics Corp. Feb 16, 200634 days
K052027 OPTI LION ELECTROLYTE ANALYZER, MODEL GD7200Osmetech, Inc. Sep 29, 200564 days
K020148 pHoenix ISE Reagents FOR Olympus AU Chemistry SystemsPhoenix Diagnostics, Inc. Feb 4, 200219 days
K013451 PHOENOIX ISE REAGENTS FOR ROCHE HITACHI, MODELS 700/900 SERIESPhoenix Diagnostics, Inc. Dec 18, 200161 days

Recalls

Recall records for product code CGZ per year, by the date the recall began. Each recalled product is one record; these 11 records belong to 10 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20170
20183
20192
20200
20210
20220
20230
20240
20250
2026 (so far)0

Latest recalls

Recall and classFirm and reasonBegan
Z-0267-2020Class II Ortho-Clinical DiagnosticsUsers may experience intermittent slide dispense issues with some cartridges due to the anti-backup platen failing to advance inside the cartridge. Sep 10, 2019
Z-0268-2020Class II Ortho-Clinical DiagnosticsUsers may experience intermittent slide dispense issues with some cartridges due to the anti-backup platen failing to advance inside the cartridge. Sep 10, 2019
Z-1296-2019Class II Abbott Laboratories, IncThere is a potential to generate falsely elevated serum or plasma chloride results when using the product. Aug 28, 2018
Z-3043-2018Class III Ortho-Clinical DiagnosticsThe slides do not meet the maximum allowable interference (MAI) claims for triglyceride interference when using serum samples. Triglyceride interference on the… Jul 11, 2018
Z-3184-2018Class III Ortho-Clinical DiagnosticsThe slides failed to meet current claims for Limit of Blank (LoB), Limit of Detection (LoD) and Limit of Quantification (LoQ) for urine samples. The measuring… Jul 11, 2018

Adverse event reports (MAUDE)

Medical device reports received per year that name product code CGZ. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
20172
201819
20193
20203
20210
20225
20239
20246
20255
2026 (so far)2

Last 12 months by event type

  • Deaths0
  • Injuries0
  • Malfunctions2
  • Other or not stated0

Most reported device problems, last 12 months

  1. High Test Results1
  2. Low Test Results1

In plain English: what the newest reports describe Written by AI

  • Some chloride results were questioned because they were accompanied by a negative anion gap, a calculated value that can signal an electrolyte imbalance. In these reports the customer repeated the test, and the repeat chloride value was judged correct.

    2 reports: 1823260-2025-04600, 1823260-2023-03486

  • In one report, a chloride result of 88 mmol/L was low and prompted a rerun of quality control, which was out of range until the customer troubleshot the analyzer. The repeat result of 101 mmol/L was judged correct.

    Report: 1823260-2023-01858

  • In one report, the same sample gave a chloride result of 138 mmol/L initially and then 119, 126 and 239 mmol/L on three repeats. The customer did not know which result was correct, and the results were not reported outside the laboratory.

    Report: 1823260-2023-02412

  • One report describes a positive bias in chloride results on an integrated sensor since 2021, but states there were no discordant patient results, no delays, and no known patient harm. It also says the manufacturer's data showed no effect on patient care.

    Report: 2432235-2025-00034

Written by AI from the narratives of the 20 newest reports that have one, received Jun 5, 2023 to Jan 7, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
480in the FDA’s Global Unique Device Identification Database
Establishment listings
260registered establishments that list this product code

Watch CGZ in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get CGZ every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in CGZ. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.