FDA product code
CDQ: Urease And Glutamic Dehydrogenase, Urea Nitrogen (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 862.1770
- Review panel
- Clinical Chemistry
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 1170 in the last 12 months
- Median review time
- 61 daysthe latest 25 decisions
- Recall records
- 60 in the last 12 months
- Adverse event reports
- 11last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code CDQ per year, by decision date, 117 since 1976. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 1 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K203771 | Urea Nitrogen2Abbott Ireland Diagnostics Division | May 31, 2022524 days |
| K140690 | BS-480 CHEMISTRY ANALYZER, BS-490 CHEMISTRY ANALYZER, CLC720I CHEMISTRY ANALYZERShenzhen Mindray Bio-Medical Electronics Co., Ltd. | Sep 2, 2014167 days |
| K110675 | EASYRA UREA NITROGEN AND CREATININE REAGENTSMedica Corp. | Apr 21, 201142 days |
| K073295 | UREA/BUN, SCAL, NORTROL, ABTROLThermofisher Scientific OY | May 29, 2008188 days |
| K070146 | ABX PENTRA GLUCOSE HK CP, UREA CP, URIC ACID CPHoriba Abx | Oct 12, 2007269 days |
| K060120 | TECO CX3 REAGENT SET FOR SYNCHRON CX SYSTEMTeco Diagnostics | Mar 27, 200669 days |
| K041051 | RAICHEM BUN RATE REAGENT (LIQUID)Hemagen Diagnostics, Inc. | Jun 30, 200472 days |
| K024182 | PRECISION SYSTEMS ANALETTE CHEMISTRY ANALYZER STANBIO LABORATORY REAGENTSPrecision Systems, Inc. | Mar 3, 200374 days |
| K021385 | ATAC PAK BUN REAGENTElan Diagnostics | Jul 12, 200271 days |
| K012649 | RAICHEM BUN RATE REAGENTHemagen Diagnostics, Inc. | Oct 15, 200163 days |
Recalls
Recall records for product code CDQ per year, by the date the recall began. Each recalled product is one record; these 6 records belong to 6 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 1 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 2 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-2666-2023Class II | Siemens Healthcare Diagnostics, Inc.There is a potential for a positive bias on Quality Control (QC) and patient sample results when using some IRON_2 reagent wedge pairs. When the issue occurs,… | Aug 8, 2023 |
| Z-2362-2023Class II | Randox Laboratories Ltd.Carryover from the Direct LDL-cholesterol assay on the Rx instruments may result in a negative bias, up to -11%, in quality control and patient samples, in ure… | Jun 13, 2023 |
| Z-2062-2018Class II | Siemens Healthcare Diagnostics, Inc.Urea Nitrogen reagent kit demonstrated an increased incidence of Calibration Failures. Results cannot be generated by the system when calibration fails. | Mar 27, 2018 |
| Z-1622-2016Class II | Siemens Healthcare Diagnostics, Inc.Siemens Healthcare Diagnostics has determined that Dimension Vista Blood Urea Nitrogen (BUN) lots 15215AE, 15243BB, 15264BA, 15299BB, 15300BA and 15341AC may e… | Feb 16, 2016 |
| Z-0189-2012 | Abbott Laboratories, IncThe reagents are being recalled due to particulate matter in some of the reagent cartridges. The particulate matter has been identified as Aureobasidium fungus… | Oct 3, 2011 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code CDQ. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 3 |
| 2021 | 1 |
| 2022 | 1 |
| 2023 | 2 |
| 2024 | 4 |
| 2025 | 6 |
| 2026 (so far) | 11 |
Last 12 months by event type
- Deaths0
- Injuries0
- Malfunctions11
- Other or not stated0
Most reported device problems, last 12 months
- Low Test Results7
- Non Reproducible Results5
- High Test Results1
- Incorrect, Inadequate or Imprecise Result or Readings1
- Output Problem1
In plain English: what the newest reports describe Written by AI
Many reports describe blood urea nitrogen (BUN, a kidney function test) giving unexpectedly low results that were later repeated at much higher values, usually on another analyzer or by rerunning the sample, with the repeat judged correct.
10 reports: 1823260-2026-04140, 1823260-2026-03332, 1823260-2026-01521, 1823260-2026-01502, 1823260-2026-00187, 1823260-2025-04189, 1823260-2025-01128, 1823260-2025-00295, 1823260-2025-00208, 1823260-2024-02176
In some reports the analyzer attached a data flag to a result and then automatically reran the sample, and the rerun value differed greatly from the first. In one of these reports the rerun used a dilution step.
3 reports: 1823260-2025-02388, 1823260-2023-01389, 1823260-2025-00295
Two reports describe the opposite pattern, where high first results were repeated at much lower values. In one, the manufacturer's repeat was judged correct, and in the other the reporter said the first result was reported outside the laboratory before the problem was noticed.
2 reports: 1823260-2025-02388, 1823260-2023-01389
Two reports involved urine samples tested on cobas 8000 modules, where the first and repeat results differed widely. In one, the first result was not reported because it did not match the patient's clinical diagnosis.
2 reports: 1823260-2024-03503, 1823260-2024-02742
One report describes a result that was repeated on a second module after it did not match the patient's creatinine result, a related kidney test. The first module gave 17 mg/dL and the second gave 134 mg/dL, and the second value was judged correct.
Report: 1823260-2026-00859
Written by AI from the narratives of the 20 newest reports that have one, received Apr 25, 2023 to Oct 7, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 108in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 62registered establishments that list this product code
Other codes under 21 CFR 862.1770
Product codes classified under the same regulation.
Watch CDQ in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get CDQ every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in CDQ. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:CDQRetrieved Nov 29, 2026, 5:51 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:CDQRetrieved Nov 29, 2026, 5:51 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:CDQ · count decision_dateRetrieved Nov 29, 2026, 5:51 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:CDQRetrieved Nov 29, 2026, 5:51 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:CDQ · count event_date_initiatedRetrieved Nov 29, 2026, 5:51 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:CDQ · count res_event_numberRetrieved Nov 29, 2026, 5:51 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-2666-2023" "Z-2362-2023" "Z-2062-2018" "Z-1622-2016" "Z-0189-2012")Retrieved Nov 29, 2026, 5:51 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:CDQ AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 5:51 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:CDQ AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 5:51 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:CDQ AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 5:51 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:CDQRetrieved Nov 29, 2026, 5:51 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:CDQRetrieved Nov 29, 2026, 5:51 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:862.1770Retrieved Nov 28, 2026, 9:21 PM UTC. FDA data updated Nov 16, 2026.