FDA product code
CDP: Radioimmunoassay, Total Triiodothyronine (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 862.1710
- Review panel
- Clinical Chemistry
- Premarket route
- 510(k) exempt
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 1440 in the last 12 months
- Median review time
- 48 daysthe latest 25 decisions
- Recall records
- 200 in the last 12 months
- Adverse event reports
- 29last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code CDP per year, by decision date, 144 since 1976. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K001607 | REPROBEAD T3 ENZYME IMMUNOASSAY REAGENT KIT (CATALOG# 104)Repromedix Corp. | Oct 5, 2000134 days |
| K991692 | CHIRON DIAGNOSTICS ACS: 180 FT3Chiron Diagnostics Corp. | Jun 18, 199931 days |
| K991311 | TOTAL T3 ENZYME IMMUNOASSAY TEST KIT, MODEL BC-1005Biocheck, Inc. | Jun 2, 199947 days |
| K991252 | AIA-PACK FT3 ASSAYTosoh Medics, Inc. | May 11, 199928 days |
| K981824 | OPTICOAT T3 EIA KITBiotecx Laboratories, Inc. | Dec 22, 1998214 days |
| K983434 | ABBOTT ARCHITECT TOTAL T3Abbott Laboratories | Nov 20, 199852 days |
| K983439 | ABBOTT ARCHITECT FREE T3Abbott Laboratories | Nov 20, 199852 days |
| K974027 | AURAFLEX FT3 150 TEST PACK, AURAFLEX FT3 CALIBRATOR PACKAlfa Biotech (Uk) , Ltd. | Feb 2, 1998102 days |
| K974634 | IMMULITE FREE T3Diagnostic Products Corp. | Jan 2, 199821 days |
| K973966 | ACCESS TOTAL T3 REAGENTS ON THE ACCESS IMMUNOASSAY ANALYZERBeckman Instruments, Inc. | Oct 31, 199714 days |
Recalls
Recall records for product code CDP per year, by the date the recall began. Each recalled product is one record; these 20 records belong to 16 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 2 |
| 2018 | 5 |
| 2019 | 2 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 1 |
| 2024 | 1 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-1818-2024Class II | QUIDELORTHOT3 Reagent Packs used with Free T3 Calibrators may produce positively biased results for patient and quality control samples. The positive bias may cause erron… | Apr 9, 2024 |
| Z-1419-2023Class II | Beckman Coulter, Inc.The reagent lot produces elevated results (approximately >30%) in some (15-20%)patient samples that do not correlate with the results of other thyroid markers… | Mar 15, 2023 |
| Z-2238-2020Class II | Beckman Coulter Inc.The recalling firm characterized the potential interference effects of biotin (up to 1,200 ng/mL) on all biotin-mediated Access assays. The following six assay… | Apr 30, 2019 |
| Z-2239-2020Class II | Beckman Coulter Inc.The recalling firm characterized the potential interference effects of biotin (up to 1,200 ng/mL) on all biotin-mediated Access assays. The following six assay… | Apr 30, 2019 |
| Z-0673-2019Class II | Tosoh Bioscience IncST AIA-PACK FT3 (FT3: Free Triiodothyronine) is designed for IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of Free Triiodothyronine (FT3) in hu… | Nov 30, 2018 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code CDP. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 151 |
| 2018 | 365 |
| 2019 | 343 |
| 2020 | 209 |
| 2021 | 120 |
| 2022 | 123 |
| 2023 | 131 |
| 2024 | 41 |
| 2025 | 25 |
| 2026 (so far) | 27 |
Last 12 months by event type
- Deaths0
- Injuries0
- Malfunctions19
- Other or not stated10
Most reported device problems, last 12 months
- High Test Results19
- Low Test Results6
- Non Reproducible Results6
- Incorrect, Inadequate or Imprecise Result or Readings5
- Incorrect Measurement2
- Output Problem2
In plain English: what the newest reports describe Written by AI
Most reports only state that results were questioned after comparison with another analyzer or method. They give no values, explain nothing about what was wrong, and do not say whether any patient was affected.
12 reports: 1823260-2026-03074, 1823260-2026-01620, 1823260-2026-01636, 1823260-2026-01642, 1823260-2026-01643, 1823260-2026-01644, 1823260-2026-01640, 1823260-2026-01648, 1823260-2026-00280, 1823260-2026-00277, 1823260-2026-00282, 1823260-2025-04312
Several reports suspect that an interfering substance in the patient sample, rather than the test itself, may have affected results. The reports do not identify the substance or confirm it, and one says the results did not match the patient's clinical picture.
3 reports: 1823260-2026-02234, 1823260-2026-01628, 1823260-2025-05274
Two reports describe free thyroid hormone results that differed from repeat tests elsewhere. One showed free T4 of 1.72 ng/dL against 1.03 ng/dL at another lab, and free T3 of 3.98 pg/mL against 2.81 pg/mL. The other showed free T3 of 7.78 pg/mL at another lab against 3.14 pg/mL on a different analyzer.
2 reports: 1823260-2025-05274, 1823260-2026-01601
Depressed thyroid hormone results for one patient were higher on alternate methods. A later sample from the same patient gave similar results, and the physician questioned them. The report says no patient intervention or adverse health consequences are known.
Report: 1219913-2026-00139
One report describes discrepant results for a single patient sample run on many different tests, including vitamin D, testosterone, cortisol, prolactin and others. The narrative does not give the values, and it ends before the details are complete.
Report: 1823260-2025-05281
Written by AI from the narratives of the 18 newest reports that have one, received Oct 23, 2025 to Jul 29, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 123in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 79registered establishments that list this product code
Watch CDP in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get CDP every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in CDP. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:CDPRetrieved Nov 29, 2026, 7:51 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:CDPRetrieved Nov 29, 2026, 7:51 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:CDP · count decision_dateRetrieved Nov 29, 2026, 7:51 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:CDPRetrieved Nov 29, 2026, 7:51 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:CDP · count event_date_initiatedRetrieved Nov 29, 2026, 7:51 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:CDP · count res_event_numberRetrieved Nov 29, 2026, 7:51 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-1818-2024" "Z-1419-2023" "Z-2238-2020" "Z-2239-2020" "Z-0673-2019")Retrieved Nov 29, 2026, 7:51 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:CDP AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 7:51 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:CDP AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 7:51 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:CDP AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 7:51 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:CDPRetrieved Nov 29, 2026, 7:51 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:CDPRetrieved Nov 29, 2026, 7:51 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:862.1710Retrieved Nov 29, 2026, 7:51 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil