FDA product code
CBL: Absorbent, Carbon-Dioxide (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class I
- Regulation
- 21 CFR 868.5300
- Review panel
- Anesthesiology
- Premarket route
- 510(k) exempt
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 50 in the last 12 months
- Median review time
- 64 daysthe latest 5 decisions
- Recall records
- 70 in the last 12 months
- Adverse event reports
- 21last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code CBL per year, by decision date, 5 since 1981. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K921803 | SOFNOLIME CARBON DIOXIDE ABSORBENTMolecular Products , Ltd. | Oct 15, 1992183 days |
| K897171 | DISPOSABLE CO2 ABSORBER/EGibeck, Inc. | Mar 13, 199075 days |
| K860147 | CARBON DIOXIDE ABSORBENTSDentsply Intl. | Feb 28, 198644 days |
| K853571 | MEDIPART SODA LIME MP-3522Medipart Jerry Alexander | Oct 29, 198564 days |
| K811260 | BARALYME CO2 ABSORBENTAllied Healthcare Products, Inc. | May 27, 198121 days |
Recalls
Recall records for product code CBL per year, by the date the recall began. Each recalled product is one record; these 7 records belong to 5 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 1 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-2433-2021Class I | Armstrong Medical Services LimitedSome devices may be associated with high and unexpected resistance to gas flow during clinical use, impacting or preventing adequate ventilation of an anaesthe… | Aug 5, 2021 |
| Z-0560-2011 | Darex Container Products Div of W.R. Grace & Co.Pre-Pak units of SODASORB CO2 absorbent manufactured prior to August 2010 may have missing or illegible expiration dates and/or batch numbers. | Oct 20, 2010 |
| Z-0561-2011 | Darex Container Products Div of W.R. Grace & Co.Pre-Pak units of SODASORB CO2 absorbent manufactured prior to August 2010 may have missing or illegible expiration dates and/or batch numbers. | Oct 20, 2010 |
| Z-2152-2011 | Allied Healthcare Products IncThe granules do not absorb as long as expected in the anesthesia machines | Aug 12, 2010 |
| Recall | Darex Container Products Div of W.R. Grace & Co.Both lots were found to be out of specification due to an elevated level of small particles (fines). | Mar 31, 2005 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code CBL. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 0 |
| 2018 | 4 |
| 2019 | 4 |
| 2020 | 10 |
| 2021 | 22 |
| 2022 | 20 |
| 2023 | 9 |
| 2024 | 6 |
| 2025 | 17 |
| 2026 (so far) | 21 |
Last 12 months by event type
- Deaths0
- Injuries3
- Malfunctions18
- Other or not stated0
Most reported device problems, last 12 months
- Energy Output Problem6
- Fitting Problem6
- Leak/Splash3
- Gas/Air Leak2
- No Flow2
- Failure to Seal1
In plain English: what the newest reports describe Written by AI
Reports describe the absorbent canister not sealing properly in the machine, which causes a circuit leak. Some say it was hard to seal during canister changes, and one says the canister fit but failed a leak test. In several cases staff had to bag the patient by hand, and one report says the machine did not register carbon dioxide from the patient.
6 reports: 1417592-2026-40682, 1417592-2026-05826, 1924066-2026-00004, 1924066-2025-00081, 1924066-2025-00082, 1924066-2025-00083
Reports describe ventilation stopping during a case and the patient going into respiratory arrest. The damaged absorber container was cited, and the patient recovered in each report. One report says the manufacturer of the damaged disposable container was not the device maker.
3 reports: MW5189851, MW5189853, 3010838917-2026-00006
Two reports describe leaks in prefilled absorber canisters from one lot, found at two hospitals. The reports say leaks were found in all customer batches, and no patient harm was reported. The manufacturer could not investigate because the devices were not returned.
2 reports: 3002226076-2026-01568, 3002226076-2026-01569
A report says the manufacturer's team found cracked canisters from one lot. The report does not describe any effect on a patient.
Report: MW5190462
A report says the flow sensors and their oxygen rings fell out of the machine along with the absorbent canisters. Staff had to ventilate the patient by hand until a technician put the part back.
Report: 1924066-2026-00004
Written by AI from the narratives of the 17 newest reports that have one, received Nov 14, 2025 to Sep 15, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 134in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 25registered establishments that list this product code
Watch CBL in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get CBL every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in CBL. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:CBLRetrieved Nov 29, 2026, 11:37 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:CBLRetrieved Nov 29, 2026, 11:37 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:CBL · count decision_dateRetrieved Nov 29, 2026, 11:37 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:CBLRetrieved Nov 29, 2026, 11:37 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:CBL · count event_date_initiatedRetrieved Nov 29, 2026, 11:37 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:CBL · count res_event_numberRetrieved Nov 29, 2026, 11:37 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-2433-2021" "Z-0560-2011" "Z-0561-2011" "Z-2152-2011")Retrieved Nov 29, 2026, 11:37 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:CBL AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 11:37 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:CBL AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 11:37 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:CBL AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 11:37 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:CBLRetrieved Nov 29, 2026, 11:37 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:CBLRetrieved Nov 29, 2026, 11:37 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:868.5300Retrieved Nov 29, 2026, 11:37 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil