Product codes

FDA product code

CBL: Absorbent, Carbon-Dioxide (FDA product code)

Follow CBL: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class I
Regulation
21 CFR 868.5300
Review panel
Anesthesiology
Premarket route
510(k) exempt
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Not eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
50 in the last 12 months
Median review time
64 daysthe latest 5 decisions
Recall records
70 in the last 12 months
Adverse event reports
21last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code CBL per year, by decision date, 5 since 1981. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20170
20180
20190
20200
20210
20220
20230
20240
20250
2026 (so far)0

Latest decisions

NumberDevice and applicantDecision and review time
K921803 SOFNOLIME CARBON DIOXIDE ABSORBENTMolecular Products , Ltd. Oct 15, 1992183 days
K897171 DISPOSABLE CO2 ABSORBER/EGibeck, Inc. Mar 13, 199075 days
K860147 CARBON DIOXIDE ABSORBENTSDentsply Intl. Feb 28, 198644 days
K853571 MEDIPART SODA LIME MP-3522Medipart Jerry Alexander Oct 29, 198564 days
K811260 BARALYME CO2 ABSORBENTAllied Healthcare Products, Inc. May 27, 198121 days

Recalls

Recall records for product code CBL per year, by the date the recall began. Each recalled product is one record; these 7 records belong to 5 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20170
20180
20190
20200
20211
20220
20230
20240
20250
2026 (so far)0

Latest recalls

Recall and classFirm and reasonBegan
Z-2433-2021Class I Armstrong Medical Services LimitedSome devices may be associated with high and unexpected resistance to gas flow during clinical use, impacting or preventing adequate ventilation of an anaesthe… Aug 5, 2021
Z-0560-2011 Darex Container Products Div of W.R. Grace & Co.Pre-Pak units of SODASORB CO2 absorbent manufactured prior to August 2010 may have missing or illegible expiration dates and/or batch numbers. Oct 20, 2010
Z-0561-2011 Darex Container Products Div of W.R. Grace & Co.Pre-Pak units of SODASORB CO2 absorbent manufactured prior to August 2010 may have missing or illegible expiration dates and/or batch numbers. Oct 20, 2010
Z-2152-2011 Allied Healthcare Products IncThe granules do not absorb as long as expected in the anesthesia machines Aug 12, 2010
Recall Darex Container Products Div of W.R. Grace & Co.Both lots were found to be out of specification due to an elevated level of small particles (fines). Mar 31, 2005

Adverse event reports (MAUDE)

Medical device reports received per year that name product code CBL. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
20170
20184
20194
202010
202122
202220
20239
20246
202517
2026 (so far)21

Last 12 months by event type

  • Deaths0
  • Injuries3
  • Malfunctions18
  • Other or not stated0

Most reported device problems, last 12 months

  1. Energy Output Problem6
  2. Fitting Problem6
  3. Leak/Splash3
  4. Gas/Air Leak2
  5. No Flow2
  6. Failure to Seal1

In plain English: what the newest reports describe Written by AI

  • Reports describe the absorbent canister not sealing properly in the machine, which causes a circuit leak. Some say it was hard to seal during canister changes, and one says the canister fit but failed a leak test. In several cases staff had to bag the patient by hand, and one report says the machine did not register carbon dioxide from the patient.

    6 reports: 1417592-2026-40682, 1417592-2026-05826, 1924066-2026-00004, 1924066-2025-00081, 1924066-2025-00082, 1924066-2025-00083

  • Reports describe ventilation stopping during a case and the patient going into respiratory arrest. The damaged absorber container was cited, and the patient recovered in each report. One report says the manufacturer of the damaged disposable container was not the device maker.

    3 reports: MW5189851, MW5189853, 3010838917-2026-00006

  • Two reports describe leaks in prefilled absorber canisters from one lot, found at two hospitals. The reports say leaks were found in all customer batches, and no patient harm was reported. The manufacturer could not investigate because the devices were not returned.

    2 reports: 3002226076-2026-01568, 3002226076-2026-01569

  • A report says the manufacturer's team found cracked canisters from one lot. The report does not describe any effect on a patient.

    Report: MW5190462

  • A report says the flow sensors and their oxygen rings fell out of the machine along with the absorbent canisters. Staff had to ventilate the patient by hand until a technician put the part back.

    Report: 1924066-2026-00004

Written by AI from the narratives of the 17 newest reports that have one, received Nov 14, 2025 to Sep 15, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
134in the FDA’s Global Unique Device Identification Database
Establishment listings
25registered establishments that list this product code

Watch CBL in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get CBL every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in CBL. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.