FDA product code
CAP: Monitor, Airway Pressure (Includes Gauge And/Or Alarm) (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 868.2600
- Review panel
- Anesthesiology
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 791 in the last 12 months
- Median review time
- 196 daysthe latest 25 decisions
- Recall records
- 00 in the last 12 months
- Adverse event reports
- 1last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code CAP per year, by decision date, 79 since 1978. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 0 |
| 2018 | 1 |
| 2019 | 2 |
| 2020 | 1 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 1 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K260671 | RP Check (RP01)MD Diagnostics, Ltd. | Oct 26, 2026238 days |
| K201666 | SINGLE USE MANOMETERFlexicare Medical Limited. | Nov 13, 2020147 days |
| K182292 | VORTRAN APM-PlusVortran Medical Technology 1, Inc. | Sep 18, 2019391 days |
| K180510 | NS 120P-TRS Airway Pressure GaugeInstrumentation Industries, Inc. | May 1, 2019428 days |
| K172284 | NAPA LP-15 Airway Pressure MonitorDrw Medical, LLC | Feb 9, 2018196 days |
| K153733 | VORTRAN ManometerVortran Medical Technology 1, Inc. | Sep 16, 2016263 days |
| K143148 | Disposable Pressure Manometer (DPM)Westmed, Inc. | Aug 11, 2015281 days |
| K140370 | DISPOSABLE PRESSURE MANOMETERBesmed Health Business Corp | Jun 19, 2014125 days |
| K122627 | GIO DIGITAL PRESSURE GUAGESGalemed Corp. | Aug 22, 2013359 days |
| K122077 | MANOMETERIntersurgical, Inc. | Jan 10, 2013178 days |
Recalls
Recall records for product code CAP per year, by the date the recall began. Each recalled product is one record.
No recalls on record for this product code.
Adverse event reports (MAUDE)
Medical device reports received per year that name product code CAP. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 4 |
| 2020 | 1 |
| 2021 | 5 |
| 2022 | 6 |
| 2023 | 11 |
| 2024 | 11 |
| 2025 | 2 |
| 2026 (so far) | 1 |
Last 12 months by event type
- Deaths0
- Injuries0
- Malfunctions0
- Other or not stated1
Most reported device problems, last 12 months
- Output Problem1
In plain English: what the newest reports describe Written by AI
Two reports describe a cuff-pressure device (a cuff pressure meter used with airway tubes) that gave wrong or failed readings. In one, the cuff was found deflated and a comparison meter showed the reading did not reach the expected level; that report also says the patient was at home and no harm was reported. The other, a near-identical report, describes the same events and patient.
2 reports: 3012307300-2023-08554, 3012307300-2023-08199
Two reports describe a measured value or pressure reading that was wrong or did not change correctly. One says the pressure decreased during use and the other says the measured value was incorrect; both state no patient harm was reported.
2 reports: 3012307300-2024-07929, 3012307300-2024-07541
Two reports describe a gauge or pointer that would not return to zero. One says the pointer did not return to zero after the air release button was pressed; another says the pressure gauge was not set to zero and did not respond to pressure. Neither reports patient harm.
2 reports: 3012307300-2023-08202, 9616567-2024-00067
Two reports describe an odor or strange smell noticed during pre-use checks. One report says the product was replaced for this reason and held at a sales office until the smell faded before delivery.
2 reports: 3012307300-2023-07331, 3012307300-2023-05216
Two reports describe a problem during priming, where air escaped under pressure in one and the rear hook was damaged in the other. Neither reports patient harm.
2 reports: 3012307300-2024-08353, 3012307300-2024-03972
Written by AI from the narratives of the 19 newest reports that have one, received May 5, 2023 to Jul 23, 2025. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 188in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 98registered establishments that list this product code
Watch CAP in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get CAP every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in CAP. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:CAPRetrieved Nov 28, 2026, 8:34 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:CAPRetrieved Nov 29, 2026, 4:49 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:CAP · count decision_dateRetrieved Nov 28, 2026, 8:34 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:CAPRetrieved Nov 29, 2026, 4:49 PM UTC.
- device/recall · search product_code:CAP · count event_date_initiatedRetrieved Nov 28, 2026, 8:34 PM UTC.
- device/recall · search product_code:CAP · count res_event_numberRetrieved Nov 28, 2026, 8:34 PM UTC.
- device/event · search device.device_report_product_code:CAP AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 4:49 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:CAP AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 4:49 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:CAP AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 4:49 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:CAPRetrieved Nov 28, 2026, 8:34 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:CAPRetrieved Nov 28, 2026, 8:34 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:868.2600Retrieved Nov 28, 2026, 8:34 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil