Product codes

FDA product code

BSN: Filter, Conduction, Anesthetic (FDA product code)

Follow BSN: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 868.5130
Review panel
Anesthesiology
Premarket route
510(k)
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
90 in the last 12 months
Median review time
121 daysthe latest 9 decisions
Recall records
10 in the last 12 months
Adverse event reports
9last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code BSN per year, by decision date, 9 since 1978. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20170
20180
20191
20200
20210
20220
20230
20240
20250
2026 (so far)0

Latest decisions

NumberDevice and applicantDecision and review time
K190663 Epidural MiniFilter LUER 80369-7, Epidural MiniFilter NRFit 80369-6PAJUNK GmbH Medizintechnologie Jun 17, 201994 days
K133004 VENTED MILLEX -GV, CATHIVEX -GVMerck Millipore , Ltd. Mar 27, 2014183 days
K083451 PORTEX EPIDURAL FILTER, MODEL 100/386/010, VARIOUS STANDARD AND CUSTOM TRAYSSmiths Medical Asd, Inc. Mar 4, 2009103 days
K023892 MILLEX VV, GV, AND HV SYRINGE FILTER UNITSMillipore Corp. Jan 17, 200356 days
K013988 MILLEX GP STERILIZING FILTER UNIT, MODEL SLGP 033 RMillipore Corp. Apr 4, 2002121 days
K963717 MILLEX - MP FILTER UNIT (SLMPO25SS/SLMPL25SS)Millipore Corp. Jun 10, 1997267 days
K960928 SYRINGEFILTERArbor Technologies, Inc. Apr 4, 1997393 days
K902666 ANESTHESIA CONDUCTION FILTERArbor Technologies, Inc. Apr 23, 1991309 days
K780519 FILTER TUBEJelco Laboratories May 9, 197839 days

Recalls

Recall records for product code BSN per year, by the date the recall began. Each recalled product is one record; these 1 records belong to 1 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20170
20180
20190
20200
20210
20220
20230
20240
20250
2026 (so far)0

Latest recalls

Recall and classFirm and reasonBegan
Z-1044-2010 Millipore CorporationMembrane integrity defects may allow the passage of viable or non-viable contaminants to pass through the filter. Nov 25, 2009

Adverse event reports (MAUDE)

Medical device reports received per year that name product code BSN. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
20170
20181
20191
20201
20213
202223
202318
202411
20256
2026 (so far)8

Last 12 months by event type

  • Deaths0
  • Injuries1
  • Malfunctions8
  • Other or not stated0

Most reported device problems, last 12 months

  1. Fluid/Blood Leak4
  2. Break2
  3. Crack1
  4. Difficult to Insert1
  5. Material Deformation1
  6. Material Separation1

In plain English: what the newest reports describe Written by AI

  • Reports describe the device leaking. One describes leakage during dosing, and others describe air leaks that led to use being stopped or that happened during patient use. One report describes water leaking from the plunger.

    4 reports: 3012307300-2026-00708, 3012307300-2024-09368, 3012307300-2024-09541, 3012307300-2024-11150

  • The filter came apart from the connector it attaches to. In one case this caused a leak during infusion, and in another, boluses stopped working until anesthetists were called. One report says the epidural was left open to air for an unknown time.

    3 reports: 3012307300-2026-06280, 3012307300-2026-04648, 3012307300-2025-03596

  • Filters had defects before use. One filter failed a pressure test meant to check its barrier, with a reported result of 35-45 PSI instead of the labeled 50 PSI. Another had a crack found when opened, and a third had black spots that looked like lint.

    3 reports: MW5179067, 9680661-2025-00024, 3012307300-2024-08622

  • Epidural catheters broke or fractured when they were being inserted or withdrawn, and the reports suspect a piece may have stayed in the epidural space. One report estimates about 4 cm was left behind and says the patient was kept under observation.

    2 reports: 3012307300-2026-07036, 3012307300-2025-13497

  • Catheters could not be flushed or were blocked. In one report, very high pressure occurred when bolusing a nerve catheter, and replacing the filter and connector did not fix it. In another, an epidural catheter appeared obstructed and was found defective when removed, with no holes found.

    2 reports: 3012307300-2026-00436, 3012307300-2025-03676

Written by AI from the narratives of the 17 newest reports that have one, received Sep 5, 2024 to Jul 6, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
50in the FDA’s Global Unique Device Identification Database
Establishment listings
32registered establishments that list this product code

Watch BSN in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get BSN every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in BSN. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.