FDA device news · Product code NRY

FDA 510(k) clearance

Route 92 Medical Full Length 046 Reperfusion System and Aspiration Tubing Set (K263266): Route 92 Medical, Inc. 510(k) clearance

San Mateo, CA

Follow NRY: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Decision
Substantially Equivalent
Decision date
Oct 16, 2026
Received
Sep 18, 2026
510(k) type
Special
Product code
NRYCatheter, Thrombus Retriever
Review panel
Cardiovascular
Device class
Class II
Regulation
21 CFR 870.1250
Third-party review
No
Summary or statement
Summary
Review time
28 daysreceipt to decision, calendar days
Median in NRY
130 daysthe latest 24 other decisions
NRY decisions
14in the last 12 months
Applicant’s 510(k)s
15in openFDA, this one included

In plain English Written by AI

What the company’s 510(k) summary says about the device, its predicate and its testing, in plain words.

What it is
The Route 92 Medical Full Length 046 Reperfusion System is a set of thin, flexible catheters (tubes threaded through blood vessels) used to remove blood clots from the brain. It includes a 046 Aspiration Catheter, a Tenzing 4 Delivery Catheter, and an Aspiration Tubing Set that connects to a vacuum pump to draw clot material out through the catheter.
What it’s cleared for
The device is cleared to revascularize (restore blood flow in) patients with acute ischemic stroke caused by a large blockage in arteries of the brain, within 8 hours of symptom onset. It is for patients who cannot receive or have failed thrombolytic drug therapy (clot-dissolving drugs). It is intended for use by physicians trained in neurovascular interventional techniques.
Compared with (the predicate)
  • K233329, Route 92 Medical Full Length 054 Reperfusion System and Aspiration Tubing Set
How it differs
The summary says the 046 system has a smaller aspiration catheter (0.046 inch inner diameter versus 0.054 inch) and a slightly shorter 143 cm length option in place of 148 cm. The indications differ only in wording: the predicate refers to patients ineligible for IV t-PA, while this device refers to thrombolytic drug therapy in general.
Testing the summary lists
  • Biocompatibility: no new testing, because patient-contacting materials were unchanged from the predicate
  • Sterilization by ethylene oxide (sterility assurance level 10^-6, AAMI TIR28:2016) and shelf life aging studies

Written by AI from the 510(k) summary the company filed with the FDA (the PDF), and checked against it: predicate numbers and figures must appear in the summary. A summary is the company’s own account, not the FDA’s review, and the AI can still get it wrong: read the PDF before you rely on it.

How the review time compares

Calendar days from the day FDA received the 510(k) to its decision. They include time the submission waited on the company for answers, so they run longer than FDA’s 90-day review goal.

K263266 took 28 calendar days from receipt to decision. Of the latest 24 other decisions in product code NRY, 22 took longer, 2 took the same time; their median was 130 days.

Latest decisions in NRY

NumberDevice and applicantDecision and review time
K262021 Toro 71 Aspiration Catheter; Toro Aspiration TubingToro Neurovascular, Inc. · Traditional Oct 21, 2026127 days
K263266 Route 92 Medical Full Length 046 Reperfusion System and Aspiration Tubing SetRoute 92 Medical, Inc. · Special Oct 16, 202628 days
K261089 Radical 064 Reperfusion Catheter; Radical 076 Reperfusion Catheter; Radical Aspiration Tubing KitMaduro Medical, Inc. · Traditional Sep 14, 2026166 days
K260617 Tigertriever 13 Ultra Revascularization DeviceRapid Medical Ltd. · Traditional Aug 28, 2026184 days
K262398 CEREGLIDE 57 Intermediate Catheter; Cerenovus Aspiration Tubing setCERENOVUS, Inc. · Special Aug 13, 202630 days
K262391 APRO 35 Swift Catheter and Alembic Aspiration TubingAlembic, LLC · Special Aug 12, 202629 days
K262399 Millipede SystemPerfuze, Ltd. · Special Aug 12, 202629 days
K260002 Socrates Aspiration System (Socrates 70 Aspiration Catheter)Scientia Vascular, Inc. · Traditional Jul 16, 2026195 days
K250690 Penumbra System (Thunderbolt Aspiration Tubing)Penumbra, Inc. · Traditional Jun 10, 2026460 days
K252707 Zenith Aspiration Catheter; Disposable Aspiration Tubing SetSuzhou Zenith Vascular SciTech Limited · Traditional May 20, 2026266 days

Other 510(k)s from Route 92 Medical, Inc.

The applicant’s other clearances in openFDA, newest first. Company names are matched as the FDA recorded them, so a renamed or differently spelled company may not appear.

NumberDevice and applicantDecision and review time
K260220 Route 92 Medical Full Length 046 Access SystemRoute 92 Medical, Inc. · Traditional Sep 16, 2026233 days
K243601 Route 92 Medical HiPoint Reperfusion System and Aspiration Tubing SetRoute 92 Medical, Inc. · Traditional May 19, 2025179 days
K240529 8F Modified Sheath SystemRoute 92 Medical, Inc. · Special Oct 2, 2024222 days
K233329 Route 92 Medical Full Length 054 Reperfusion System and Aspiration Tubing SetRoute 92 Medical, Inc. · Special Apr 23, 2024207 days
K231583 Route 92 Medical Full Length 054 Access SystemRoute 92 Medical, Inc. · Special Aug 15, 202376 days
K223530 Route 92 Medical Full Length 070 Reperfusion System and Aspiration Tubing SetRoute 92 Medical, Inc. · Traditional Apr 11, 2023139 days
K222743 Route 92 Medical Full Length 070 Access SystemRoute 92 Medical, Inc. · Special Oct 4, 202225 days
K210635 Route 92 Medical Full Length 088 Access SystemRoute 92 Medical, Inc. · Special Jul 27, 2021146 days
K203043 Route 92 Medical 070 Access SystemRoute 92 Medical, Inc. · Special Nov 11, 202036 days
K200121 Route 92 Medical 088 Access System, 110 cmRoute 92 Medical, Inc. · Traditional Apr 18, 202088 days
K191717 Route 92 Medical Sheath SystemRoute 92 Medical, Inc. · Traditional Jan 13, 2020201 days
K190431 Route 92 Medical Delivery CatheterRoute 92 Medical, Inc. · Special Mar 21, 201927 days

Watch NRY in DeviceVigil

DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and it tells you about new clearances, recalls and MAUDE reports for them, right beside the FDA’s own pages.

Add to Chrome — free Add to Edge

Get NRY every Monday

One short email after openFDA’s weekly update with that week’s new 510(k)s, De Novos, PMAs and recalls in NRY. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.