FDA device news · Product code NBC
FDA 510(k) clearance
Access BNP II (K262597): Beckman Coulter, Inc. 510(k) clearance
Chaska, MN
- Decision
- Substantially Equivalent
- Decision date
- Oct 20, 2026
- Received
- Jul 27, 2026
- 510(k) type
- Traditional
- Product code
- NBCTest, Natriuretic Peptide
- Review panel
- Clinical Chemistry
- Device class
- Class II
- Regulation
- 21 CFR 862.1117
- Third-party review
- No
- Summary or statement
- Summary
- Review time
- 85 daysreceipt to decision, calendar days
- Median in NBC
- 175 daysthe latest 24 other decisions
- NBC decisions
- 4in the last 12 months
- Applicant’s 510(k)s
- 273in openFDA, this one included
In plain English Written by AI
What the company’s 510(k) summary says about the device, its predicate and its testing, in plain words.
- What it is
- The Access BNP II is an in vitro diagnostic test, meaning it is run on a sample outside the body. It measures B-type natriuretic peptide (BNP), a protein the heart releases when its walls are stretched, in plasma (the liquid part of blood collected in an EDTA tube).
- What it’s cleared for
- The test is cleared for use with the Beckman Coulter Access Family of Immunoassay Systems to measure BNP quantitatively in EDTA plasma. It is intended as an aid in diagnosing heart failure in patients who come to the emergency department with suspected new onset, acutely worsening, or flare-up heart failure. It is also intended to risk stratify patients with acute coronary syndromes and patients with heart failure. The test is prescription use only.
- Compared with (the predicate)
- K252169, Access BNP II
- How it differs
- The summary says the candidate device runs on the Access 2 Immunoassay System, while the predicate runs on the DxI 9000 Access Immunoassay Analyzer. The sample volume also differs: 30 μL for the candidate device versus 13 μL for the predicate. The summary lists the intended use, measuring range, and other listed characteristics as the same.
- Testing the summary lists
- Linearity (CLSI EP06 2nd Edition), performed on the Access 2 Immunoassay System
- Imprecision (CLSI EP05-A3), performed on the Access 2 Immunoassay System over 20 or more days
- High dose hook effect, tested up to 150,000 pg/mL
- Limit of blank, limit of detection, and limit of quantitation (CLSI EP17-A2), performed on the Access 2 Immunoassay System
- Method comparison against the DxI 9000 Access Immunoassay Analyzer (CLSI EP09c, Passing-Bablok regression)
Written by AI from the 510(k) summary the company filed with the FDA (the PDF), and checked against it: predicate numbers and figures must appear in the summary. A summary is the company’s own account, not the FDA’s review, and the AI can still get it wrong: read the PDF before you rely on it.
How the review time compares
Calendar days from the day FDA received the 510(k) to its decision. They include time the submission waited on the company for answers, so they run longer than FDA’s 90-day review goal.
K262597 took 85 calendar days from receipt to decision. Of the latest 24 other decisions in product code NBC, 19 took longer, 5 were faster; their median was 175 days.
Latest decisions in NBC
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K262597 | Access BNP IIBeckman Coulter, Inc. · Traditional | Oct 20, 202685 days |
| K262109 | Access NT-proBNPBeckman Coulter, Inc. · Traditional | Sep 21, 202690 days |
| K252169 | Access BNP IIBeckman Coulter, Inc. · Traditional | Mar 24, 2026257 days |
| K253539 | Alere NT-proBNP for Alinity iAxis-Shield Diagnostics, Ltd. · Traditional | Feb 18, 202697 days |
| K241176 | Alere NT-proBNP for Alinity i Reagent KitAxis-Shield Diagnostics, Ltd. · Traditional | Jan 16, 2025262 days |
| K232164 | Access NT-proBNPBeckman Coulter, Inc. · Traditional | Apr 12, 2024266 days |
| K220265 | ADVIA Centaur® NT-proBNPII (PBNPII)Siemens Healthcare Diagnostics, Inc. · Traditional | Sep 24, 2023601 days |
| K223637 | Elecsys proBNP II, Elecsys proBNP II STATRoche Diagnostics · Traditional | Jul 21, 2023228 days |
| K210546 | Elecsys proBNP II, Elecsys proBNP II STATRoche Diagnostics · Traditional | Mar 31, 2022399 days |
| K211199 | ST AIA-PACK BNP AssayTosoh Bioscience, Inc. · Special | Nov 8, 2021200 days |
Other 510(k)s from Beckman Coulter, Inc.
The applicant’s other clearances in openFDA, newest first. Company names are matched as the FDA recorded them, so a renamed or differently spelled company may not appear.
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K261109 | Access MeMed BVBeckman Coulter, Inc. · Traditional | Oct 19, 2026199 days |
| K262109 | Access NT-proBNPBeckman Coulter, Inc. · Traditional | Sep 21, 202690 days |
| K252169 | Access BNP IIBeckman Coulter, Inc. · Traditional | Mar 24, 2026257 days |
| K254059 | Access anti-HBc IgMBeckman Coulter, Inc. · Traditional | Mar 13, 202686 days |
| K253687 | Access anti-HBc TotalBeckman Coulter, Inc. · Traditional | Feb 19, 202690 days |
| K251995 | Access anti-HAV IgMBeckman Coulter, Inc. · Traditional | Jan 27, 2026214 days |
| K250588 | Access Rubella IgGBeckman Coulter, Inc. · Traditional | Nov 17, 2025263 days |
| K252580 | iQ200 SeriesBeckman Coulter, Inc. · Special | Sep 10, 202526 days |
| K243846 | Access anti-HAVBeckman Coulter, Inc. · Traditional | Sep 9, 2025267 days |
| K243804 | MicroScan Dried Gram-Negative MIC/Combo Panels with Cefepime (CPE) (0.12-64 µg/mL) (MicroScan)Beckman Coulter, Inc. · Traditional | Aug 20, 2025252 days |
| K250036 | MicroScan Dried Gram-Positive MIC/Combo Panels with Daptomycin (DAP) (0.06-32 µg/mL)Beckman Coulter, Inc. · Traditional | Aug 15, 2025219 days |
| K243483 | Access hsTnIBeckman Coulter, Inc. · Traditional | Aug 1, 2025266 days |
Watch NBC in DeviceVigil
DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and it tells you about new clearances, recalls and MAUDE reports for them, right beside the FDA’s own pages.
Get NBC every Monday
One short email after openFDA’s weekly update with that week’s new 510(k)s, De Novos, PMAs and recalls in NBC. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/510k · search k_number:K262597Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:NBCRetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/classification · search product_code:NBCRetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:NBC AND decision_date:[20251129 TO 20261129]Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search applicant:"Beckman Coulter, Inc."Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil