FDA device news · Product code SEM
FDA 510(k) clearance
Minder® System (K262318): Epi-Minder Pty, Ltd. 510(k) clearance
Melbourne, AU
- Decision
- Substantially Equivalent
- Decision date
- Oct 29, 2026
- Received
- Jul 8, 2026
- 510(k) type
- Traditional
- Product code
- SEMSub-Scalp Implanted Electroencephalogram System For Remote Patient Monitoring
- Review panel
- Neurology
- Device class
- Class II
- Regulation
- 21 CFR 882.1360
- Third-party review
- No
- Summary or statement
- Summary
- Review time
- 113 daysreceipt to decision, calendar days
- Median in SEM
- 193 daysthe latest 2 other decisions
- SEM decisions
- 2in the last 12 months
- Applicant’s 510(k)s
- 2in openFDA, this one included
In plain English Written by AI
What the company’s 510(k) summary says about the device, its predicate and its testing, in plain words.
- What it is
- The Minder System is an electroencephalographic (EEG, a recording of the brain's electrical activity) recording and transmitting device implanted under the scalp. An implant with two recording channels, each using two electrodes, sends EEG signals through the skin to an externally worn wearable behind the ear, which passes them over Bluetooth Low Energy to a Minder App on a locked-down Android phone.
- What it’s cleared for
- The Minder System is cleared to acquire, transmit, and store EEG data continuously from patients aged 12 to 75 with drug-resistant epilepsy who are intolerant of, or not indicated for, more conservative monitoring tools. It is meant to aid a physician's remote assessment and monitoring, and it does not provide diagnostic conclusions.
- Compared with (the predicate)
- DEN240062, Minder System
- How it differs
- The subject device's indicated age range is 12 to 75 years, while the predicate's is 18 to 75 years. The summary says the hardware and software are unchanged, and it supports the younger age group through a pediatric extrapolation, which uses existing clinical data from the predicate study.
- Testing the summary lists
- Performance bench testing, leveraged from DEN240062
- Clinical performance data, pediatric extrapolation for ages 12 to 17 (justified from existing predicate clinical data)
- Risk analysis update per ISO 14971 for growth between ages 12 and 17
- Biocompatibility (ISO 10993-1)
Written by AI from the 510(k) summary the company filed with the FDA (the PDF), and checked against it: predicate numbers and figures must appear in the summary. A summary is the company’s own account, not the FDA’s review, and the AI can still get it wrong: read the PDF before you rely on it.
How the review time compares
Calendar days from the day FDA received the 510(k) to its decision. They include time the submission waited on the company for answers, so they run longer than FDA’s 90-day review goal.
K262318 took 113 calendar days from receipt to decision. Of the latest 2 other decisions in product code SEM, 2 took longer; their median was 193 days.
Latest decisions in SEM
Other 510(k)s from Epi-Minder Pty, Ltd.
The applicant’s other clearances in openFDA, newest first. Company names are matched as the FDA recorded them, so a renamed or differently spelled company may not appear.
| Number | Device and applicant | Decision and review time |
|---|---|---|
| DEN240062 | Minder SystemEpi-Minder Pty, Ltd. · Direct | Apr 17, 2025181 days |
Watch SEM in DeviceVigil
DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and it tells you about new clearances, recalls and MAUDE reports for them, right beside the FDA’s own pages.
Get SEM every Monday
One short email after openFDA’s weekly update with that week’s new 510(k)s, De Novos, PMAs and recalls in SEM. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/510k · search k_number:K262318Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:SEMRetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/classification · search product_code:SEMRetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:SEM AND decision_date:[20251129 TO 20261129]Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search applicant:"Epi-Minder Pty, Ltd."Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.