FDA device news · Product code SEM

FDA 510(k) clearance

Minder® System (K262318): Epi-Minder Pty, Ltd. 510(k) clearance

Melbourne, AU

Follow SEM: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Decision
Substantially Equivalent
Decision date
Oct 29, 2026
Received
Jul 8, 2026
510(k) type
Traditional
Product code
SEMSub-Scalp Implanted Electroencephalogram System For Remote Patient Monitoring
Review panel
Neurology
Device class
Class II
Regulation
21 CFR 882.1360
Third-party review
No
Summary or statement
Summary
Review time
113 daysreceipt to decision, calendar days
Median in SEM
193 daysthe latest 2 other decisions
SEM decisions
2in the last 12 months
Applicant’s 510(k)s
2in openFDA, this one included

In plain English Written by AI

What the company’s 510(k) summary says about the device, its predicate and its testing, in plain words.

What it is
The Minder System is an electroencephalographic (EEG, a recording of the brain's electrical activity) recording and transmitting device implanted under the scalp. An implant with two recording channels, each using two electrodes, sends EEG signals through the skin to an externally worn wearable behind the ear, which passes them over Bluetooth Low Energy to a Minder App on a locked-down Android phone.
What it’s cleared for
The Minder System is cleared to acquire, transmit, and store EEG data continuously from patients aged 12 to 75 with drug-resistant epilepsy who are intolerant of, or not indicated for, more conservative monitoring tools. It is meant to aid a physician's remote assessment and monitoring, and it does not provide diagnostic conclusions.
Compared with (the predicate)
  • DEN240062, Minder System
How it differs
The subject device's indicated age range is 12 to 75 years, while the predicate's is 18 to 75 years. The summary says the hardware and software are unchanged, and it supports the younger age group through a pediatric extrapolation, which uses existing clinical data from the predicate study.
Testing the summary lists
  • Performance bench testing, leveraged from DEN240062
  • Clinical performance data, pediatric extrapolation for ages 12 to 17 (justified from existing predicate clinical data)
  • Risk analysis update per ISO 14971 for growth between ages 12 and 17
  • Biocompatibility (ISO 10993-1)

Written by AI from the 510(k) summary the company filed with the FDA (the PDF), and checked against it: predicate numbers and figures must appear in the summary. A summary is the company’s own account, not the FDA’s review, and the AI can still get it wrong: read the PDF before you rely on it.

How the review time compares

Calendar days from the day FDA received the 510(k) to its decision. They include time the submission waited on the company for answers, so they run longer than FDA’s 90-day review goal.

K262318 took 113 calendar days from receipt to decision. Of the latest 2 other decisions in product code SEM, 2 took longer; their median was 193 days.

Latest decisions in SEM

NumberDevice and applicantDecision and review time
K262318 Minder® SystemEpi-Minder Pty, Ltd. · Traditional Oct 29, 2026113 days
K253607 UNEEG EpiSight SystemUneeg Medical A/S · Traditional Jun 11, 2026205 days
DEN240062 Minder SystemEpi-Minder Pty, Ltd. · Direct Apr 17, 2025181 days

Other 510(k)s from Epi-Minder Pty, Ltd.

The applicant’s other clearances in openFDA, newest first. Company names are matched as the FDA recorded them, so a renamed or differently spelled company may not appear.

NumberDevice and applicantDecision and review time
DEN240062 Minder SystemEpi-Minder Pty, Ltd. · Direct Apr 17, 2025181 days

Watch SEM in DeviceVigil

DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and it tells you about new clearances, recalls and MAUDE reports for them, right beside the FDA’s own pages.

Add to Chrome — free Add to Edge

Get SEM every Monday

One short email after openFDA’s weekly update with that week’s new 510(k)s, De Novos, PMAs and recalls in SEM. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.