FDA device news · Product code ONO
FDA 510(k) clearance
NeuroBlate System (K262315): Monteris Medical 510(k) clearance
Minnetonka, MN
- Decision
- Substantially Equivalent
- Decision date
- Sep 3, 2026
- Received
- Jul 8, 2026
- 510(k) type
- Traditional
- Product code
- ONONeurosurgical Laser With Mr Thermography
- Review panel
- General, Plastic Surgery
- Device class
- Class II
- Regulation
- 21 CFR 878.4810
- Third-party review
- No
- Summary or statement
- Summary
- Review time
- 57 daysreceipt to decision, calendar days
- Median in ONO
- 129 daysthe latest 7 other decisions
- ONO decisions
- 3in the last 12 months
- Applicant’s 510(k)s
- 22in openFDA, this one included
In plain English Written by AI
What the company’s 510(k) summary says about the device, its predicate and its testing, in plain words.
- What it is
- The NeuroBlate System is a set of MRI-compatible laser devices and accessories used in neurosurgery to deliver precise thermal (heat) therapy to a target zone in the brain. Gas-cooled laser probes deliver controlled energy, a robotic probe driver helps the surgeon position and steer the probe, and control software supports procedure planning and real-time monitoring under MRI.
- What it’s cleared for
- The NeuroBlate System is cleared as a neurosurgical tool to ablate, necrotize, or coagulate (destroy or heat-treat) intracranial soft tissue, including brain tumors, radiation necrosis, and epileptogenic foci, using interstitial irradiation or thermal therapy with 1064 nm lasers. It is indicated for pediatric and adult patients, with the intended patients being adults and pediatric patients from age 2 and older. The submission describes the change as a clarification of existing wording, not a new indication.
- Compared with (the predicate)
- How it differs
- The summary says the device proposes no physical, hardware, software, material, or technology changes. The only change is clarified wording in the Indications for Use, such as updated terminology describing how the device is used today.
- Testing the summary lists
- In-vitro (bench) data from prior submissions, considered still applicable because only labeling changed
- In-vivo (animal) data from prior submissions, considered still applicable because only labeling changed
- Risk assessment and risk mitigation documentation from prior submissions, considered still applicable
Written by AI from the 510(k) summary the company filed with the FDA (the PDF), and checked against it: predicate numbers and figures must appear in the summary. A summary is the company’s own account, not the FDA’s review, and the AI can still get it wrong: read the PDF before you rely on it.
How the review time compares
Calendar days from the day FDA received the 510(k) to its decision. They include time the submission waited on the company for answers, so they run longer than FDA’s 90-day review goal.
K262315 took 57 calendar days from receipt to decision. Of the latest 7 other decisions in product code ONO, 6 took longer, 1 was faster; their median was 129 days.
Latest decisions in ONO
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K262315 | NeuroBlate SystemMonteris Medical · Traditional | Sep 3, 202657 days |
| K260976 | NeuroBlate SystemMonteris Medical · Special | Apr 23, 202630 days |
| K253391 | Visualase Cooled Laser Applicator System (9735559, 9735560, 9735561)Medtronic Navigation, Inc. · Traditional | Mar 13, 2026164 days |
| K251298 | Mobile Laser Unit (1001-N2); Thermoguide Workstation (1100-N1); Laser Applicator (4012-N5, 4017-N2, 4017-N4); MR Introducer (4013-N6)Clinical Laserthermia Systems, AB · Traditional | Sep 4, 2025129 days |
| K250307 | Visualase V2 MRI-guided Laser Ablation System (9736422)Medtronic Navigation · Traditional | May 29, 2025115 days |
| K240877 | Monteris Medical NeuroBlate SystemMonteris Medical, · Traditional | Jun 4, 202467 days |
| K214125 | TRANBERG Thermoguide Therapy SystemClinical Laserthermia Systems, AB · Traditional | Sep 22, 2022266 days |
| K211269 | Visualase MRI-Guided Laser Ablation System (SW 3.4)Medtronic Navigation, Inc. · Traditional | Jan 7, 2022255 days |
Other 510(k)s from Monteris Medical
The applicant’s other clearances in openFDA, newest first. Company names are matched as the FDA recorded them, so a renamed or differently spelled company may not appear.
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K260976 | NeuroBlate SystemMonteris Medical · Special | Apr 23, 202630 days |
| K240877 | Monteris Medical NeuroBlate SystemMonteris Medical, · Traditional | Jun 4, 202467 days |
| K231061 | NeuroBlate SystemMonteris Medical, · Traditional | Dec 8, 2023239 days |
| K222983 | NeuroBlate Fusion-S Software V3.17Monteris Medical, · Traditional | Mar 15, 2023168 days |
| K201056 | NeuroBlate SystemMonteris Medical, Inc. · Traditional | Aug 21, 2020122 days |
| K193375 | NeuroBlate SystemMonteris Medical, · Special | Jan 3, 202029 days |
| K182036 | NeuroBlate SystemMonteris Medical, · Traditional | Oct 15, 201877 days |
| K173305 | NeuroBlate SystemMonteris Medical, · Special | Nov 17, 201730 days |
| K172881 | NeuroBlate SystemMonteris Medical, · Special | Oct 20, 201729 days |
| K171255 | Monteris Medical NeuroBlate SystemMonteris Medical, Inc. · Traditional | Jul 25, 201788 days |
| K170724 | NeuroBlate SystemMonteris Medical, · Special | Apr 7, 201729 days |
| K162762 | NeuroBlate SystemMonteris Medical, Inc. · Special | Oct 26, 201626 days |
Watch ONO in DeviceVigil
DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and it tells you about new clearances, recalls and MAUDE reports for them, right beside the FDA’s own pages.
Get ONO every Monday
One short email after openFDA’s weekly update with that week’s new 510(k)s, De Novos, PMAs and recalls in ONO. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/510k · search k_number:K262315Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:ONORetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/classification · search product_code:ONORetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:ONO AND decision_date:[20251129 TO 20261129]Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search applicant:"Monteris Medical"Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.