FDA device news · Product code KRH

FDA 510(k) clearance

MEMO 4D Curve™ Semirigid mitral annuloplasty ring (K261707): Corcym S.r.l. 510(k) clearance

Saluggia, IT

Follow KRH: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Decision
Substantially Equivalent
Decision date
Jun 18, 2026
Received
May 22, 2026
510(k) type
Special
Product code
KRHRing, Annuloplasty
Review panel
Cardiovascular
Device class
Class II
Regulation
21 CFR 870.3800
Third-party review
No
Summary or statement
Summary
Review time
27 daysreceipt to decision, calendar days
Median in KRH
76 daysthe latest 24 other decisions
KRH decisions
2in the last 12 months
Applicant’s 510(k)s
3in openFDA, this one included

In plain English Written by AI

What the company’s 510(k) summary says about the device, its predicate and its testing, in plain words.

What it is
MEMO 4D Curve is a semirigid open mitral annuloplasty ring, an implant used to reshape the mitral annulus (the ring of tissue around the mitral valve) so the valve leaflets close properly. It has a Nitinol metal core (a superelastic nickel-titanium alloy) enclosed in silicone and covered with knitted polyester fabric. It comes in 10 sizes, is supplied sterile, and is for single use.
What it’s cleared for
MEMO 4D Curve is indicated for patients with congenital or acquired mitral insufficiency or steno-insufficiency (leaking or narrowed and leaking mitral valves) with dilatation and/or deformation of the mitral annulus. The summary states these indications are the same as those of the predicate device.
Compared with (the predicate)
  • K230318, Memo 4D Semirigid Annuloplasty Ring
How it differs
The summary says MEMO 4D Curve is an open ring, while the predicate's design is not described as open. It also says there are slight material differences: the ReChord System is present only on the Memo 4D predicate, and the fabric differs in texturing and tubular diameter, and the Tyvek packaging adhesive differs.
Testing the summary lists
  • Computational structural analysis (finite element analysis) per ISO 5910 and ASME V&V 40-2018
  • Mechanical resistance testing and mechanical characterization of PET fabric
  • Suture pull-out test
  • Fatigue testing per ISO 5910
  • Usability summative evaluation per FDA human factors guidance

Written by AI from the 510(k) summary the company filed with the FDA (the PDF), and checked against it: predicate numbers and figures must appear in the summary. A summary is the company’s own account, not the FDA’s review, and the AI can still get it wrong: read the PDF before you rely on it.

How the review time compares

Calendar days from the day FDA received the 510(k) to its decision. They include time the submission waited on the company for answers, so they run longer than FDA’s 90-day review goal.

K261707 took 27 calendar days from receipt to decision. Of the latest 24 other decisions in product code KRH, 23 took longer, 1 was faster; their median was 76 days.

Latest decisions in KRH

NumberDevice and applicantDecision and review time
K261707 MEMO 4D Curve™ Semirigid mitral annuloplasty ringCorcym S.r.l. · Special Jun 18, 202627 days
K260498 TriMemo™ SEMIRIGID ANNULOPLASTY RINGCorcym S.r.l. · Abbreviated Apr 14, 202660 days
K253232 Seguin Annuloplasty RingABBOTT MEDICAL · Special Oct 29, 202530 days
K251982 Edwards MC3 Tricuspid annuloplasty ring (4900)Edwards Lifesciences, LLC · Abbreviated Sep 23, 202588 days
K251688 Carpentier-Edwards Physio Annuloplasty Ring (4450)Edwards Lifesciences, LLC · Abbreviated Sep 17, 2025107 days
K250859 TransForm McCarthy Mitral Annuloplasty Ring (TF)Genesee Biomedical, Inc. · Traditional Jul 25, 2025126 days
K232599 TransForm McCarthy Mitral Annuloplasty Ring (TF)Genesee Biomedical, Inc. · Traditional Apr 9, 2024225 days
K230679 WellsForm Tricuspid Annuloplasty Band (WF)Genesee Biomedical, Inc. · Traditional May 26, 202374 days
K230318 Memo 3D Semirigid Annuloplasty Ring; Memo 3D ReChord Semirigid Annuloplasty Ring; Memo 4D Semirigid Annuloplasty RingCorcym S.r.l. · Traditional Apr 7, 202360 days
K202253 TruForm Sievers Annuloplasty RingGenesee Biomedical, Inc. · Traditional Oct 6, 202057 days

Other 510(k)s from Corcym S.r.l.

The applicant’s other clearances in openFDA, newest first. Company names are matched as the FDA recorded them, so a renamed or differently spelled company may not appear.

NumberDevice and applicantDecision and review time
K260498 TriMemo™ SEMIRIGID ANNULOPLASTY RINGCorcym S.r.l. · Abbreviated Apr 14, 202660 days
K230318 Memo 3D Semirigid Annuloplasty Ring; Memo 3D ReChord Semirigid Annuloplasty Ring; Memo 4D Semirigid Annuloplasty RingCorcym S.r.l. · Traditional Apr 7, 202360 days

Watch KRH in DeviceVigil

DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and it tells you about new clearances, recalls and MAUDE reports for them, right beside the FDA’s own pages.

Add to Chrome — free Add to Edge

Get KRH every Monday

One short email after openFDA’s weekly update with that week’s new 510(k)s, De Novos, PMAs and recalls in KRH. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.