FDA device news · Product code HSB
FDA 510(k) clearance
Precice™ Max System and Precice™ Ankle Salvage System (K261651): NuVasive Specialized Orthopedics 510(k) clearance
Audubon, PA
- Decision
- Substantially Equivalent
- Decision date
- Jun 18, 2026
- Received
- May 19, 2026
- 510(k) type
- Special
- Product code
- HSBRod, Fixation, Intramedullary And Accessories
- Review panel
- Orthopedic
- Device class
- Class II
- Regulation
- 21 CFR 888.3020
- Third-party review
- No
- Summary or statement
- Summary
- Review time
- 30 daysreceipt to decision, calendar days
- Median in HSB
- 157 daysthe latest 24 other decisions
- HSB decisions
- 20in the last 12 months
- Applicant’s 510(k)s
- 25in openFDA, this one included
In plain English Written by AI
What the company’s 510(k) summary says about the device, its predicate and its testing, in plain words.
- What it is
- The Precice Max System and Precice Ankle Salvage System are implant systems made of titanium alloy. Each uses intramedullary nails (rods placed inside the bone's hollow center), locking screws and end caps, and works with an External Remote Controller (ERC) to lengthen or compress limbs gradually.
- What it’s cleared for
- The Precice Max System is indicated for limb lengthening, open and closed fracture fixation, pseudarthrosis, malunions, nonunions, or bone transport of long bones in patients 18 years and older, and for limb lengthening of the femur and tibia in pediatric patients older than 12.
- Compared with (the predicate)
- How it differs
- The summary says the subject locking screws and end caps have the same design, intended use, material, function, and range of sizes as the predicates. The only difference is that the subject devices are supplied non-sterile, while the predicate devices are supplied sterile.
- Testing the summary lists
- Steam sterilization validation (ISO 17665-1), existing validations adopted for the non-sterile products
Written by AI from the 510(k) summary the company filed with the FDA (the PDF), and checked against it: predicate numbers and figures must appear in the summary. A summary is the company’s own account, not the FDA’s review, and the AI can still get it wrong: read the PDF before you rely on it.
How the review time compares
Calendar days from the day FDA received the 510(k) to its decision. They include time the submission waited on the company for answers, so they run longer than FDA’s 90-day review goal.
K261651 took 30 calendar days from receipt to decision. Of the latest 24 other decisions in product code HSB, 22 took longer, 2 were faster; their median was 157 days.
Latest decisions in HSB
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K261189 | Freedom Nail Ankle Fusion SystemMcGinley Orthopaedic Innovations · Traditional | Oct 19, 2026192 days |
| K261345 | Pediatric Nailing Platform FemurOrthoPediatrics Corp. · Traditional | Sep 24, 2026154 days |
| K261481 | Pangea Femur Reconstruction SystemStryker GmbH · Traditional | Jun 29, 202655 days |
| K253080 | AND Medical Femoral Nail System (PFNA01, PFNA02)Suzhou and Science & Technology Development Corp. · Traditional | Jun 18, 2026268 days |
| K261651 | Precice™ Max System and Precice™ Ankle Salvage SystemNuVasive Specialized Orthopedics · Special | Jun 18, 202630 days |
| K253155 | Fule Metallic Lockable Intramedullary Nail SystemsBeijing Fule Science & Technology Development Co., Ltd. · Traditional | Jun 17, 2026264 days |
| K253046 | HERA Interlocking Intramedullary Nailing SystemZheJiang Decans Medical Devices Co., Ltd. · Traditional | Jun 16, 2026267 days |
| K260146 | FITBONE® TRANSPORT AND LENGTHENING SYSTEMOrthofix Srl · Traditional | May 19, 2026119 days |
| K253991 | Fitbone™ Trochanteric, FITBONE® Transport and Lengthening System, FITBONE® TAAOrthofix Srl · Traditional | May 8, 2026147 days |
| K261043 | AUTOBAHN™ Nailing SystemGlobus Medical, Inc. · Special | Apr 24, 202624 days |
Other 510(k)s from NuVasive Specialized Orthopedics
The applicant’s other clearances in openFDA, newest first. Company names are matched as the FDA recorded them, so a renamed or differently spelled company may not appear.
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K233593 | MAGEC Spinal Bracing and Distraction SystemNuvasive Specialized Orthopedics, Incorporated · Traditional | Feb 6, 202490 days |
| K232267 | Precice Max SystemNuvasive Specialized Orthopedics, Incorporated · Traditional | Dec 15, 2023137 days |
| K230765 | Precice Ankle Salvage SystemNuvasive Specialized Orthopedics, Inc. · Traditional | Aug 29, 2023162 days |
| K220234 | Precice Intramedullary Limb Lengthening SystemNuvasive Specialized Orthopedics, Inc. · Traditional | Mar 15, 2023412 days |
| K202348 | External Remote Controller ERC 4PNuvasive Specialized Orthopedics, Inc. · Traditional | Dec 22, 2020126 days |
| K200430 | Precice® Ankle Salvage SystemNuvasive Specialized Orthopedics, Inc. · Traditional | Sep 16, 2020208 days |
| K201567 | Precice® Bone Transport SystemNuvasive Specialized Orthopedics, Inc. · Traditional | Aug 10, 202060 days |
| K201543 | MAGEC® SystemNuvasive Specialized Orthopedics, Inc. · Traditional | Jul 30, 202051 days |
| K193016 | PRECICE Bone Transport SystemNuvasive Specialized Orthopedics, Inc. · Traditional | May 8, 2020192 days |
| K193617 | Precice® ScrewsNuvasive Specialized Orthopedics, Inc. · Traditional | May 8, 2020134 days |
| K192181 | PRECICE Plating SystemNuvasive Specialized Orthopedics, Inc. · Traditional | Nov 7, 201987 days |
| K191336 | PRECICE SystemNuvasive Specialized Orthopedics, Inc. · Traditional | Sep 26, 2019132 days |
Watch HSB in DeviceVigil
DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and it tells you about new clearances, recalls and MAUDE reports for them, right beside the FDA’s own pages.
Get HSB every Monday
One short email after openFDA’s weekly update with that week’s new 510(k)s, De Novos, PMAs and recalls in HSB. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/510k · search k_number:K261651Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:HSBRetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/classification · search product_code:HSBRetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:HSB AND decision_date:[20251129 TO 20261129]Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search applicant:"NuVasive Specialized Orthopedics"Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil