FDA device news · Product code QDQ
FDA 510(k) clearance
qCT Lung (v1.0) (K261517): Qure.Ai Technologies Pvt. , Ltd. 510(k) clearance
Mumbai, IN
- Decision
- Substantially Equivalent
- Decision date
- Sep 14, 2026
- Received
- May 7, 2026
- 510(k) type
- Traditional
- Product code
- QDQRadiological Computer Assisted Detection/Diagnosis Software For Lesions Suspicious For Cancer
- Review panel
- Radiology
- Device class
- Class II
- Regulation
- 21 CFR 892.2090
- Third-party review
- No
- Summary or statement
- Summary
- Review time
- 130 daysreceipt to decision, calendar days
- Median in QDQ
- 164 daysthe latest 24 other decisions
- QDQ decisions
- 11in the last 12 months
- Applicant’s 510(k)s
- 1in openFDA, this one included
In plain English Written by AI
What the company’s 510(k) summary says about the device, its predicate and its testing, in plain words.
- What it is
- qCT Lung is artificial intelligence software that analyzes non-contrast CT scans of the chest. It uses a deep learning model, a type of convolutional neural network (a pattern-recognition system trained on large sets of images), to find pulmonary nodules (small spots in the lung) and describe them by texture, spiculation (spiky edges), calcification (calcium deposits), volume and diameter.
- What it’s cleared for
- The software is cleared as a concurrent reader for CT scans from the lung cancer screening population. It helps radiologists and pulmonologists detect nodules sized 4 to 30 mm and provides characterization and measurements. It is adjunctive information only, is not a replacement for a physician's review or clinical judgment, and must be used with the original CT series.
- Compared with (the predicate)
- K242652, Lunit INSIGHT DBT v1.1
- How it differs
- The summary names Lunit INSIGHT DBT v1.1 as the predicate. It is a breast cancer tool that reads digital breast tomosynthesis (3D mammography) images from women, while qCT Lung reads non-contrast chest CT images for lung nodules.
- Testing the summary lists
- Software testing (section 10.1 heading only; the summary text is cut off before any details)
Written by AI from the 510(k) summary the company filed with the FDA (the PDF), and checked against it: predicate numbers and figures must appear in the summary. A summary is the company’s own account, not the FDA’s review, and the AI can still get it wrong: read the PDF before you rely on it.
How the review time compares
Calendar days from the day FDA received the 510(k) to its decision. They include time the submission waited on the company for answers, so they run longer than FDA’s 90-day review goal.
K261517 took 130 calendar days from receipt to decision. Of the latest 24 other decisions in product code QDQ, 16 took longer, 7 were faster, 1 took the same time; their median was 164 days.
Latest decisions in QDQ
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K263749 | CadAI-B DxBeamWorks, Inc. · Special | Nov 6, 202621 days |
| K261517 | qCT Lung (v1.0)Qure.Ai Technologies Pvt. , Ltd. · Traditional | Sep 14, 2026130 days |
| K260898 | Transpara (2.1.0)Screenpoint Medical B.V. · Special | Aug 14, 2026149 days |
| K260086 | CadAI-B DxBeamWorks, Inc. · Traditional | Aug 12, 2026212 days |
| K260303 | See-Mode Augmented Reporting Tool, Breast (SMART-B)See-Mode Technologies Pty, Ltd. · Traditional | Jul 28, 2026179 days |
| K260714 | MammoScreen® (5)Therapixel · Special | Jun 26, 2026113 days |
| K253825 | Saige-DxDeepHealth, Inc. · Traditional | Jun 15, 2026196 days |
| K253682 | DeepHealth ProstateAIQuantib B.V. · Traditional | Apr 29, 2026159 days |
| K260320 | Lunit INSIGHT MMG (v1.1.10)Lunit, Inc. · Special | Apr 23, 202683 days |
| K253796 | Lunit INSIGHT DBT (V1.2)Lunit, Inc. · Traditional | Mar 26, 2026118 days |
Other 510(k)s from Qure.Ai Technologies Pvt. , Ltd.
The applicant’s other clearances in openFDA, newest first. Company names are matched as the FDA recorded them, so a renamed or differently spelled company may not appear.
This is the only 510(k) in openFDA under this applicant name.
Watch QDQ in DeviceVigil
DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and it tells you about new clearances, recalls and MAUDE reports for them, right beside the FDA’s own pages.
Get QDQ every Monday
One short email after openFDA’s weekly update with that week’s new 510(k)s, De Novos, PMAs and recalls in QDQ. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/510k · search k_number:K261517Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:QDQRetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/classification · search product_code:QDQRetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:QDQ AND decision_date:[20251129 TO 20261129]Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search applicant:"Qure.Ai Technologies Pvt. , Ltd."Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.