FDA device news · Product code NOT
FDA 510(k) clearance
OptraSCAN System (K260755): Optrascan, Inc. 510(k) clearance
San Jose, CA
- Decision
- Substantially Equivalent
- Decision date
- Jul 2, 2026
- Received
- Mar 9, 2026
- 510(k) type
- Traditional
- Product code
- NOTMicroscope, Automated, Image Analysis, Operator Intervention
- Review panel
- Hematology
- Device class
- Class II
- Regulation
- 21 CFR 864.1860
- Third-party review
- No
- Summary or statement
- Summary
- Review time
- 115 daysreceipt to decision, calendar days
- Median in NOT
- 169 daysthe latest 12 other decisions
- NOT decisions
- 1in the last 12 months
- Applicant’s 510(k)s
- 1in openFDA, this one included
In plain English Written by AI
What the company’s 510(k) summary says about the device, its predicate and its testing, in plain words.
- What it is
- The OptraSCAN System is an automated digital pathology system that converts glass microscope slides into high-resolution digital whole slide images, then lets a pathologist view, manage and analyze them on a computer. It includes a scanner, a color monitor, and a workstation loaded with image management and HER2 image analysis software.
- What it’s cleared for
- The system is cleared as an in vitro diagnostic aid for pathologists. It is used to review and semi-quantitatively score HER2 protein expression in scanned images of formalin-fixed, paraffin-embedded breast tissue slides stained with the Dako HercepTest. The pathologist makes the final score decision, and the software makes no independent interpretations.
- Compared with (the predicate)
- K141109, Aperio ePathology eIHC IVD System, AT Turbo and CS2 models
- How it differs
- The summary lists the subject device and predicate side by side. It says the subject device uses the OS-Ultra Scanner, Dell monitor and ImagePath software, while the predicate uses the ScanScope AT Turbo scanner, a color monitor and eSlideManager software.
- Testing the summary lists
- Precision and reproducibility analytical study, covering scanners, readers, timepoints and laboratory environments
Written by AI from the 510(k) summary the company filed with the FDA (the PDF), and checked against it: predicate numbers and figures must appear in the summary. A summary is the company’s own account, not the FDA’s review, and the AI can still get it wrong: read the PDF before you rely on it.
How the review time compares
Calendar days from the day FDA received the 510(k) to its decision. They include time the submission waited on the company for answers, so they run longer than FDA’s 90-day review goal.
K260755 took 115 calendar days from receipt to decision. Of the latest 12 other decisions in product code NOT, 9 took longer, 3 were faster; their median was 169 days.
Latest decisions in NOT
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K260755 | OptraSCAN SystemOptrascan, Inc. · Traditional | Jul 2, 2026115 days |
| K141109 | APERIO EPATHOLOGY EIHC IVD SYSTEMLeica Biosystems Imaging, Inc. · Traditional | Jul 29, 201491 days |
| K121033 | VIRTUOSO SYSTEM FOR IHC KI-67 (30-9)Ventana Medical Systems, Inc. · Traditional | Sep 6, 2013519 days |
| K121350 | VIRTUSO SYSTEM FOR IHC (DO-7)Ventana Medical Systems, Inc. · Traditional | Jun 1, 201228 days |
| K111755 | VIRTUOSO SYSTEM FOR IHC KI-67 (30-9)Ventana Medical Systems, Inc. · Traditional | Feb 22, 2012245 days |
| K111543 | VIRTUOSO (TM) SYSTEM FOR IHC HER2 (4B5)Ventana Medical Systems, Inc. · Traditional | Oct 12, 2011132 days |
| K080564 | SCANSCOPE XT SYSTEMAperio Technologies · Traditional | Aug 14, 2009533 days |
| K080910 | PATHIAM SYSTEM FOR HER2/NEU IMMUNOHISTOCHEMISTRY REAGENTS & KITSBioimagene, Inc. · Traditional | Feb 4, 2009309 days |
| K071128 | SCANSCOPE XT SYSTEMAperio Technologies · Traditional | Oct 10, 2007170 days |
| K062756 | PATHIAM IMAGING SOFTWARE FOR HER2/NEUBioimagene, Inc. · Traditional | Feb 20, 2007159 days |
Other 510(k)s from Optrascan, Inc.
The applicant’s other clearances in openFDA, newest first. Company names are matched as the FDA recorded them, so a renamed or differently spelled company may not appear.
This is the only 510(k) in openFDA under this applicant name.
Watch NOT in DeviceVigil
DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and it tells you about new clearances, recalls and MAUDE reports for them, right beside the FDA’s own pages.
Get NOT every Monday
One short email after openFDA’s weekly update with that week’s new 510(k)s, De Novos, PMAs and recalls in NOT. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/510k · search k_number:K260755Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:NOTRetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/classification · search product_code:NOTRetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:NOT AND decision_date:[20251129 TO 20261129]Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search applicant:"Optrascan, Inc."Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil