FDA device news · Product code NOT

FDA 510(k) clearance

OptraSCAN System (K260755): Optrascan, Inc. 510(k) clearance

San Jose, CA

Follow NOT: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Decision
Substantially Equivalent
Decision date
Jul 2, 2026
Received
Mar 9, 2026
510(k) type
Traditional
Product code
NOTMicroscope, Automated, Image Analysis, Operator Intervention
Review panel
Hematology
Device class
Class II
Regulation
21 CFR 864.1860
Third-party review
No
Summary or statement
Summary
Review time
115 daysreceipt to decision, calendar days
Median in NOT
169 daysthe latest 12 other decisions
NOT decisions
1in the last 12 months
Applicant’s 510(k)s
1in openFDA, this one included

In plain English Written by AI

What the company’s 510(k) summary says about the device, its predicate and its testing, in plain words.

What it is
The OptraSCAN System is an automated digital pathology system that converts glass microscope slides into high-resolution digital whole slide images, then lets a pathologist view, manage and analyze them on a computer. It includes a scanner, a color monitor, and a workstation loaded with image management and HER2 image analysis software.
What it’s cleared for
The system is cleared as an in vitro diagnostic aid for pathologists. It is used to review and semi-quantitatively score HER2 protein expression in scanned images of formalin-fixed, paraffin-embedded breast tissue slides stained with the Dako HercepTest. The pathologist makes the final score decision, and the software makes no independent interpretations.
Compared with (the predicate)
  • K141109, Aperio ePathology eIHC IVD System, AT Turbo and CS2 models
How it differs
The summary lists the subject device and predicate side by side. It says the subject device uses the OS-Ultra Scanner, Dell monitor and ImagePath software, while the predicate uses the ScanScope AT Turbo scanner, a color monitor and eSlideManager software.
Testing the summary lists
  • Precision and reproducibility analytical study, covering scanners, readers, timepoints and laboratory environments

Written by AI from the 510(k) summary the company filed with the FDA (the PDF), and checked against it: predicate numbers and figures must appear in the summary. A summary is the company’s own account, not the FDA’s review, and the AI can still get it wrong: read the PDF before you rely on it.

How the review time compares

Calendar days from the day FDA received the 510(k) to its decision. They include time the submission waited on the company for answers, so they run longer than FDA’s 90-day review goal.

K260755 took 115 calendar days from receipt to decision. Of the latest 12 other decisions in product code NOT, 9 took longer, 3 were faster; their median was 169 days.

Latest decisions in NOT

NumberDevice and applicantDecision and review time
K260755 OptraSCAN SystemOptrascan, Inc. · Traditional Jul 2, 2026115 days
K141109 APERIO EPATHOLOGY EIHC IVD SYSTEMLeica Biosystems Imaging, Inc. · Traditional Jul 29, 201491 days
K121033 VIRTUOSO SYSTEM FOR IHC KI-67 (30-9)Ventana Medical Systems, Inc. · Traditional Sep 6, 2013519 days
K121350 VIRTUSO SYSTEM FOR IHC (DO-7)Ventana Medical Systems, Inc. · Traditional Jun 1, 201228 days
K111755 VIRTUOSO SYSTEM FOR IHC KI-67 (30-9)Ventana Medical Systems, Inc. · Traditional Feb 22, 2012245 days
K111543 VIRTUOSO (TM) SYSTEM FOR IHC HER2 (4B5)Ventana Medical Systems, Inc. · Traditional Oct 12, 2011132 days
K080564 SCANSCOPE XT SYSTEMAperio Technologies · Traditional Aug 14, 2009533 days
K080910 PATHIAM SYSTEM FOR HER2/NEU IMMUNOHISTOCHEMISTRY REAGENTS & KITSBioimagene, Inc. · Traditional Feb 4, 2009309 days
K071128 SCANSCOPE XT SYSTEMAperio Technologies · Traditional Oct 10, 2007170 days
K062756 PATHIAM IMAGING SOFTWARE FOR HER2/NEUBioimagene, Inc. · Traditional Feb 20, 2007159 days

Other 510(k)s from Optrascan, Inc.

The applicant’s other clearances in openFDA, newest first. Company names are matched as the FDA recorded them, so a renamed or differently spelled company may not appear.

This is the only 510(k) in openFDA under this applicant name.

Watch NOT in DeviceVigil

DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and it tells you about new clearances, recalls and MAUDE reports for them, right beside the FDA’s own pages.

Add to Chrome — free Add to Edge

Get NOT every Monday

One short email after openFDA’s weekly update with that week’s new 510(k)s, De Novos, PMAs and recalls in NOT. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.