FDA device news · Product code OMP

FDA 510(k) clearance

MeCare Alpha (MCA2019) (K260426): IVT Medical, Ltd. 510(k) clearance

Ra'Anana, IL

Follow OMP: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Decision
Substantially Equivalent
Decision date
Oct 8, 2026
Received
Feb 10, 2026
510(k) type
Traditional
Product code
OMPNegative Pressure Wound Therapy Powered Suction Pump
Review panel
General, Plastic Surgery
Device class
Class II
Regulation
21 CFR 878.4780
Third-party review
No
Summary or statement
Summary
Review time
240 daysreceipt to decision, calendar days
Median in OMP
141 daysthe latest 24 other decisions
OMP decisions
11in the last 12 months
Applicant’s 510(k)s
2in openFDA, this one included

In plain English Written by AI

What the company’s 510(k) summary says about the device, its predicate and its testing, in plain words.

What it is
The MeCare α System (Model MCA2019) is a portable, battery-powered negative pressure wound therapy device for use in professional healthcare facilities. A software-regulated vacuum pump in a reusable control unit applies a set level of suction through tubing and a collecting canister to a sealed wound dressing, removing wound fluid (exudate) into the canister.
What it’s cleared for
It is cleared for wound management by applying a pre-set level of negative pressure to remove fluids, including wound exudate and infectious materials. It is indicated for small- to medium-sized chronic, acute, traumatic, sub-acute and dehisced wounds, partial-thickness burns, diabetic or pressure ulcers, flaps and grafts, in patients with a low risk of bleeding. It is for use only by trained healthcare professionals in professional healthcare facilities, not at home, and is prescription only.
Compared with (the predicate)
How it differs
The summary says the device runs at a narrower, pre-set pressure range (40, 50 or 75 mmHg, with a 90 mmHg software maximum) instead of the predicate's adjustable 30 to 200 mmHg.
Testing the summary lists
  • Electrical safety (IEC 60601-1, ANSI/AAMI ES60601-1)
  • Electromagnetic compatibility (IEC 60601-1-2)
  • Battery safety (IEC 62133-2, including Clause 5.6)
  • Delivered negative pressure under simulated clinical use over seven days, using three units in a simulated wound model
  • Battery endurance, full canister alarm, low vacuum alarm and tilt alarm testing, with three units for most tests

Written by AI from the 510(k) summary the company filed with the FDA (the PDF), and checked against it: predicate numbers and figures must appear in the summary. A summary is the company’s own account, not the FDA’s review, and the AI can still get it wrong: read the PDF before you rely on it.

How the review time compares

Calendar days from the day FDA received the 510(k) to its decision. They include time the submission waited on the company for answers, so they run longer than FDA’s 90-day review goal.

K260426 took 240 calendar days from receipt to decision. Of the latest 24 other decisions in product code OMP, 9 took longer, 15 were faster; their median was 141 days.

Latest decisions in OMP

NumberDevice and applicantDecision and review time
K262232 Wound VOX NPWT System (V1-01-01); Wound VOX 300ml Collection Canister & Supply Tubing (CC-00-03); Wound VOX Y-Connector (YK-01-13); Wound VOX Dressing Kit Small (DK-01-01); Wound VOX Dressing Kit Medium (DK-01-02); Wound VOX Dressing Kit Large (DK-01-03); Wound VOX Adhesive Drape (DK-01-04); Wound VOX Foam Dressing (DK-01-05); Wound VOX Suction Bell Assembly (DK-01-06); Wound VOX Wound Contact Layer (10cmx8cm) (DK-01-07); Wound VOX Wound Contact Layer (25cmx17cm) (DK-01-14)Bechtel Medical, Inc. · Traditional Nov 6, 2026129 days
K260306 RENASYS-G 15Fr Channel Drain Gauze Dressing Kit (66802257); RENASYS 15Fr Channel Drain Accessory Kit (66802253)Smith & Nephew Medical, Ltd. · Traditional Oct 30, 2026273 days
K261982 RENASYS ABDO SHIELD Temporary Abdominal Closure Kit (66807747)Smith & Nephew Medical, Ltd. · Traditional Oct 23, 2026133 days
K260426 MeCare Alpha (MCA2019)IVT Medical, Ltd. · Traditional Oct 8, 2026240 days
K261338 RENASYS DYNAFLOW Kit (66807741)Smith & Nephew Medical, Ltd. · Traditional Sep 11, 2026141 days
K253263 ConvaVAC™ Negative Pressure Wound Therapy SystemConvatec Limited, Gdc Building · Traditional Jun 24, 2026268 days
K253199 CURASYS 2 (EMDN0003)CG Bio Co., Ltd. · Traditional Jun 23, 2026270 days
K260646 RENASYS EDGE (66803126)Smith & Nephew Medical, Ltd. · Traditional May 27, 202689 days
K261291 Nisus NPWT Canister 500-mL (CPC-500)Cork Medical · Special May 21, 202631 days
K254019 Wound Geni NPWT SystemBechtel Medical, Inc. · Traditional May 4, 2026140 days

Other 510(k)s from IVT Medical, Ltd.

The applicant’s other clearances in openFDA, newest first. Company names are matched as the FDA recorded them, so a renamed or differently spelled company may not appear.

NumberDevice and applicantDecision and review time
K121817 VCARE AIVT Medical, Ltd. · Traditional Sep 17, 2013454 days

Watch OMP in DeviceVigil

DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and it tells you about new clearances, recalls and MAUDE reports for them, right beside the FDA’s own pages.

Add to Chrome — free Add to Edge

Get OMP every Monday

One short email after openFDA’s weekly update with that week’s new 510(k)s, De Novos, PMAs and recalls in OMP. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.